Cargando…
An International Comparison of Absolute Radiant Power Measurement Capabilities
We report the results of an intercomparison of monochromatic radiant power measurement capabilities recently completed by 11 national laboratories. The intercomparison radiometers, distributed in pairs, included an amplifier with six decades of precision gain and one of two types of silicon photodio...
Autor principal: | |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
[Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology
1990
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4930016/ https://www.ncbi.nlm.nih.gov/pubmed/28179788 http://dx.doi.org/10.6028/jres.095.040 |
Sumario: | We report the results of an intercomparison of monochromatic radiant power measurement capabilities recently completed by 11 national laboratories. The intercomparison radiometers, distributed in pairs, included an amplifier with six decades of precision gain and one of two types of silicon photodiode (pn or np-type construction). Eleven of the laboratories measured the absolute responsivity of the radiometers at 633 nm and nine at 488 nm. The standard deviation of the overall difference was 0.36% at both wavelengths. The agreement between the various participating laboratories and NIST was within the measurement accuracy stated by the participants. |
---|