High False Positives and False Negatives in Yeast Parameter in an Automated Urine Sediment Analyzer
BACKGROUND: Automated urine sediment analyzers have proven their feasibility in medical laboratories. However, editing manual microscopic review of some specimens severely limits the usefulness of such systems. This study aims to give feedback on the practical experience on »Yeast«, which is one of...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Society of Medical Biochemists of Serbia
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4922345/ https://www.ncbi.nlm.nih.gov/pubmed/28356844 http://dx.doi.org/10.2478/jomb-2014-0052 |
_version_ | 1782439608311611392 |
---|---|
author | Aydin, Ozgur Ellidag, Hamit Yasar Eren, Esin Yilmaz, Necat |
author_facet | Aydin, Ozgur Ellidag, Hamit Yasar Eren, Esin Yilmaz, Necat |
author_sort | Aydin, Ozgur |
collection | PubMed |
description | BACKGROUND: Automated urine sediment analyzers have proven their feasibility in medical laboratories. However, editing manual microscopic review of some specimens severely limits the usefulness of such systems. This study aims to give feedback on the practical experience on »Yeast«, which is one of the parameters that compel frequent manual reviews. METHODS: 5448 freshly collected urine specimens submitted from various departments of our hospital for diagnostic urinalysis were studied by the UriSed® (77 Elektronika, Hungary). A specialist medical doctor inspected every image on-board, and reviewed the ones with a »Yeast« alarm by traditional manual microscopy. RESULTS: UriSed alarmed in 491 samples (9%) for yeast. In 59 samples (1%) the number of particles exceeded the cut-off and a »positive for yeast« was set. A false positive report of yeast +1 to 3+/HPF was found in 51 samples (0.9%). There were 8 cases with positive for yeast from both microscopic methods. Thirty-three »negative for yeast« samples were corrected as positive after the manual microscopic review. CONCLUSIONS: We report a high percentage of false positives and negatives in the yeast parameter, in line with other studies on UriSed as well as on other instruments in the market. As an important feedback, our observations showed that the major concern in false results was »the focusing problem«. We believe in the necessity of a focus check and comparison of alarms between images on board. |
format | Online Article Text |
id | pubmed-4922345 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | Society of Medical Biochemists of Serbia |
record_format | MEDLINE/PubMed |
spelling | pubmed-49223452017-03-29 High False Positives and False Negatives in Yeast Parameter in an Automated Urine Sediment Analyzer Aydin, Ozgur Ellidag, Hamit Yasar Eren, Esin Yilmaz, Necat J Med Biochem Original Paper BACKGROUND: Automated urine sediment analyzers have proven their feasibility in medical laboratories. However, editing manual microscopic review of some specimens severely limits the usefulness of such systems. This study aims to give feedback on the practical experience on »Yeast«, which is one of the parameters that compel frequent manual reviews. METHODS: 5448 freshly collected urine specimens submitted from various departments of our hospital for diagnostic urinalysis were studied by the UriSed® (77 Elektronika, Hungary). A specialist medical doctor inspected every image on-board, and reviewed the ones with a »Yeast« alarm by traditional manual microscopy. RESULTS: UriSed alarmed in 491 samples (9%) for yeast. In 59 samples (1%) the number of particles exceeded the cut-off and a »positive for yeast« was set. A false positive report of yeast +1 to 3+/HPF was found in 51 samples (0.9%). There were 8 cases with positive for yeast from both microscopic methods. Thirty-three »negative for yeast« samples were corrected as positive after the manual microscopic review. CONCLUSIONS: We report a high percentage of false positives and negatives in the yeast parameter, in line with other studies on UriSed as well as on other instruments in the market. As an important feedback, our observations showed that the major concern in false results was »the focusing problem«. We believe in the necessity of a focus check and comparison of alarms between images on board. Society of Medical Biochemists of Serbia 2015-07 2015-07-14 /pmc/articles/PMC4922345/ /pubmed/28356844 http://dx.doi.org/10.2478/jomb-2014-0052 Text en © by Ozgur Aydin http://creativecommons.org/licenses/by-nc-nd/3.0/ This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License. |
spellingShingle | Original Paper Aydin, Ozgur Ellidag, Hamit Yasar Eren, Esin Yilmaz, Necat High False Positives and False Negatives in Yeast Parameter in an Automated Urine Sediment Analyzer |
title | High False Positives and False Negatives in Yeast Parameter in an Automated Urine Sediment Analyzer |
title_full | High False Positives and False Negatives in Yeast Parameter in an Automated Urine Sediment Analyzer |
title_fullStr | High False Positives and False Negatives in Yeast Parameter in an Automated Urine Sediment Analyzer |
title_full_unstemmed | High False Positives and False Negatives in Yeast Parameter in an Automated Urine Sediment Analyzer |
title_short | High False Positives and False Negatives in Yeast Parameter in an Automated Urine Sediment Analyzer |
title_sort | high false positives and false negatives in yeast parameter in an automated urine sediment analyzer |
topic | Original Paper |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4922345/ https://www.ncbi.nlm.nih.gov/pubmed/28356844 http://dx.doi.org/10.2478/jomb-2014-0052 |
work_keys_str_mv | AT aydinozgur highfalsepositivesandfalsenegativesinyeastparameterinanautomatedurinesedimentanalyzer AT ellidaghamityasar highfalsepositivesandfalsenegativesinyeastparameterinanautomatedurinesedimentanalyzer AT erenesin highfalsepositivesandfalsenegativesinyeastparameterinanautomatedurinesedimentanalyzer AT yilmaznecat highfalsepositivesandfalsenegativesinyeastparameterinanautomatedurinesedimentanalyzer |