Cargando…

Custom titanium sleeve for surgical treatment of mechanically assisted crevice corrosion in the well-fixed, noncontemporary stem in total hip arthroplasty

Adverse local tissue reaction associated with total hip replacement may occur when mechanically assisted crevice corrosion occurs at metal-metal modular junctions in which at least one of the components is fabricated from cobalt-chromium alloy. Complete removal of components may be associated with s...

Descripción completa

Detalles Bibliográficos
Autores principales: Leibiger, Thomas, McGrory, Brian J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4958117/
https://www.ncbi.nlm.nih.gov/pubmed/28326384
http://dx.doi.org/10.1016/j.artd.2015.10.001
Descripción
Sumario:Adverse local tissue reaction associated with total hip replacement may occur when mechanically assisted crevice corrosion occurs at metal-metal modular junctions in which at least one of the components is fabricated from cobalt-chromium alloy. Complete removal of components may be associated with significant morbidity; when components are well fixed and in acceptable position, it may be appropriate to consider modular rather than complete revision. We have diagnosed mechanically assisted crevice corrosion in total hip arthroplasty patients with noncontemporary but well-fixed femoral components and found that modular conversion to a ceramic femoral head to remove a source of CoCr corrosion and fretting products was only possible by having a custom titanium sleeve manufactured. Surgical implantation with a revision style Biolox ceramic head (CeramTec, Plochingen, Germany) was then achieved.