Magnetic Resonance Three-dimensional Cube Technique in the Measurement of Piglet Femoral Anteversion
BACKGROUND: The accurate measurement of the femoral anteversion (FA) angle is always a topic of much debate in the orthopedic surgery and radiology research. We aimed to explore a new FA measurement method to acquire accurate results without radiation damage using piglet model. METHODS: A total of t...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications & Media Pvt Ltd
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4931266/ https://www.ncbi.nlm.nih.gov/pubmed/27364796 http://dx.doi.org/10.4103/0366-6999.184462 |
Sumario: | BACKGROUND: The accurate measurement of the femoral anteversion (FA) angle is always a topic of much debate in the orthopedic surgery and radiology research. We aimed to explore a new FA measurement method to acquire accurate results without radiation damage using piglet model. METHODS: A total of thirty piglets were assigned to two groups based on the age. Bilateral femora were imaged with 3.0-T magnetic resonance (MR) and 64-slice computed tomography (CT) examinations on all piglets. FA was measured on MR-three-dimensional (3D) postprocessing software with a four-step method: initial validation of the femoral condylar axis, validation of the condylar plane, validation of the femoral neck axis, and line-plane angle measurement of FA. After MR and CT examinations, all piglets were sacrificed and their degree of FA was measured using their excised, dried femora. MR, CT, and dried-femur measurement results were analyzed statistically; MR and CT measurements were compared for accuracy against each other and against the gold standard dried femur measurement. RESULTS: In both groups, the mean FA value measured by MR was lower than that measured by CT. A statistically significant difference was observed between CT- and dried-femur measurements but not between MR- and dried-femur measurements. A higher correlation (0.783 vs. 0.408) and a higher consistency (0.863 vs. 0.578) with dried-femur measurement results were seen for MR measurements than CT measurements in the 1-week age group. However, in the 8-week age group, similar correlations (0.707 vs. 0.669) and consistencies (0.864 vs. 0.821) were observed. CONCLUSIONS: Noninvasive MR-3D-Cube reconstruction was able to accurately measure FA in piglets. Particularly in the 1-week age group with a larger proportion of cartilaginous structures, the correlation and consistency between MR- and dried-femur measurement results were higher than those between CT- and dried-femur measurements, suggesting that MR may be a new useful examination tool for FA-related diseases in children. |
---|