Comparison of the Multiple Reaction Monitoring and Enhanced Product Ion Scan Modes for Confirmation of Stilbenes in Bovine Urine Samples Using LC–MS/MS QTRAP(®) System

In accordance with Commission Decision 2002/657/EC, confirmatory methods for the detection of prohibited substances should comply with specific requirements, including the criteria for confirmation. Two strategies: multiple reaction monitoring (MRM) and enhanced product ion (EPI) scanning functions...

Descripción completa

Detalles Bibliográficos
Autores principales: Matraszek-Zuchowska, Iwona, Wozniak, Barbara, Posyniak, Andrzej
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4961723/
https://www.ncbi.nlm.nih.gov/pubmed/27512157
http://dx.doi.org/10.1007/s10337-016-3121-1
_version_ 1782444699293843456
author Matraszek-Zuchowska, Iwona
Wozniak, Barbara
Posyniak, Andrzej
author_facet Matraszek-Zuchowska, Iwona
Wozniak, Barbara
Posyniak, Andrzej
author_sort Matraszek-Zuchowska, Iwona
collection PubMed
description In accordance with Commission Decision 2002/657/EC, confirmatory methods for the detection of prohibited substances should comply with specific requirements, including the criteria for confirmation. Two strategies: multiple reaction monitoring (MRM) and enhanced product ion (EPI) scanning functions were compared for confirming the anabolic compounds from synthetic stilbenes group in bovine urine samples. In the research, twenty samples fortified at the Recommended Concentration (RC) of 1 µg L(−1) with diethylstilbestrol, dienestrol and hexestrol were analyzed by liquid chromatography-tandem mass spectrometry on a QTRAP 5500 instrument. The analytical procedure, validated in accordance with the Commission Decision 2002/657/EC, used in the official control of hormones in Poland was applied. The validation parameters were in agreement with 2002/657/EC performance criteria. The effectiveness of MRM and EPI scanning modes for confirmation purposes was evaluated based on the percentage of the results confirmed. In all urine samples recorded in the MRM mode, the confirmation criteria (retention time, relative intensities between transitions) have been fulfilled. The presence of stilbenes in all urine samples using EPI scan mode was confirmed too as evidenced by a good matching of stilbenes spectra in the samples to the reference spectra with critical match factor above 0.7. The results of the research show that EPI scanning function provides the same effectiveness for confirmation of banned compounds as the mostly used MRM scan mode and can be an additional tool to confirm the doubtful case results in the analysis of hormones residues, even at such low concentration levels.
format Online
Article
Text
id pubmed-4961723
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Springer Berlin Heidelberg
record_format MEDLINE/PubMed
spelling pubmed-49617232016-08-08 Comparison of the Multiple Reaction Monitoring and Enhanced Product Ion Scan Modes for Confirmation of Stilbenes in Bovine Urine Samples Using LC–MS/MS QTRAP(®) System Matraszek-Zuchowska, Iwona Wozniak, Barbara Posyniak, Andrzej Chromatographia Original In accordance with Commission Decision 2002/657/EC, confirmatory methods for the detection of prohibited substances should comply with specific requirements, including the criteria for confirmation. Two strategies: multiple reaction monitoring (MRM) and enhanced product ion (EPI) scanning functions were compared for confirming the anabolic compounds from synthetic stilbenes group in bovine urine samples. In the research, twenty samples fortified at the Recommended Concentration (RC) of 1 µg L(−1) with diethylstilbestrol, dienestrol and hexestrol were analyzed by liquid chromatography-tandem mass spectrometry on a QTRAP 5500 instrument. The analytical procedure, validated in accordance with the Commission Decision 2002/657/EC, used in the official control of hormones in Poland was applied. The validation parameters were in agreement with 2002/657/EC performance criteria. The effectiveness of MRM and EPI scanning modes for confirmation purposes was evaluated based on the percentage of the results confirmed. In all urine samples recorded in the MRM mode, the confirmation criteria (retention time, relative intensities between transitions) have been fulfilled. The presence of stilbenes in all urine samples using EPI scan mode was confirmed too as evidenced by a good matching of stilbenes spectra in the samples to the reference spectra with critical match factor above 0.7. The results of the research show that EPI scanning function provides the same effectiveness for confirmation of banned compounds as the mostly used MRM scan mode and can be an additional tool to confirm the doubtful case results in the analysis of hormones residues, even at such low concentration levels. Springer Berlin Heidelberg 2016-07-05 2016 /pmc/articles/PMC4961723/ /pubmed/27512157 http://dx.doi.org/10.1007/s10337-016-3121-1 Text en © The Author(s) 2016 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
spellingShingle Original
Matraszek-Zuchowska, Iwona
Wozniak, Barbara
Posyniak, Andrzej
Comparison of the Multiple Reaction Monitoring and Enhanced Product Ion Scan Modes for Confirmation of Stilbenes in Bovine Urine Samples Using LC–MS/MS QTRAP(®) System
title Comparison of the Multiple Reaction Monitoring and Enhanced Product Ion Scan Modes for Confirmation of Stilbenes in Bovine Urine Samples Using LC–MS/MS QTRAP(®) System
title_full Comparison of the Multiple Reaction Monitoring and Enhanced Product Ion Scan Modes for Confirmation of Stilbenes in Bovine Urine Samples Using LC–MS/MS QTRAP(®) System
title_fullStr Comparison of the Multiple Reaction Monitoring and Enhanced Product Ion Scan Modes for Confirmation of Stilbenes in Bovine Urine Samples Using LC–MS/MS QTRAP(®) System
title_full_unstemmed Comparison of the Multiple Reaction Monitoring and Enhanced Product Ion Scan Modes for Confirmation of Stilbenes in Bovine Urine Samples Using LC–MS/MS QTRAP(®) System
title_short Comparison of the Multiple Reaction Monitoring and Enhanced Product Ion Scan Modes for Confirmation of Stilbenes in Bovine Urine Samples Using LC–MS/MS QTRAP(®) System
title_sort comparison of the multiple reaction monitoring and enhanced product ion scan modes for confirmation of stilbenes in bovine urine samples using lc–ms/ms qtrap(®) system
topic Original
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4961723/
https://www.ncbi.nlm.nih.gov/pubmed/27512157
http://dx.doi.org/10.1007/s10337-016-3121-1
work_keys_str_mv AT matraszekzuchowskaiwona comparisonofthemultiplereactionmonitoringandenhancedproductionscanmodesforconfirmationofstilbenesinbovineurinesamplesusinglcmsmsqtrapsystem
AT wozniakbarbara comparisonofthemultiplereactionmonitoringandenhancedproductionscanmodesforconfirmationofstilbenesinbovineurinesamplesusinglcmsmsqtrapsystem
AT posyniakandrzej comparisonofthemultiplereactionmonitoringandenhancedproductionscanmodesforconfirmationofstilbenesinbovineurinesamplesusinglcmsmsqtrapsystem