Differential impact of multi-focus fan beam collimation with L-mode and conventional systems on the accuracy of myocardial perfusion imaging: Quantitative evaluation using phantoms
OBJECTIVE(S): A novel IQ-SPECT™ method has become widely used in clinical studies. The present study compares the quality of myocardial perfusion images (MPI) acquired using the IQ-SPECT™ (IQ-mode), conventional (180° apart: C-mode) and L-mode (90° apart: L-mode) systems. We assessed spatial resolut...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Asia Oceania Journal of Nuclear Medicine & Biology
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4927048/ https://www.ncbi.nlm.nih.gov/pubmed/27408847 |
_version_ | 1782440207369371648 |
---|---|
author | Onishi, Hideo Matsutomo, Norikazu Kangai, Yoshiharu Saho, Tatsunori Amijima, Hizuru |
author_facet | Onishi, Hideo Matsutomo, Norikazu Kangai, Yoshiharu Saho, Tatsunori Amijima, Hizuru |
author_sort | Onishi, Hideo |
collection | PubMed |
description | OBJECTIVE(S): A novel IQ-SPECT™ method has become widely used in clinical studies. The present study compares the quality of myocardial perfusion images (MPI) acquired using the IQ-SPECT™ (IQ-mode), conventional (180° apart: C-mode) and L-mode (90° apart: L-mode) systems. We assessed spatial resolution, image reproducibility and quantifiability using various physical phantoms. METHODS: SPECT images were acquired using a dual-headed gamma camera with C-mode, L-mode, and IQ-mode acquisition systems from line source, pai and cardiac phantoms containing solutions of (99m)Tc. The line source phantom was placed in the center of the orbit and at ± 4.0, ± 8.0, ± 12.0, ± 16.0 and ± 20.0 cm off center. We examined quantifiability using the pai phantom comprising six chambers containing 0.0, 0.016, 0.03, 0.045, 0.062, and 0.074 MBq/mL of 99m-Tc and cross-calibrating the SPECT counts. Image resolution and reproducibility were quantified as myocardial wall thickness (MWT) and %uptake using polar maps. RESULTS: The full width at half maximum (FWHM) of the IQ-mode in the center was increased by 11% as compared with C-mode, and FWHM in the periphery was increased 41% compared with FWHM at the center. Calibrated SPECT counts were essentially the same when quantified using IQ-and C-modes. IQ-SPECT images of MWT were significantly improved (P<0.001) over L-mode, and C-mode SPECT imaging with IQ-mode became increasingly inhomogeneous, both visually and quantitatively (C-mode vs. L-mode, ns; C-mode vs. IQ-mode, P<0.05). CONCLUSION: Myocardial perfusion images acquired by IQ-SPECT were comparable to those acquired by conventional and L-mode SPECT, but with significantly improved resolution and quality. Our results suggest that IQ-SPECT is the optimal technology for myocardial perfusion SPECT imaging. |
format | Online Article Text |
id | pubmed-4927048 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2013 |
publisher | Asia Oceania Journal of Nuclear Medicine & Biology |
record_format | MEDLINE/PubMed |
spelling | pubmed-49270482016-07-12 Differential impact of multi-focus fan beam collimation with L-mode and conventional systems on the accuracy of myocardial perfusion imaging: Quantitative evaluation using phantoms Onishi, Hideo Matsutomo, Norikazu Kangai, Yoshiharu Saho, Tatsunori Amijima, Hizuru Asia Ocean J Nucl Med Biol Original Article OBJECTIVE(S): A novel IQ-SPECT™ method has become widely used in clinical studies. The present study compares the quality of myocardial perfusion images (MPI) acquired using the IQ-SPECT™ (IQ-mode), conventional (180° apart: C-mode) and L-mode (90° apart: L-mode) systems. We assessed spatial resolution, image reproducibility and quantifiability using various physical phantoms. METHODS: SPECT images were acquired using a dual-headed gamma camera with C-mode, L-mode, and IQ-mode acquisition systems from line source, pai and cardiac phantoms containing solutions of (99m)Tc. The line source phantom was placed in the center of the orbit and at ± 4.0, ± 8.0, ± 12.0, ± 16.0 and ± 20.0 cm off center. We examined quantifiability using the pai phantom comprising six chambers containing 0.0, 0.016, 0.03, 0.045, 0.062, and 0.074 MBq/mL of 99m-Tc and cross-calibrating the SPECT counts. Image resolution and reproducibility were quantified as myocardial wall thickness (MWT) and %uptake using polar maps. RESULTS: The full width at half maximum (FWHM) of the IQ-mode in the center was increased by 11% as compared with C-mode, and FWHM in the periphery was increased 41% compared with FWHM at the center. Calibrated SPECT counts were essentially the same when quantified using IQ-and C-modes. IQ-SPECT images of MWT were significantly improved (P<0.001) over L-mode, and C-mode SPECT imaging with IQ-mode became increasingly inhomogeneous, both visually and quantitatively (C-mode vs. L-mode, ns; C-mode vs. IQ-mode, P<0.05). CONCLUSION: Myocardial perfusion images acquired by IQ-SPECT were comparable to those acquired by conventional and L-mode SPECT, but with significantly improved resolution and quality. Our results suggest that IQ-SPECT is the optimal technology for myocardial perfusion SPECT imaging. Asia Oceania Journal of Nuclear Medicine & Biology 2013 /pmc/articles/PMC4927048/ /pubmed/27408847 Text en Copyright: © mums.ac.ir http://creativecommons.org/licenses/by/3.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Onishi, Hideo Matsutomo, Norikazu Kangai, Yoshiharu Saho, Tatsunori Amijima, Hizuru Differential impact of multi-focus fan beam collimation with L-mode and conventional systems on the accuracy of myocardial perfusion imaging: Quantitative evaluation using phantoms |
title | Differential impact of multi-focus fan beam collimation with L-mode and conventional systems on the accuracy of myocardial perfusion imaging: Quantitative evaluation using phantoms |
title_full | Differential impact of multi-focus fan beam collimation with L-mode and conventional systems on the accuracy of myocardial perfusion imaging: Quantitative evaluation using phantoms |
title_fullStr | Differential impact of multi-focus fan beam collimation with L-mode and conventional systems on the accuracy of myocardial perfusion imaging: Quantitative evaluation using phantoms |
title_full_unstemmed | Differential impact of multi-focus fan beam collimation with L-mode and conventional systems on the accuracy of myocardial perfusion imaging: Quantitative evaluation using phantoms |
title_short | Differential impact of multi-focus fan beam collimation with L-mode and conventional systems on the accuracy of myocardial perfusion imaging: Quantitative evaluation using phantoms |
title_sort | differential impact of multi-focus fan beam collimation with l-mode and conventional systems on the accuracy of myocardial perfusion imaging: quantitative evaluation using phantoms |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4927048/ https://www.ncbi.nlm.nih.gov/pubmed/27408847 |
work_keys_str_mv | AT onishihideo differentialimpactofmultifocusfanbeamcollimationwithlmodeandconventionalsystemsontheaccuracyofmyocardialperfusionimagingquantitativeevaluationusingphantoms AT matsutomonorikazu differentialimpactofmultifocusfanbeamcollimationwithlmodeandconventionalsystemsontheaccuracyofmyocardialperfusionimagingquantitativeevaluationusingphantoms AT kangaiyoshiharu differentialimpactofmultifocusfanbeamcollimationwithlmodeandconventionalsystemsontheaccuracyofmyocardialperfusionimagingquantitativeevaluationusingphantoms AT sahotatsunori differentialimpactofmultifocusfanbeamcollimationwithlmodeandconventionalsystemsontheaccuracyofmyocardialperfusionimagingquantitativeevaluationusingphantoms AT amijimahizuru differentialimpactofmultifocusfanbeamcollimationwithlmodeandconventionalsystemsontheaccuracyofmyocardialperfusionimagingquantitativeevaluationusingphantoms |