Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth
Purpose. To compare between a new optical biometer (AL-Scan, Nidek Co., Aichi, Japan) and an anterior segment optical coherence tomographer (Visante AS-OCT, Carl Zeiss Meditec, Dublin, USA) for measuring central corneal thickness (CCT), anterior chamber depth (ACD), and aqueous depth (AD). Methods....
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Hindawi Publishing Corporation
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4923600/ https://www.ncbi.nlm.nih.gov/pubmed/27403339 http://dx.doi.org/10.1155/2016/6347236 |
_version_ | 1782439738434650112 |
---|---|
author | Huang, Jinhai Lu, Weicong Savini, Giacomo Chen, Hao Wang, Chengfang Yu, Xinxin Bao, Fangjun Wang, Qinmei |
author_facet | Huang, Jinhai Lu, Weicong Savini, Giacomo Chen, Hao Wang, Chengfang Yu, Xinxin Bao, Fangjun Wang, Qinmei |
author_sort | Huang, Jinhai |
collection | PubMed |
description | Purpose. To compare between a new optical biometer (AL-Scan, Nidek Co., Aichi, Japan) and an anterior segment optical coherence tomographer (Visante AS-OCT, Carl Zeiss Meditec, Dublin, USA) for measuring central corneal thickness (CCT), anterior chamber depth (ACD), and aqueous depth (AD). Methods. Sixty-three eyes of 63 normal subjects were examined with AL-Scan and Visante AS-OCT in this prospective study. One eye per subject was measured three times with both devices to record their CCT, ACD, and AD. All procedures were performed by the same operator. Agreement between the two devices was assessed using paired t-tests, Bland-Altman plots, and 95% limits of agreement (LoA). Results. The mean CCT, ACD, and AD measured by AL-Scan were 538.59 ± 27.37 μm, 3.70 ± 0.30 mm, and 3.16 ± 0.30 mm, respectively. The mean values obtained by the Visante OCT were 536.14 ± 26.61 μm for CCT, 3.71 ± 0.29 mm for ACD, and 3.17 ± 0.29 mm for AD. The mean CCT by the AL-Scan was higher than that obtained by the Visante AS-OCT (difference = 2.45 ± 6.07 μm, P < 0.05). The differences in ACD and AD measurements were not statistically significant. The 95% LoA of CCT, ACD, and AD were between −9.44 and 14.35 μm, −0.15 and 0.12 mm, and −0.15 and 0.12 mm, respectively. Conclusions. Since these two devices were comparable for measuring CCT, ACD, and AD, their results can be interchangeably used in the clinic. |
format | Online Article Text |
id | pubmed-4923600 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | Hindawi Publishing Corporation |
record_format | MEDLINE/PubMed |
spelling | pubmed-49236002016-07-11 Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth Huang, Jinhai Lu, Weicong Savini, Giacomo Chen, Hao Wang, Chengfang Yu, Xinxin Bao, Fangjun Wang, Qinmei J Ophthalmol Research Article Purpose. To compare between a new optical biometer (AL-Scan, Nidek Co., Aichi, Japan) and an anterior segment optical coherence tomographer (Visante AS-OCT, Carl Zeiss Meditec, Dublin, USA) for measuring central corneal thickness (CCT), anterior chamber depth (ACD), and aqueous depth (AD). Methods. Sixty-three eyes of 63 normal subjects were examined with AL-Scan and Visante AS-OCT in this prospective study. One eye per subject was measured three times with both devices to record their CCT, ACD, and AD. All procedures were performed by the same operator. Agreement between the two devices was assessed using paired t-tests, Bland-Altman plots, and 95% limits of agreement (LoA). Results. The mean CCT, ACD, and AD measured by AL-Scan were 538.59 ± 27.37 μm, 3.70 ± 0.30 mm, and 3.16 ± 0.30 mm, respectively. The mean values obtained by the Visante OCT were 536.14 ± 26.61 μm for CCT, 3.71 ± 0.29 mm for ACD, and 3.17 ± 0.29 mm for AD. The mean CCT by the AL-Scan was higher than that obtained by the Visante AS-OCT (difference = 2.45 ± 6.07 μm, P < 0.05). The differences in ACD and AD measurements were not statistically significant. The 95% LoA of CCT, ACD, and AD were between −9.44 and 14.35 μm, −0.15 and 0.12 mm, and −0.15 and 0.12 mm, respectively. Conclusions. Since these two devices were comparable for measuring CCT, ACD, and AD, their results can be interchangeably used in the clinic. Hindawi Publishing Corporation 2016 2016-06-14 /pmc/articles/PMC4923600/ /pubmed/27403339 http://dx.doi.org/10.1155/2016/6347236 Text en Copyright © 2016 Jinhai Huang et al. https://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Huang, Jinhai Lu, Weicong Savini, Giacomo Chen, Hao Wang, Chengfang Yu, Xinxin Bao, Fangjun Wang, Qinmei Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth |
title | Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth |
title_full | Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth |
title_fullStr | Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth |
title_full_unstemmed | Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth |
title_short | Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth |
title_sort | comparison between a new optical biometry device and an anterior segment optical coherence tomographer for measuring central corneal thickness and anterior chamber depth |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4923600/ https://www.ncbi.nlm.nih.gov/pubmed/27403339 http://dx.doi.org/10.1155/2016/6347236 |
work_keys_str_mv | AT huangjinhai comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth AT luweicong comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth AT savinigiacomo comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth AT chenhao comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth AT wangchengfang comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth AT yuxinxin comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth AT baofangjun comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth AT wangqinmei comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth |