Accuracy of currently available neonatal respiratory function monitors for neonatal resuscitation

This study aimed to test the accuracy in volume measurements of three available respiratory function monitors (RFMs) for neonatal resuscitation and the effect of changing gas conditions. The Florian, New Life Box Neo-RSD (NLB Neo-RSD) and NICO RFM were tested on accuracy with volumes of 10 and 20 mL...

Descripción completa

Detalles Bibliográficos
Autores principales: Verbeek, Charlotte, van Zanten, Henriëtte A., van Vonderen, Jeroen J., Kitchen, Marcus J, Hooper, Stuart B., te Pas, Arjan B.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4930469/
https://www.ncbi.nlm.nih.gov/pubmed/27279013
http://dx.doi.org/10.1007/s00431-016-2739-1
Descripción
Sumario:This study aimed to test the accuracy in volume measurements of three available respiratory function monitors (RFMs) for neonatal resuscitation and the effect of changing gas conditions. The Florian, New Life Box Neo-RSD (NLB Neo-RSD) and NICO RFM were tested on accuracy with volumes of 10 and 20 mL and on changes in volume measurements under changing gas conditions (oxygen level 21–100 % and from cold dry air (24 ± 2 °C) to heated humidified air (37 °C). Volume differences >10 % were considered clinically relevant. We found that the mean (SD) volume difference was clinically acceptable for all devices (10, 20 mL): Florian (+8.4 (1.2)%, +8.4 (0.5)%); NLB Neo-RSD (+5.8 (1.1)%, +4.3 (1.4)%); and NICO (−8.2 (0.9)%, −8.7 (0.8)%). Changing from cold dry to heated humidified air increased the volume difference using the Florian (cold dry air, heated humidified air (+5.2 (1.2)%, +12.2 (0.9)%) but not NLB Neo-RSD (+2.0(1.6)%, +3.4(2.8)%) and NICO (−2.3 % (0.8), +0.1 (0.6)%). Similarly, when using heated humidified air, increasing oxygen enlarged increased the volume difference using the Florian (oxygen 21 %, 100 %: +12.2(1.0)%, +19.8(1.1)%), but not NLB Neo-RSD (+0.2(1.9)%, +1.1(2.8)%) and NICO (−5.6(0.9)%, −3.7(0.9)%). Clinically relevant changes occurred when changing both gas conditions (Florian +25.7(1.7)%; NLB Neo-RSD +3.8(2.4)%; NICO −5.7(1.4)%). Conclusion: The available RFMs demonstrated clinically acceptable deviations in volume measurements, except for the Florian when changing gas conditions.