Cargando…
Effectiveness of Inspiratory Termination Synchrony with Automatic Cycling During Noninvasive Pressure Support Ventilation
BACKGROUND: Pressure support ventilation (PSV) is a standard method for non-invasive home ventilation. A bench study was designed to compare the effectiveness of patient-ventilator inspiratory termination synchronization with automated and conventional triggering in various respiratory mechanics mod...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
International Scientific Literature, Inc.
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4915317/ https://www.ncbi.nlm.nih.gov/pubmed/27198165 http://dx.doi.org/10.12659/MSM.896059 |
_version_ | 1782438684659810304 |
---|---|
author | Chen, Yuqing Cheng, Kewen Zhou, Xin |
author_facet | Chen, Yuqing Cheng, Kewen Zhou, Xin |
author_sort | Chen, Yuqing |
collection | PubMed |
description | BACKGROUND: Pressure support ventilation (PSV) is a standard method for non-invasive home ventilation. A bench study was designed to compare the effectiveness of patient-ventilator inspiratory termination synchronization with automated and conventional triggering in various respiratory mechanics models. MATERIAL/METHODS: Two ventilators, the Respironics V60 and Curative Flexo ST 30, connected to a Hans Rudolph Series 1101 lung simulator, were evaluated using settings that simulate lung mechanics in patients with chronic obstructive pulmonary disease (COPD), acute respiratory distress syndrome (ARDS), or normal lungs. Ventilators were operated with automated (Auto-Trak) or conventional high-, moderate-, and low-sensitivity flow-cycling software algorithms, 5 cmH(2)O or 15 cmH(2)O pressure support, 5 cmH(2)O positive end-expiratory pressure (PEEP), and an air leak of 25–28 L/min. RESULTS: Both ventilators adapted to the system leak without requiring adjustment of triggering settings. In all simulated lung conditions, automated cycling resulted in shorter triggering delay times (<100 ms) and lower triggering pressure-time product (PTPt) values. Tidal volumes (V(T)) increased with lower conventional cycling sensitivity level. In the COPD model, automated cycling had higher leak volumes and shorter cycling delay times than in conventional cycling. Asynchronous events were rare. Inspiratory time (Tinsp), peak expiratory flow (PEF), and cycling off delay time (Cdelay) increased as a result of reduction in conventional cycling sensitivity level. In the ARDS and normal adult lung models, premature cycling was frequent at the high-sensitive cycling level. CONCLUSIONS: Overall, the Auto-Trak protocol showed better patient-machine cycling synchronization than conventional triggering. This was evident by shorter triggering time delays and lower PTPt. |
format | Online Article Text |
id | pubmed-4915317 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | International Scientific Literature, Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-49153172016-06-28 Effectiveness of Inspiratory Termination Synchrony with Automatic Cycling During Noninvasive Pressure Support Ventilation Chen, Yuqing Cheng, Kewen Zhou, Xin Med Sci Monit Medical Technology BACKGROUND: Pressure support ventilation (PSV) is a standard method for non-invasive home ventilation. A bench study was designed to compare the effectiveness of patient-ventilator inspiratory termination synchronization with automated and conventional triggering in various respiratory mechanics models. MATERIAL/METHODS: Two ventilators, the Respironics V60 and Curative Flexo ST 30, connected to a Hans Rudolph Series 1101 lung simulator, were evaluated using settings that simulate lung mechanics in patients with chronic obstructive pulmonary disease (COPD), acute respiratory distress syndrome (ARDS), or normal lungs. Ventilators were operated with automated (Auto-Trak) or conventional high-, moderate-, and low-sensitivity flow-cycling software algorithms, 5 cmH(2)O or 15 cmH(2)O pressure support, 5 cmH(2)O positive end-expiratory pressure (PEEP), and an air leak of 25–28 L/min. RESULTS: Both ventilators adapted to the system leak without requiring adjustment of triggering settings. In all simulated lung conditions, automated cycling resulted in shorter triggering delay times (<100 ms) and lower triggering pressure-time product (PTPt) values. Tidal volumes (V(T)) increased with lower conventional cycling sensitivity level. In the COPD model, automated cycling had higher leak volumes and shorter cycling delay times than in conventional cycling. Asynchronous events were rare. Inspiratory time (Tinsp), peak expiratory flow (PEF), and cycling off delay time (Cdelay) increased as a result of reduction in conventional cycling sensitivity level. In the ARDS and normal adult lung models, premature cycling was frequent at the high-sensitive cycling level. CONCLUSIONS: Overall, the Auto-Trak protocol showed better patient-machine cycling synchronization than conventional triggering. This was evident by shorter triggering time delays and lower PTPt. International Scientific Literature, Inc. 2016-05-20 /pmc/articles/PMC4915317/ /pubmed/27198165 http://dx.doi.org/10.12659/MSM.896059 Text en © Med Sci Monit, 2016 This work is licensed under Creative Common Attribution-NonCommercial-NoDerivatives 4.0 International (CC BY-NC-ND 4.0) |
spellingShingle | Medical Technology Chen, Yuqing Cheng, Kewen Zhou, Xin Effectiveness of Inspiratory Termination Synchrony with Automatic Cycling During Noninvasive Pressure Support Ventilation |
title | Effectiveness of Inspiratory Termination Synchrony with Automatic Cycling During Noninvasive Pressure Support Ventilation |
title_full | Effectiveness of Inspiratory Termination Synchrony with Automatic Cycling During Noninvasive Pressure Support Ventilation |
title_fullStr | Effectiveness of Inspiratory Termination Synchrony with Automatic Cycling During Noninvasive Pressure Support Ventilation |
title_full_unstemmed | Effectiveness of Inspiratory Termination Synchrony with Automatic Cycling During Noninvasive Pressure Support Ventilation |
title_short | Effectiveness of Inspiratory Termination Synchrony with Automatic Cycling During Noninvasive Pressure Support Ventilation |
title_sort | effectiveness of inspiratory termination synchrony with automatic cycling during noninvasive pressure support ventilation |
topic | Medical Technology |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4915317/ https://www.ncbi.nlm.nih.gov/pubmed/27198165 http://dx.doi.org/10.12659/MSM.896059 |
work_keys_str_mv | AT chenyuqing effectivenessofinspiratoryterminationsynchronywithautomaticcyclingduringnoninvasivepressuresupportventilation AT chengkewen effectivenessofinspiratoryterminationsynchronywithautomaticcyclingduringnoninvasivepressuresupportventilation AT zhouxin effectivenessofinspiratoryterminationsynchronywithautomaticcyclingduringnoninvasivepressuresupportventilation |