Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth

Purpose. To compare between a new optical biometer (AL-Scan, Nidek Co., Aichi, Japan) and an anterior segment optical coherence tomographer (Visante AS-OCT, Carl Zeiss Meditec, Dublin, USA) for measuring central corneal thickness (CCT), anterior chamber depth (ACD), and aqueous depth (AD). Methods....

Descripción completa

Detalles Bibliográficos
Autores principales: Huang, Jinhai, Lu, Weicong, Savini, Giacomo, Chen, Hao, Wang, Chengfang, Yu, Xinxin, Bao, Fangjun, Wang, Qinmei
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4923600/
https://www.ncbi.nlm.nih.gov/pubmed/27403339
http://dx.doi.org/10.1155/2016/6347236
_version_ 1782439738434650112
author Huang, Jinhai
Lu, Weicong
Savini, Giacomo
Chen, Hao
Wang, Chengfang
Yu, Xinxin
Bao, Fangjun
Wang, Qinmei
author_facet Huang, Jinhai
Lu, Weicong
Savini, Giacomo
Chen, Hao
Wang, Chengfang
Yu, Xinxin
Bao, Fangjun
Wang, Qinmei
author_sort Huang, Jinhai
collection PubMed
description Purpose. To compare between a new optical biometer (AL-Scan, Nidek Co., Aichi, Japan) and an anterior segment optical coherence tomographer (Visante AS-OCT, Carl Zeiss Meditec, Dublin, USA) for measuring central corneal thickness (CCT), anterior chamber depth (ACD), and aqueous depth (AD). Methods. Sixty-three eyes of 63 normal subjects were examined with AL-Scan and Visante AS-OCT in this prospective study. One eye per subject was measured three times with both devices to record their CCT, ACD, and AD. All procedures were performed by the same operator. Agreement between the two devices was assessed using paired t-tests, Bland-Altman plots, and 95% limits of agreement (LoA). Results. The mean CCT, ACD, and AD measured by AL-Scan were 538.59 ± 27.37 μm, 3.70 ± 0.30 mm, and 3.16 ± 0.30 mm, respectively. The mean values obtained by the Visante OCT were 536.14 ± 26.61 μm for CCT, 3.71 ± 0.29 mm for ACD, and 3.17 ± 0.29 mm for AD. The mean CCT by the AL-Scan was higher than that obtained by the Visante AS-OCT (difference = 2.45 ± 6.07 μm, P < 0.05). The differences in ACD and AD measurements were not statistically significant. The 95% LoA of CCT, ACD, and AD were between −9.44 and 14.35 μm, −0.15 and 0.12 mm, and −0.15 and 0.12 mm, respectively. Conclusions. Since these two devices were comparable for measuring CCT, ACD, and AD, their results can be interchangeably used in the clinic.
format Online
Article
Text
id pubmed-4923600
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Hindawi Publishing Corporation
record_format MEDLINE/PubMed
spelling pubmed-49236002016-07-11 Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth Huang, Jinhai Lu, Weicong Savini, Giacomo Chen, Hao Wang, Chengfang Yu, Xinxin Bao, Fangjun Wang, Qinmei J Ophthalmol Research Article Purpose. To compare between a new optical biometer (AL-Scan, Nidek Co., Aichi, Japan) and an anterior segment optical coherence tomographer (Visante AS-OCT, Carl Zeiss Meditec, Dublin, USA) for measuring central corneal thickness (CCT), anterior chamber depth (ACD), and aqueous depth (AD). Methods. Sixty-three eyes of 63 normal subjects were examined with AL-Scan and Visante AS-OCT in this prospective study. One eye per subject was measured three times with both devices to record their CCT, ACD, and AD. All procedures were performed by the same operator. Agreement between the two devices was assessed using paired t-tests, Bland-Altman plots, and 95% limits of agreement (LoA). Results. The mean CCT, ACD, and AD measured by AL-Scan were 538.59 ± 27.37 μm, 3.70 ± 0.30 mm, and 3.16 ± 0.30 mm, respectively. The mean values obtained by the Visante OCT were 536.14 ± 26.61 μm for CCT, 3.71 ± 0.29 mm for ACD, and 3.17 ± 0.29 mm for AD. The mean CCT by the AL-Scan was higher than that obtained by the Visante AS-OCT (difference = 2.45 ± 6.07 μm, P < 0.05). The differences in ACD and AD measurements were not statistically significant. The 95% LoA of CCT, ACD, and AD were between −9.44 and 14.35 μm, −0.15 and 0.12 mm, and −0.15 and 0.12 mm, respectively. Conclusions. Since these two devices were comparable for measuring CCT, ACD, and AD, their results can be interchangeably used in the clinic. Hindawi Publishing Corporation 2016 2016-06-14 /pmc/articles/PMC4923600/ /pubmed/27403339 http://dx.doi.org/10.1155/2016/6347236 Text en Copyright © 2016 Jinhai Huang et al. https://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Huang, Jinhai
Lu, Weicong
Savini, Giacomo
Chen, Hao
Wang, Chengfang
Yu, Xinxin
Bao, Fangjun
Wang, Qinmei
Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth
title Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth
title_full Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth
title_fullStr Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth
title_full_unstemmed Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth
title_short Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth
title_sort comparison between a new optical biometry device and an anterior segment optical coherence tomographer for measuring central corneal thickness and anterior chamber depth
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4923600/
https://www.ncbi.nlm.nih.gov/pubmed/27403339
http://dx.doi.org/10.1155/2016/6347236
work_keys_str_mv AT huangjinhai comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth
AT luweicong comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth
AT savinigiacomo comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth
AT chenhao comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth
AT wangchengfang comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth
AT yuxinxin comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth
AT baofangjun comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth
AT wangqinmei comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth