Accuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants
Background. In premature infants, maintaining blood partial pressure of carbon dioxide (pCO(2)) value within a narrow range is important to avoid cerebral lesions. The aim of this study was to assess the accuracy of a noninvasive transcutaneous method (TcpCO(2)), compared to blood partial pressure o...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Hindawi Publishing Corporation
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4916268/ https://www.ncbi.nlm.nih.gov/pubmed/27375901 http://dx.doi.org/10.1155/2016/8041967 |
_version_ | 1782438800119562240 |
---|---|
author | Janaillac, Marie Labarinas, Sonia Pfister, Riccardo E. Karam, Oliver |
author_facet | Janaillac, Marie Labarinas, Sonia Pfister, Riccardo E. Karam, Oliver |
author_sort | Janaillac, Marie |
collection | PubMed |
description | Background. In premature infants, maintaining blood partial pressure of carbon dioxide (pCO(2)) value within a narrow range is important to avoid cerebral lesions. The aim of this study was to assess the accuracy of a noninvasive transcutaneous method (TcpCO(2)), compared to blood partial pressure of carbon dioxide (pCO(2)). Methods. Retrospective observational study in a tertiary neonatal intensive care unit. We analyzed the correlation between blood pCO(2) and transcutaneous values and the accuracy between the trends of blood pCO(2) and TcpCO(2) in all consecutive premature infants born at <33 weeks' gestational age. Results. 248 infants were included (median gestational age: 29 + 5 weeks and median birth weight: 1250 g), providing 1365 pairs of TcpCO(2) and blood pCO(2) values. Pearson's R correlation between these values was 0.58. The mean bias was −0.93 kPa with a 95% confidence limit of agreement of −4.05 to +2.16 kPa. Correlation between the trends of TcpCO(2) and blood pCO(2) values was good in only 39.6%. Conclusions. In premature infants, TcpCO(2) was poorly correlated to blood pCO(2), with a wide limit of agreement. Furthermore, concordance between trends was equally low. We warn about clinical decision-making on TcpCO(2) alone when used as continuous monitoring. |
format | Online Article Text |
id | pubmed-4916268 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | Hindawi Publishing Corporation |
record_format | MEDLINE/PubMed |
spelling | pubmed-49162682016-07-03 Accuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants Janaillac, Marie Labarinas, Sonia Pfister, Riccardo E. Karam, Oliver Crit Care Res Pract Research Article Background. In premature infants, maintaining blood partial pressure of carbon dioxide (pCO(2)) value within a narrow range is important to avoid cerebral lesions. The aim of this study was to assess the accuracy of a noninvasive transcutaneous method (TcpCO(2)), compared to blood partial pressure of carbon dioxide (pCO(2)). Methods. Retrospective observational study in a tertiary neonatal intensive care unit. We analyzed the correlation between blood pCO(2) and transcutaneous values and the accuracy between the trends of blood pCO(2) and TcpCO(2) in all consecutive premature infants born at <33 weeks' gestational age. Results. 248 infants were included (median gestational age: 29 + 5 weeks and median birth weight: 1250 g), providing 1365 pairs of TcpCO(2) and blood pCO(2) values. Pearson's R correlation between these values was 0.58. The mean bias was −0.93 kPa with a 95% confidence limit of agreement of −4.05 to +2.16 kPa. Correlation between the trends of TcpCO(2) and blood pCO(2) values was good in only 39.6%. Conclusions. In premature infants, TcpCO(2) was poorly correlated to blood pCO(2), with a wide limit of agreement. Furthermore, concordance between trends was equally low. We warn about clinical decision-making on TcpCO(2) alone when used as continuous monitoring. Hindawi Publishing Corporation 2016 2016-06-08 /pmc/articles/PMC4916268/ /pubmed/27375901 http://dx.doi.org/10.1155/2016/8041967 Text en Copyright © 2016 Marie Janaillac et al. https://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Janaillac, Marie Labarinas, Sonia Pfister, Riccardo E. Karam, Oliver Accuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants |
title | Accuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants |
title_full | Accuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants |
title_fullStr | Accuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants |
title_full_unstemmed | Accuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants |
title_short | Accuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants |
title_sort | accuracy of transcutaneous carbon dioxide measurement in premature infants |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4916268/ https://www.ncbi.nlm.nih.gov/pubmed/27375901 http://dx.doi.org/10.1155/2016/8041967 |
work_keys_str_mv | AT janaillacmarie accuracyoftranscutaneouscarbondioxidemeasurementinprematureinfants AT labarinassonia accuracyoftranscutaneouscarbondioxidemeasurementinprematureinfants AT pfisterriccardoe accuracyoftranscutaneouscarbondioxidemeasurementinprematureinfants AT karamoliver accuracyoftranscutaneouscarbondioxidemeasurementinprematureinfants |