Cargando…

Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device

Failure of a vascular closure device most commonly results in a hemorrhage or pseudoaneurysm formation. In this paper a rare case of severe acute limb ischemia following incorrect deployment of a clip-based closure device (Starclose SE, Abbott Vascular) in a 31-year-old woman is presented. Symptoms...

Descripción completa

Detalles Bibliográficos
Autores principales: Dzieciuchowicz, Łukasz, Zmysłowski, Maciej, Stefaniak, Karolina, Oszkinis, Grzegorz
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Termedia Publishing House 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4945604/
https://www.ncbi.nlm.nih.gov/pubmed/27458492
http://dx.doi.org/10.5114/wiitm.2016.59577
Descripción
Sumario:Failure of a vascular closure device most commonly results in a hemorrhage or pseudoaneurysm formation. In this paper a rare case of severe acute limb ischemia following incorrect deployment of a clip-based closure device (Starclose SE, Abbott Vascular) in a 31-year-old woman is presented. Symptoms of acute limb ischemia occurred at the start of the ambulation, 6 h after completion of the procedure. Because of the severity of ischemia the patient was treated surgically, and limb perfusion was successfully restored. An attempt of closure of an inadvertently punctured narrow superficial femoral artery was identified as the cause of this complication.