Energy Dependence of Measured CT Numbers on Substituted Materials Used for CT Number Calibration of Radiotherapy Treatment Planning Systems

INTRODUCTION: For accurate dose calculations, it is necessary to provide a correct relationship between the CT numbers and electron density in radiotherapy treatment planning systems (TPSs). The purpose of this study was to investigate the energy dependence of measured CT numbers on substituted mate...

Descripción completa

Detalles Bibliográficos
Autores principales: Mahmoudi, Reza, Jabbari, Nasrollah, aghdasi, Mehdi, Khalkhali, Hamid Reza
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4938553/
https://www.ncbi.nlm.nih.gov/pubmed/27391672
http://dx.doi.org/10.1371/journal.pone.0158828
_version_ 1782441878143107072
author Mahmoudi, Reza
Jabbari, Nasrollah
aghdasi, Mehdi
Khalkhali, Hamid Reza
author_facet Mahmoudi, Reza
Jabbari, Nasrollah
aghdasi, Mehdi
Khalkhali, Hamid Reza
author_sort Mahmoudi, Reza
collection PubMed
description INTRODUCTION: For accurate dose calculations, it is necessary to provide a correct relationship between the CT numbers and electron density in radiotherapy treatment planning systems (TPSs). The purpose of this study was to investigate the energy dependence of measured CT numbers on substituted materials used for CT number calibration of radiotherapy TPSs and the resulting errors in the treatment planning calculation doses. MATERIALS AND METHODS: In this study, we designed a cylindrical water phantom with different materials used as tissue equivalent materials for the simulation of tissues and obtaining the related CT numbers. For evaluating the effect of CT number variations of substituted materials due to energy changing of scanner (kVp) on the dose calculation of TPS, the slices of the scanned phantom at three kVp's were imported into the desired TPSs (MIRS and CorePLAN). Dose calculations were performed on two TPSs. RESULTS: The mean absolute percentage differences between the CT numbers of CT scanner and two treatment planning systems for all the samples were 3.22%±2.57% for CorePLAN and 2.88%±2.11% for MIRS. It was also found that the maximum absolute percentage difference between all of the calculated doses from each photon beam of linac (6 and 15 MV) at three kVp's was less than 1.2%. DISCUSSION: The present study revealed that, for the materials with effective low atomic number, the mean CT number increased with increasing energy, which was opposite for the materials with an effective high atomic number. We concluded that the tissue substitute materials had a different behavior in the energy ranges from 80 to 130 kVp. So, it is necessary to consider the energy dependence of the substitute materials used for the measurement or calibration of CT number for radiotherapy treatment planning systems.
format Online
Article
Text
id pubmed-4938553
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-49385532016-07-22 Energy Dependence of Measured CT Numbers on Substituted Materials Used for CT Number Calibration of Radiotherapy Treatment Planning Systems Mahmoudi, Reza Jabbari, Nasrollah aghdasi, Mehdi Khalkhali, Hamid Reza PLoS One Research Article INTRODUCTION: For accurate dose calculations, it is necessary to provide a correct relationship between the CT numbers and electron density in radiotherapy treatment planning systems (TPSs). The purpose of this study was to investigate the energy dependence of measured CT numbers on substituted materials used for CT number calibration of radiotherapy TPSs and the resulting errors in the treatment planning calculation doses. MATERIALS AND METHODS: In this study, we designed a cylindrical water phantom with different materials used as tissue equivalent materials for the simulation of tissues and obtaining the related CT numbers. For evaluating the effect of CT number variations of substituted materials due to energy changing of scanner (kVp) on the dose calculation of TPS, the slices of the scanned phantom at three kVp's were imported into the desired TPSs (MIRS and CorePLAN). Dose calculations were performed on two TPSs. RESULTS: The mean absolute percentage differences between the CT numbers of CT scanner and two treatment planning systems for all the samples were 3.22%±2.57% for CorePLAN and 2.88%±2.11% for MIRS. It was also found that the maximum absolute percentage difference between all of the calculated doses from each photon beam of linac (6 and 15 MV) at three kVp's was less than 1.2%. DISCUSSION: The present study revealed that, for the materials with effective low atomic number, the mean CT number increased with increasing energy, which was opposite for the materials with an effective high atomic number. We concluded that the tissue substitute materials had a different behavior in the energy ranges from 80 to 130 kVp. So, it is necessary to consider the energy dependence of the substitute materials used for the measurement or calibration of CT number for radiotherapy treatment planning systems. Public Library of Science 2016-07-08 /pmc/articles/PMC4938553/ /pubmed/27391672 http://dx.doi.org/10.1371/journal.pone.0158828 Text en © 2016 Mahmoudi et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Mahmoudi, Reza
Jabbari, Nasrollah
aghdasi, Mehdi
Khalkhali, Hamid Reza
Energy Dependence of Measured CT Numbers on Substituted Materials Used for CT Number Calibration of Radiotherapy Treatment Planning Systems
title Energy Dependence of Measured CT Numbers on Substituted Materials Used for CT Number Calibration of Radiotherapy Treatment Planning Systems
title_full Energy Dependence of Measured CT Numbers on Substituted Materials Used for CT Number Calibration of Radiotherapy Treatment Planning Systems
title_fullStr Energy Dependence of Measured CT Numbers on Substituted Materials Used for CT Number Calibration of Radiotherapy Treatment Planning Systems
title_full_unstemmed Energy Dependence of Measured CT Numbers on Substituted Materials Used for CT Number Calibration of Radiotherapy Treatment Planning Systems
title_short Energy Dependence of Measured CT Numbers on Substituted Materials Used for CT Number Calibration of Radiotherapy Treatment Planning Systems
title_sort energy dependence of measured ct numbers on substituted materials used for ct number calibration of radiotherapy treatment planning systems
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4938553/
https://www.ncbi.nlm.nih.gov/pubmed/27391672
http://dx.doi.org/10.1371/journal.pone.0158828
work_keys_str_mv AT mahmoudireza energydependenceofmeasuredctnumbersonsubstitutedmaterialsusedforctnumbercalibrationofradiotherapytreatmentplanningsystems
AT jabbarinasrollah energydependenceofmeasuredctnumbersonsubstitutedmaterialsusedforctnumbercalibrationofradiotherapytreatmentplanningsystems
AT aghdasimehdi energydependenceofmeasuredctnumbersonsubstitutedmaterialsusedforctnumbercalibrationofradiotherapytreatmentplanningsystems
AT khalkhalihamidreza energydependenceofmeasuredctnumbersonsubstitutedmaterialsusedforctnumbercalibrationofradiotherapytreatmentplanningsystems