Differential impact of multi-focus fan beam collimation with L-mode and conventional systems on the accuracy of myocardial perfusion imaging: Quantitative evaluation using phantoms

OBJECTIVE(S): A novel IQ-SPECT™ method has become widely used in clinical studies. The present study compares the quality of myocardial perfusion images (MPI) acquired using the IQ-SPECT™ (IQ-mode), conventional (180° apart: C-mode) and L-mode (90° apart: L-mode) systems. We assessed spatial resolut...

Descripción completa

Detalles Bibliográficos
Autores principales: Onishi, Hideo, Matsutomo, Norikazu, Kangai, Yoshiharu, Saho, Tatsunori, Amijima, Hizuru
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Asia Oceania Journal of Nuclear Medicine & Biology 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4927048/
https://www.ncbi.nlm.nih.gov/pubmed/27408847
_version_ 1782440207369371648
author Onishi, Hideo
Matsutomo, Norikazu
Kangai, Yoshiharu
Saho, Tatsunori
Amijima, Hizuru
author_facet Onishi, Hideo
Matsutomo, Norikazu
Kangai, Yoshiharu
Saho, Tatsunori
Amijima, Hizuru
author_sort Onishi, Hideo
collection PubMed
description OBJECTIVE(S): A novel IQ-SPECT™ method has become widely used in clinical studies. The present study compares the quality of myocardial perfusion images (MPI) acquired using the IQ-SPECT™ (IQ-mode), conventional (180° apart: C-mode) and L-mode (90° apart: L-mode) systems. We assessed spatial resolution, image reproducibility and quantifiability using various physical phantoms. METHODS: SPECT images were acquired using a dual-headed gamma camera with C-mode, L-mode, and IQ-mode acquisition systems from line source, pai and cardiac phantoms containing solutions of (99m)Tc. The line source phantom was placed in the center of the orbit and at ± 4.0, ± 8.0, ± 12.0, ± 16.0 and ± 20.0 cm off center. We examined quantifiability using the pai phantom comprising six chambers containing 0.0, 0.016, 0.03, 0.045, 0.062, and 0.074 MBq/mL of 99m-Tc and cross-calibrating the SPECT counts. Image resolution and reproducibility were quantified as myocardial wall thickness (MWT) and %uptake using polar maps. RESULTS: The full width at half maximum (FWHM) of the IQ-mode in the center was increased by 11% as compared with C-mode, and FWHM in the periphery was increased 41% compared with FWHM at the center. Calibrated SPECT counts were essentially the same when quantified using IQ-and C-modes. IQ-SPECT images of MWT were significantly improved (P<0.001) over L-mode, and C-mode SPECT imaging with IQ-mode became increasingly inhomogeneous, both visually and quantitatively (C-mode vs. L-mode, ns; C-mode vs. IQ-mode, P<0.05). CONCLUSION: Myocardial perfusion images acquired by IQ-SPECT were comparable to those acquired by conventional and L-mode SPECT, but with significantly improved resolution and quality. Our results suggest that IQ-SPECT is the optimal technology for myocardial perfusion SPECT imaging.
format Online
Article
Text
id pubmed-4927048
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher Asia Oceania Journal of Nuclear Medicine & Biology
record_format MEDLINE/PubMed
spelling pubmed-49270482016-07-12 Differential impact of multi-focus fan beam collimation with L-mode and conventional systems on the accuracy of myocardial perfusion imaging: Quantitative evaluation using phantoms Onishi, Hideo Matsutomo, Norikazu Kangai, Yoshiharu Saho, Tatsunori Amijima, Hizuru Asia Ocean J Nucl Med Biol Original Article OBJECTIVE(S): A novel IQ-SPECT™ method has become widely used in clinical studies. The present study compares the quality of myocardial perfusion images (MPI) acquired using the IQ-SPECT™ (IQ-mode), conventional (180° apart: C-mode) and L-mode (90° apart: L-mode) systems. We assessed spatial resolution, image reproducibility and quantifiability using various physical phantoms. METHODS: SPECT images were acquired using a dual-headed gamma camera with C-mode, L-mode, and IQ-mode acquisition systems from line source, pai and cardiac phantoms containing solutions of (99m)Tc. The line source phantom was placed in the center of the orbit and at ± 4.0, ± 8.0, ± 12.0, ± 16.0 and ± 20.0 cm off center. We examined quantifiability using the pai phantom comprising six chambers containing 0.0, 0.016, 0.03, 0.045, 0.062, and 0.074 MBq/mL of 99m-Tc and cross-calibrating the SPECT counts. Image resolution and reproducibility were quantified as myocardial wall thickness (MWT) and %uptake using polar maps. RESULTS: The full width at half maximum (FWHM) of the IQ-mode in the center was increased by 11% as compared with C-mode, and FWHM in the periphery was increased 41% compared with FWHM at the center. Calibrated SPECT counts were essentially the same when quantified using IQ-and C-modes. IQ-SPECT images of MWT were significantly improved (P<0.001) over L-mode, and C-mode SPECT imaging with IQ-mode became increasingly inhomogeneous, both visually and quantitatively (C-mode vs. L-mode, ns; C-mode vs. IQ-mode, P<0.05). CONCLUSION: Myocardial perfusion images acquired by IQ-SPECT were comparable to those acquired by conventional and L-mode SPECT, but with significantly improved resolution and quality. Our results suggest that IQ-SPECT is the optimal technology for myocardial perfusion SPECT imaging. Asia Oceania Journal of Nuclear Medicine & Biology 2013 /pmc/articles/PMC4927048/ /pubmed/27408847 Text en Copyright: © mums.ac.ir http://creativecommons.org/licenses/by/3.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Onishi, Hideo
Matsutomo, Norikazu
Kangai, Yoshiharu
Saho, Tatsunori
Amijima, Hizuru
Differential impact of multi-focus fan beam collimation with L-mode and conventional systems on the accuracy of myocardial perfusion imaging: Quantitative evaluation using phantoms
title Differential impact of multi-focus fan beam collimation with L-mode and conventional systems on the accuracy of myocardial perfusion imaging: Quantitative evaluation using phantoms
title_full Differential impact of multi-focus fan beam collimation with L-mode and conventional systems on the accuracy of myocardial perfusion imaging: Quantitative evaluation using phantoms
title_fullStr Differential impact of multi-focus fan beam collimation with L-mode and conventional systems on the accuracy of myocardial perfusion imaging: Quantitative evaluation using phantoms
title_full_unstemmed Differential impact of multi-focus fan beam collimation with L-mode and conventional systems on the accuracy of myocardial perfusion imaging: Quantitative evaluation using phantoms
title_short Differential impact of multi-focus fan beam collimation with L-mode and conventional systems on the accuracy of myocardial perfusion imaging: Quantitative evaluation using phantoms
title_sort differential impact of multi-focus fan beam collimation with l-mode and conventional systems on the accuracy of myocardial perfusion imaging: quantitative evaluation using phantoms
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4927048/
https://www.ncbi.nlm.nih.gov/pubmed/27408847
work_keys_str_mv AT onishihideo differentialimpactofmultifocusfanbeamcollimationwithlmodeandconventionalsystemsontheaccuracyofmyocardialperfusionimagingquantitativeevaluationusingphantoms
AT matsutomonorikazu differentialimpactofmultifocusfanbeamcollimationwithlmodeandconventionalsystemsontheaccuracyofmyocardialperfusionimagingquantitativeevaluationusingphantoms
AT kangaiyoshiharu differentialimpactofmultifocusfanbeamcollimationwithlmodeandconventionalsystemsontheaccuracyofmyocardialperfusionimagingquantitativeevaluationusingphantoms
AT sahotatsunori differentialimpactofmultifocusfanbeamcollimationwithlmodeandconventionalsystemsontheaccuracyofmyocardialperfusionimagingquantitativeevaluationusingphantoms
AT amijimahizuru differentialimpactofmultifocusfanbeamcollimationwithlmodeandconventionalsystemsontheaccuracyofmyocardialperfusionimagingquantitativeevaluationusingphantoms