Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device

Failure of a vascular closure device most commonly results in a hemorrhage or pseudoaneurysm formation. In this paper a rare case of severe acute limb ischemia following incorrect deployment of a clip-based closure device (Starclose SE, Abbott Vascular) in a 31-year-old woman is presented. Symptoms...

Descripción completa

Detalles Bibliográficos
Autores principales: Dzieciuchowicz, Łukasz, Zmysłowski, Maciej, Stefaniak, Karolina, Oszkinis, Grzegorz
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Termedia Publishing House 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4945604/
https://www.ncbi.nlm.nih.gov/pubmed/27458492
http://dx.doi.org/10.5114/wiitm.2016.59577
_version_ 1782442893248561152
author Dzieciuchowicz, Łukasz
Zmysłowski, Maciej
Stefaniak, Karolina
Oszkinis, Grzegorz
author_facet Dzieciuchowicz, Łukasz
Zmysłowski, Maciej
Stefaniak, Karolina
Oszkinis, Grzegorz
author_sort Dzieciuchowicz, Łukasz
collection PubMed
description Failure of a vascular closure device most commonly results in a hemorrhage or pseudoaneurysm formation. In this paper a rare case of severe acute limb ischemia following incorrect deployment of a clip-based closure device (Starclose SE, Abbott Vascular) in a 31-year-old woman is presented. Symptoms of acute limb ischemia occurred at the start of the ambulation, 6 h after completion of the procedure. Because of the severity of ischemia the patient was treated surgically, and limb perfusion was successfully restored. An attempt of closure of an inadvertently punctured narrow superficial femoral artery was identified as the cause of this complication.
format Online
Article
Text
id pubmed-4945604
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Termedia Publishing House
record_format MEDLINE/PubMed
spelling pubmed-49456042016-07-25 Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device Dzieciuchowicz, Łukasz Zmysłowski, Maciej Stefaniak, Karolina Oszkinis, Grzegorz Wideochir Inne Tech Maloinwazyjne Case Report Failure of a vascular closure device most commonly results in a hemorrhage or pseudoaneurysm formation. In this paper a rare case of severe acute limb ischemia following incorrect deployment of a clip-based closure device (Starclose SE, Abbott Vascular) in a 31-year-old woman is presented. Symptoms of acute limb ischemia occurred at the start of the ambulation, 6 h after completion of the procedure. Because of the severity of ischemia the patient was treated surgically, and limb perfusion was successfully restored. An attempt of closure of an inadvertently punctured narrow superficial femoral artery was identified as the cause of this complication. Termedia Publishing House 2016-04-27 2016-06 /pmc/articles/PMC4945604/ /pubmed/27458492 http://dx.doi.org/10.5114/wiitm.2016.59577 Text en Copyright © 2016 Fundacja Videochirurgii http://creativecommons.org/licenses/by-nc-sa/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License, allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.
spellingShingle Case Report
Dzieciuchowicz, Łukasz
Zmysłowski, Maciej
Stefaniak, Karolina
Oszkinis, Grzegorz
Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device
title Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device
title_full Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device
title_fullStr Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device
title_full_unstemmed Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device
title_short Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device
title_sort acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4945604/
https://www.ncbi.nlm.nih.gov/pubmed/27458492
http://dx.doi.org/10.5114/wiitm.2016.59577
work_keys_str_mv AT dzieciuchowiczłukasz acutelimbischemiacausedbyincorrectdeploymentofaclipbasedarterialclosuredevice
AT zmysłowskimaciej acutelimbischemiacausedbyincorrectdeploymentofaclipbasedarterialclosuredevice
AT stefaniakkarolina acutelimbischemiacausedbyincorrectdeploymentofaclipbasedarterialclosuredevice
AT oszkinisgrzegorz acutelimbischemiacausedbyincorrectdeploymentofaclipbasedarterialclosuredevice