Thoracic endovascular aortic repair of a severely angulated aorta using a double-wire technique
When endovascular treatment is performed, angulation of the access route for a device can make the operative procedure difficult. We encountered a case in which we successfully completed thoracic endovascular aortic repair (TEVAR) in a patient with severely angulated aorta by applying ‘double-wire t...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Oxford University Press
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4946623/ https://www.ncbi.nlm.nih.gov/pubmed/27421301 http://dx.doi.org/10.1093/jscr/rjw125 |
Sumario: | When endovascular treatment is performed, angulation of the access route for a device can make the operative procedure difficult. We encountered a case in which we successfully completed thoracic endovascular aortic repair (TEVAR) in a patient with severely angulated aorta by applying ‘double-wire technique’. The patient was an 80-year-old woman. An aneurysm with a 71-mm diameter was observed in the descending aorta. We performed TEVAR. Device delivery could not be achieved by a conventional procedure using one guide wire since the peripheral aorta was severely angulated. Therefore, in addition to a guide wire for main body, a stiff wire and a stiff sheath were introduced to straighten the angulation. The device was successfully introduced and TEVAR was completed. We used the Relay Plus(®) that facilitates tracking through the angulation. The device has a dual structure consisting of a hard sheath and a flexible sheath. We performed TEVAR successfully. |
---|