Cargando…

Cross-calibration of the Siemens mMR: easily acquired accurate PET phantom measurements, long-term stability and reproducibility

BACKGROUND: We present a quick and easy method to perform quantitatively accurate PET scans of typical water-filled PET plastic shell phantoms on the Siemens Biograph mMR PET/MR system. We perform regular cross-calibrations (Xcal) of our PET systems, including the PET/MR, using a Siemens mCT water p...

Descripción completa

Detalles Bibliográficos
Autores principales: Keller, Sune H., Jakoby, Björn, Svalling, Susanne, Kjaer, Andreas, Højgaard, Liselotte, Klausen, Thomas L.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4936986/
https://www.ncbi.nlm.nih.gov/pubmed/27387738
http://dx.doi.org/10.1186/s40658-016-0146-3
_version_ 1782441626475429888
author Keller, Sune H.
Jakoby, Björn
Svalling, Susanne
Kjaer, Andreas
Højgaard, Liselotte
Klausen, Thomas L.
author_facet Keller, Sune H.
Jakoby, Björn
Svalling, Susanne
Kjaer, Andreas
Højgaard, Liselotte
Klausen, Thomas L.
author_sort Keller, Sune H.
collection PubMed
description BACKGROUND: We present a quick and easy method to perform quantitatively accurate PET scans of typical water-filled PET plastic shell phantoms on the Siemens Biograph mMR PET/MR system. We perform regular cross-calibrations (Xcal) of our PET systems, including the PET/MR, using a Siemens mCT water phantom. LONG-TERM STABILITY: The mMR calibration stability was evaluated over a 3-year period where 54 cross-calibrations were acquired, showing that the mMR on average underestimated the concentration by 16 %, consistently due to the use of MR-based μ-maps. The mMR produced the narrowest calibration ratio range with the lowest standard deviation, implying it is the most stable of the six systems in the study over a 3-year period. MMR ACCURACY WITH PREDEFINED μ-MAPS: With the latest mMR software version, VB20P, it is possible to utilize predefined phantom μ-maps. We evaluated both the system-integrated, predefined μ-map of the long mMR water phantom and our own user-defined CT-based μ-map of the mCT water phantom, which is used for cross-calibration. For seven scans, which were reconstructed with correctly segmented μ-maps, the mMR produced cross-calibration ratios of 1.00–1.02, well within the acceptance range [0.95–1.05], showing high accuracy. CONCLUSIONS: The mMR is the most stable PET system in this study, and the mean underestimation is no longer an issue with the easily accessible μ-map, which resulted in correct cross-calibration ratios in all seven tests. We will share the user-defined μ-map of the mCT phantom and the protocol with interested mMR users.
format Online
Article
Text
id pubmed-4936986
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-49369862016-07-08 Cross-calibration of the Siemens mMR: easily acquired accurate PET phantom measurements, long-term stability and reproducibility Keller, Sune H. Jakoby, Björn Svalling, Susanne Kjaer, Andreas Højgaard, Liselotte Klausen, Thomas L. EJNMMI Phys Short Communication BACKGROUND: We present a quick and easy method to perform quantitatively accurate PET scans of typical water-filled PET plastic shell phantoms on the Siemens Biograph mMR PET/MR system. We perform regular cross-calibrations (Xcal) of our PET systems, including the PET/MR, using a Siemens mCT water phantom. LONG-TERM STABILITY: The mMR calibration stability was evaluated over a 3-year period where 54 cross-calibrations were acquired, showing that the mMR on average underestimated the concentration by 16 %, consistently due to the use of MR-based μ-maps. The mMR produced the narrowest calibration ratio range with the lowest standard deviation, implying it is the most stable of the six systems in the study over a 3-year period. MMR ACCURACY WITH PREDEFINED μ-MAPS: With the latest mMR software version, VB20P, it is possible to utilize predefined phantom μ-maps. We evaluated both the system-integrated, predefined μ-map of the long mMR water phantom and our own user-defined CT-based μ-map of the mCT water phantom, which is used for cross-calibration. For seven scans, which were reconstructed with correctly segmented μ-maps, the mMR produced cross-calibration ratios of 1.00–1.02, well within the acceptance range [0.95–1.05], showing high accuracy. CONCLUSIONS: The mMR is the most stable PET system in this study, and the mean underestimation is no longer an issue with the easily accessible μ-map, which resulted in correct cross-calibration ratios in all seven tests. We will share the user-defined μ-map of the mCT phantom and the protocol with interested mMR users. Springer International Publishing 2016-07-07 /pmc/articles/PMC4936986/ /pubmed/27387738 http://dx.doi.org/10.1186/s40658-016-0146-3 Text en © The Author(s). 2016 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
spellingShingle Short Communication
Keller, Sune H.
Jakoby, Björn
Svalling, Susanne
Kjaer, Andreas
Højgaard, Liselotte
Klausen, Thomas L.
Cross-calibration of the Siemens mMR: easily acquired accurate PET phantom measurements, long-term stability and reproducibility
title Cross-calibration of the Siemens mMR: easily acquired accurate PET phantom measurements, long-term stability and reproducibility
title_full Cross-calibration of the Siemens mMR: easily acquired accurate PET phantom measurements, long-term stability and reproducibility
title_fullStr Cross-calibration of the Siemens mMR: easily acquired accurate PET phantom measurements, long-term stability and reproducibility
title_full_unstemmed Cross-calibration of the Siemens mMR: easily acquired accurate PET phantom measurements, long-term stability and reproducibility
title_short Cross-calibration of the Siemens mMR: easily acquired accurate PET phantom measurements, long-term stability and reproducibility
title_sort cross-calibration of the siemens mmr: easily acquired accurate pet phantom measurements, long-term stability and reproducibility
topic Short Communication
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4936986/
https://www.ncbi.nlm.nih.gov/pubmed/27387738
http://dx.doi.org/10.1186/s40658-016-0146-3
work_keys_str_mv AT kellersuneh crosscalibrationofthesiemensmmreasilyacquiredaccuratepetphantommeasurementslongtermstabilityandreproducibility
AT jakobybjorn crosscalibrationofthesiemensmmreasilyacquiredaccuratepetphantommeasurementslongtermstabilityandreproducibility
AT svallingsusanne crosscalibrationofthesiemensmmreasilyacquiredaccuratepetphantommeasurementslongtermstabilityandreproducibility
AT kjaerandreas crosscalibrationofthesiemensmmreasilyacquiredaccuratepetphantommeasurementslongtermstabilityandreproducibility
AT højgaardliselotte crosscalibrationofthesiemensmmreasilyacquiredaccuratepetphantommeasurementslongtermstabilityandreproducibility
AT klausenthomasl crosscalibrationofthesiemensmmreasilyacquiredaccuratepetphantommeasurementslongtermstabilityandreproducibility