Cargando…
Acute limb ischemia caused by incorrect deployment of a clip-based arterial closure device
Failure of a vascular closure device most commonly results in a hemorrhage or pseudoaneurysm formation. In this paper a rare case of severe acute limb ischemia following incorrect deployment of a clip-based closure device (Starclose SE, Abbott Vascular) in a 31-year-old woman is presented. Symptoms...
Autores principales: | Dzieciuchowicz, Łukasz, Zmysłowski, Maciej, Stefaniak, Karolina, Oszkinis, Grzegorz |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Termedia Publishing House
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4945604/ https://www.ncbi.nlm.nih.gov/pubmed/27458492 http://dx.doi.org/10.5114/wiitm.2016.59577 |