Failure of Emperion modular femoral stem with implant analysis

Modularity in total hip arthroplasty provides multiple benefits to the surgeon in restoring the appropriate alignment and position to a previously damaged hip joint. The vast majority of modern implants incorporate modularity into their design with some implants having multiple modular interfaces. T...

Descripción completa

Detalles Bibliográficos
Autores principales: Stronach, Benjamin M., Roach, Michael D., St. John, Kenneth R.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4957168/
https://www.ncbi.nlm.nih.gov/pubmed/28326390
http://dx.doi.org/10.1016/j.artd.2015.11.004
_version_ 1782444135354990592
author Stronach, Benjamin M.
Roach, Michael D.
St. John, Kenneth R.
author_facet Stronach, Benjamin M.
Roach, Michael D.
St. John, Kenneth R.
author_sort Stronach, Benjamin M.
collection PubMed
description Modularity in total hip arthroplasty provides multiple benefits to the surgeon in restoring the appropriate alignment and position to a previously damaged hip joint. The vast majority of modern implants incorporate modularity into their design with some implants having multiple modular interfaces. There is the potential for failure at modular junctions because of fretting and crevice corrosion in combination with mechanical loading. This case report details the failure of an Emperion (Smith and Nephew, Memphis, TN) femoral stem in a 67-year-old male patient 6 years after total hip replacement. Analysis of the implant revealed mechanically assisted crevice corrosion that likely accelerated fatigue crack initiation in the hip stem. The benefits of modularity come with the potential drawback of a combination of fretting and crevice corrosion at the modular junction, which may accelerate fatigue, crack initiation and ultimately reduce the hip longevity.
format Online
Article
Text
id pubmed-4957168
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-49571682017-03-21 Failure of Emperion modular femoral stem with implant analysis Stronach, Benjamin M. Roach, Michael D. St. John, Kenneth R. Arthroplasty Today Case Report Modularity in total hip arthroplasty provides multiple benefits to the surgeon in restoring the appropriate alignment and position to a previously damaged hip joint. The vast majority of modern implants incorporate modularity into their design with some implants having multiple modular interfaces. There is the potential for failure at modular junctions because of fretting and crevice corrosion in combination with mechanical loading. This case report details the failure of an Emperion (Smith and Nephew, Memphis, TN) femoral stem in a 67-year-old male patient 6 years after total hip replacement. Analysis of the implant revealed mechanically assisted crevice corrosion that likely accelerated fatigue crack initiation in the hip stem. The benefits of modularity come with the potential drawback of a combination of fretting and crevice corrosion at the modular junction, which may accelerate fatigue, crack initiation and ultimately reduce the hip longevity. Elsevier 2016-01-11 /pmc/articles/PMC4957168/ /pubmed/28326390 http://dx.doi.org/10.1016/j.artd.2015.11.004 Text en © 2016 The Authors http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Case Report
Stronach, Benjamin M.
Roach, Michael D.
St. John, Kenneth R.
Failure of Emperion modular femoral stem with implant analysis
title Failure of Emperion modular femoral stem with implant analysis
title_full Failure of Emperion modular femoral stem with implant analysis
title_fullStr Failure of Emperion modular femoral stem with implant analysis
title_full_unstemmed Failure of Emperion modular femoral stem with implant analysis
title_short Failure of Emperion modular femoral stem with implant analysis
title_sort failure of emperion modular femoral stem with implant analysis
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4957168/
https://www.ncbi.nlm.nih.gov/pubmed/28326390
http://dx.doi.org/10.1016/j.artd.2015.11.004
work_keys_str_mv AT stronachbenjaminm failureofemperionmodularfemoralstemwithimplantanalysis
AT roachmichaeld failureofemperionmodularfemoralstemwithimplantanalysis
AT stjohnkennethr failureofemperionmodularfemoralstemwithimplantanalysis