Cargando…

Custom titanium sleeve for surgical treatment of mechanically assisted crevice corrosion in the well-fixed, noncontemporary stem in total hip arthroplasty

Adverse local tissue reaction associated with total hip replacement may occur when mechanically assisted crevice corrosion occurs at metal-metal modular junctions in which at least one of the components is fabricated from cobalt-chromium alloy. Complete removal of components may be associated with s...

Descripción completa

Detalles Bibliográficos
Autores principales: Leibiger, Thomas, McGrory, Brian J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4958117/
https://www.ncbi.nlm.nih.gov/pubmed/28326384
http://dx.doi.org/10.1016/j.artd.2015.10.001
_version_ 1782444271044919296
author Leibiger, Thomas
McGrory, Brian J.
author_facet Leibiger, Thomas
McGrory, Brian J.
author_sort Leibiger, Thomas
collection PubMed
description Adverse local tissue reaction associated with total hip replacement may occur when mechanically assisted crevice corrosion occurs at metal-metal modular junctions in which at least one of the components is fabricated from cobalt-chromium alloy. Complete removal of components may be associated with significant morbidity; when components are well fixed and in acceptable position, it may be appropriate to consider modular rather than complete revision. We have diagnosed mechanically assisted crevice corrosion in total hip arthroplasty patients with noncontemporary but well-fixed femoral components and found that modular conversion to a ceramic femoral head to remove a source of CoCr corrosion and fretting products was only possible by having a custom titanium sleeve manufactured. Surgical implantation with a revision style Biolox ceramic head (CeramTec, Plochingen, Germany) was then achieved.
format Online
Article
Text
id pubmed-4958117
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-49581172017-03-21 Custom titanium sleeve for surgical treatment of mechanically assisted crevice corrosion in the well-fixed, noncontemporary stem in total hip arthroplasty Leibiger, Thomas McGrory, Brian J. Arthroplasty Today Surgical Technique Adverse local tissue reaction associated with total hip replacement may occur when mechanically assisted crevice corrosion occurs at metal-metal modular junctions in which at least one of the components is fabricated from cobalt-chromium alloy. Complete removal of components may be associated with significant morbidity; when components are well fixed and in acceptable position, it may be appropriate to consider modular rather than complete revision. We have diagnosed mechanically assisted crevice corrosion in total hip arthroplasty patients with noncontemporary but well-fixed femoral components and found that modular conversion to a ceramic femoral head to remove a source of CoCr corrosion and fretting products was only possible by having a custom titanium sleeve manufactured. Surgical implantation with a revision style Biolox ceramic head (CeramTec, Plochingen, Germany) was then achieved. Elsevier 2015-10-31 /pmc/articles/PMC4958117/ /pubmed/28326384 http://dx.doi.org/10.1016/j.artd.2015.10.001 Text en © 2015 The Authors http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Surgical Technique
Leibiger, Thomas
McGrory, Brian J.
Custom titanium sleeve for surgical treatment of mechanically assisted crevice corrosion in the well-fixed, noncontemporary stem in total hip arthroplasty
title Custom titanium sleeve for surgical treatment of mechanically assisted crevice corrosion in the well-fixed, noncontemporary stem in total hip arthroplasty
title_full Custom titanium sleeve for surgical treatment of mechanically assisted crevice corrosion in the well-fixed, noncontemporary stem in total hip arthroplasty
title_fullStr Custom titanium sleeve for surgical treatment of mechanically assisted crevice corrosion in the well-fixed, noncontemporary stem in total hip arthroplasty
title_full_unstemmed Custom titanium sleeve for surgical treatment of mechanically assisted crevice corrosion in the well-fixed, noncontemporary stem in total hip arthroplasty
title_short Custom titanium sleeve for surgical treatment of mechanically assisted crevice corrosion in the well-fixed, noncontemporary stem in total hip arthroplasty
title_sort custom titanium sleeve for surgical treatment of mechanically assisted crevice corrosion in the well-fixed, noncontemporary stem in total hip arthroplasty
topic Surgical Technique
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4958117/
https://www.ncbi.nlm.nih.gov/pubmed/28326384
http://dx.doi.org/10.1016/j.artd.2015.10.001
work_keys_str_mv AT leibigerthomas customtitaniumsleeveforsurgicaltreatmentofmechanicallyassistedcrevicecorrosioninthewellfixednoncontemporarystemintotalhiparthroplasty
AT mcgrorybrianj customtitaniumsleeveforsurgicaltreatmentofmechanicallyassistedcrevicecorrosioninthewellfixednoncontemporarystemintotalhiparthroplasty