Navigated MRI-guided liver biopsies in a closed-bore scanner: experience in 52 patients

OBJECTIVES: To evaluate clinical effectiveness and diagnostic efficiency of a navigation device for MR-guided biopsies of focal liver lesions in a closed-bore scanner. METHODS: In 52 patients, 55 biopsies were performed. An add-on MR navigation system with optical instrument tracking was used for im...

Descripción completa

Detalles Bibliográficos
Autores principales: Moche, Michael, Heinig, Susann, Garnov, Nikita, Fuchs, Jochen, Petersen, Tim-Ole, Seider, Daniel, Brandmaier, Philipp, Kahn, Thomas, Busse, Harald
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4927599/
https://www.ncbi.nlm.nih.gov/pubmed/26563349
http://dx.doi.org/10.1007/s00330-015-4097-1
Descripción
Sumario:OBJECTIVES: To evaluate clinical effectiveness and diagnostic efficiency of a navigation device for MR-guided biopsies of focal liver lesions in a closed-bore scanner. METHODS: In 52 patients, 55 biopsies were performed. An add-on MR navigation system with optical instrument tracking was used for image guidance and biopsy device insertion outside the bore. Fast control imaging allowed visualization of the true needle position at any time. The biopsy workflow and procedure duration were recorded. Histological analysis and clinical course/outcome were used to calculate sensitivity, specificity and diagnostic accuracy. RESULTS: Fifty-four of 55 liver biopsies were performed successfully with the system. No major and four minor complications occurred. Mean tumour size was 23 ± 14 mm and the skin-to-target length ranged from 22 to 177 mm. In 39 cases, access path was double oblique. Sensitivity, specificity and diagnostic accuracy were 88 %, 100 % and 92 %, respectively. The mean procedure time was 51 ± 12 min, whereas the puncture itself lasted 16 ± 6 min. On average, four control scans were taken. CONCLUSIONS: Using this navigation device, biopsies of poorly visible and difficult accessible liver lesions could be performed safely and reliably in a closed-bore MRI scanner. The system can be easily implemented in clinical routine workflow. KEY POINTS: • Targeted liver biopsies could be reliably performed in a closed-bore MRI. • The navigation system allows for image guidance outside of the scanner bore. • Assisted MRI-guided biopsies are helpful for focal lesions with a difficult access. • Successful integration of the method in clinical workflow was shown. • Subsequent system installation in an existing MRI environment is feasible.