Double-low protocol for hepatic dynamic CT scan: Effect of low tube voltage and low-dose iodine contrast agent on image quality
The radiation-induced carcinogenesis from computed tomography (CT) and iodine contrast agent induced nephropathy has attracted international attention. The reduction of the radiation dose and iodine intake in CT scan is always a direction for researchers to strive. The purpose of this study was to e...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer Health
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4937926/ https://www.ncbi.nlm.nih.gov/pubmed/27368012 http://dx.doi.org/10.1097/MD.0000000000004004 |
_version_ | 1782441793852276736 |
---|---|
author | Zhang, Xiuli Li, Shaodong Liu, Wenlou Huang, Ning Li, Jingjing Cheng, Li Xu, Kai |
author_facet | Zhang, Xiuli Li, Shaodong Liu, Wenlou Huang, Ning Li, Jingjing Cheng, Li Xu, Kai |
author_sort | Zhang, Xiuli |
collection | PubMed |
description | The radiation-induced carcinogenesis from computed tomography (CT) and iodine contrast agent induced nephropathy has attracted international attention. The reduction of the radiation dose and iodine intake in CT scan is always a direction for researchers to strive. The purpose of this study was to evaluate the feasibility of a “double-low” (i.e., low tube voltage and low-dose iodine contrast agent) scanning protocol for dynamic hepatic CT with the adaptive statistical iterative reconstruction (ASIR) in patients with a body mass index (BMI) of 18.5 to 27.9 kg/m(2). A total of 128 consecutive patients with a BMI between 18.5 and 27.9 kg/m(2) were randomly assigned into 3 groups according to tube voltage, iodine contrast agent, and reconstruction algorithms. Group A (the “double-low” protocol): 100 kVp tube voltage with 40% ASIR, iodixanol at 270 mg I/mL, group B: 120 kVp tube voltage with filtered back projection (FBP), iodixanol at 270 mg I/ mL, and group C: 120 kVp tube voltage with FBP, ioversol at 350 mg I/ mL. The volume CT dose index (CTDI(vol)) and effective dose (ED) in group A were lower than those in group B and C (all P < 0.01). The iodine intake in group A was decreased by approximately 26.5% than group C, whereas no statistical difference was observed between group A and B (P > 0.05). There was no significant difference of the CT values between group A and C (P > 0.05), which both showed higher CT values than that in group B (P < 0.001). However, no statistic difference was observed in the contrast-to-noise ratio (CNR), the signal-to-noise ratio (SNR), and image-quality scores among the 3 groups (all P > 0.05). Near-perfect consistency of the evaluation for group A, B, and C (Kenall's W = 0.921, 0.874, and 0.949, respectively) was obtained by the 4 readers with respect to the overall image quality. These results suggested that the “double-low” protocol with ASIR algorithm for multi-phase hepatic CT scan can dramatically decrease radiation dose and iodine intake with adequate image quality in patients with BMI of 18.5 to 27.9 kg/m(2). |
format | Online Article Text |
id | pubmed-4937926 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | Wolters Kluwer Health |
record_format | MEDLINE/PubMed |
spelling | pubmed-49379262016-08-18 Double-low protocol for hepatic dynamic CT scan: Effect of low tube voltage and low-dose iodine contrast agent on image quality Zhang, Xiuli Li, Shaodong Liu, Wenlou Huang, Ning Li, Jingjing Cheng, Li Xu, Kai Medicine (Baltimore) 6800 The radiation-induced carcinogenesis from computed tomography (CT) and iodine contrast agent induced nephropathy has attracted international attention. The reduction of the radiation dose and iodine intake in CT scan is always a direction for researchers to strive. The purpose of this study was to evaluate the feasibility of a “double-low” (i.e., low tube voltage and low-dose iodine contrast agent) scanning protocol for dynamic hepatic CT with the adaptive statistical iterative reconstruction (ASIR) in patients with a body mass index (BMI) of 18.5 to 27.9 kg/m(2). A total of 128 consecutive patients with a BMI between 18.5 and 27.9 kg/m(2) were randomly assigned into 3 groups according to tube voltage, iodine contrast agent, and reconstruction algorithms. Group A (the “double-low” protocol): 100 kVp tube voltage with 40% ASIR, iodixanol at 270 mg I/mL, group B: 120 kVp tube voltage with filtered back projection (FBP), iodixanol at 270 mg I/ mL, and group C: 120 kVp tube voltage with FBP, ioversol at 350 mg I/ mL. The volume CT dose index (CTDI(vol)) and effective dose (ED) in group A were lower than those in group B and C (all P < 0.01). The iodine intake in group A was decreased by approximately 26.5% than group C, whereas no statistical difference was observed between group A and B (P > 0.05). There was no significant difference of the CT values between group A and C (P > 0.05), which both showed higher CT values than that in group B (P < 0.001). However, no statistic difference was observed in the contrast-to-noise ratio (CNR), the signal-to-noise ratio (SNR), and image-quality scores among the 3 groups (all P > 0.05). Near-perfect consistency of the evaluation for group A, B, and C (Kenall's W = 0.921, 0.874, and 0.949, respectively) was obtained by the 4 readers with respect to the overall image quality. These results suggested that the “double-low” protocol with ASIR algorithm for multi-phase hepatic CT scan can dramatically decrease radiation dose and iodine intake with adequate image quality in patients with BMI of 18.5 to 27.9 kg/m(2). Wolters Kluwer Health 2016-07-01 /pmc/articles/PMC4937926/ /pubmed/27368012 http://dx.doi.org/10.1097/MD.0000000000004004 Text en Copyright © 2016 Wolters Kluwer Health, Inc. All rights reserved. http://creativecommons.org/licenses/by-nd/4.0 This is an open access article distributed under the Creative Commons Attribution-NoDerivatives License 4.0, which allows for redistribution, commercial and non-commercial, as long as it is passed along unchanged and in whole, with credit to the author. http://creativecommons.org/licenses/by-nd/4.0 |
spellingShingle | 6800 Zhang, Xiuli Li, Shaodong Liu, Wenlou Huang, Ning Li, Jingjing Cheng, Li Xu, Kai Double-low protocol for hepatic dynamic CT scan: Effect of low tube voltage and low-dose iodine contrast agent on image quality |
title | Double-low protocol for hepatic dynamic CT scan: Effect of low tube voltage and low-dose iodine contrast agent on image quality |
title_full | Double-low protocol for hepatic dynamic CT scan: Effect of low tube voltage and low-dose iodine contrast agent on image quality |
title_fullStr | Double-low protocol for hepatic dynamic CT scan: Effect of low tube voltage and low-dose iodine contrast agent on image quality |
title_full_unstemmed | Double-low protocol for hepatic dynamic CT scan: Effect of low tube voltage and low-dose iodine contrast agent on image quality |
title_short | Double-low protocol for hepatic dynamic CT scan: Effect of low tube voltage and low-dose iodine contrast agent on image quality |
title_sort | double-low protocol for hepatic dynamic ct scan: effect of low tube voltage and low-dose iodine contrast agent on image quality |
topic | 6800 |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4937926/ https://www.ncbi.nlm.nih.gov/pubmed/27368012 http://dx.doi.org/10.1097/MD.0000000000004004 |
work_keys_str_mv | AT zhangxiuli doublelowprotocolforhepaticdynamicctscaneffectoflowtubevoltageandlowdoseiodinecontrastagentonimagequality AT lishaodong doublelowprotocolforhepaticdynamicctscaneffectoflowtubevoltageandlowdoseiodinecontrastagentonimagequality AT liuwenlou doublelowprotocolforhepaticdynamicctscaneffectoflowtubevoltageandlowdoseiodinecontrastagentonimagequality AT huangning doublelowprotocolforhepaticdynamicctscaneffectoflowtubevoltageandlowdoseiodinecontrastagentonimagequality AT lijingjing doublelowprotocolforhepaticdynamicctscaneffectoflowtubevoltageandlowdoseiodinecontrastagentonimagequality AT chengli doublelowprotocolforhepaticdynamicctscaneffectoflowtubevoltageandlowdoseiodinecontrastagentonimagequality AT xukai doublelowprotocolforhepaticdynamicctscaneffectoflowtubevoltageandlowdoseiodinecontrastagentonimagequality |