Fractured metallic tracheostomy tube: A rare complication of tracheostomy

Although tracheostomy is a well-accepted procedure for airway management, some early and late complications may occur. Fracture of the tracheostomy tube (TT) is a rare complication, particularly in a patient with long-term use. Herein we report a case of fractured metallic TT migrating into the trac...

Descripción completa

Detalles Bibliográficos
Autores principales: So-ngern, Apichart, Boonsarngsuk, Viboon
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4961798/
https://www.ncbi.nlm.nih.gov/pubmed/27489759
http://dx.doi.org/10.1016/j.rmcr.2016.07.007
_version_ 1782444707685597184
author So-ngern, Apichart
Boonsarngsuk, Viboon
author_facet So-ngern, Apichart
Boonsarngsuk, Viboon
author_sort So-ngern, Apichart
collection PubMed
description Although tracheostomy is a well-accepted procedure for airway management, some early and late complications may occur. Fracture of the tracheostomy tube (TT) is a rare complication, particularly in a patient with long-term use. Herein we report a case of fractured metallic TT migrating into the tracheobronchial tree. Rigid bronchoscopy was performed through the tracheostomy stoma and the fractured tube was successfully removed by a balloon catheter. Appropriate cleaning, routine careful examination, and scheduled replacement of the TT may help prevent this complication.
format Online
Article
Text
id pubmed-4961798
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-49617982016-08-03 Fractured metallic tracheostomy tube: A rare complication of tracheostomy So-ngern, Apichart Boonsarngsuk, Viboon Respir Med Case Rep Case Report Although tracheostomy is a well-accepted procedure for airway management, some early and late complications may occur. Fracture of the tracheostomy tube (TT) is a rare complication, particularly in a patient with long-term use. Herein we report a case of fractured metallic TT migrating into the tracheobronchial tree. Rigid bronchoscopy was performed through the tracheostomy stoma and the fractured tube was successfully removed by a balloon catheter. Appropriate cleaning, routine careful examination, and scheduled replacement of the TT may help prevent this complication. Elsevier 2016-07-14 /pmc/articles/PMC4961798/ /pubmed/27489759 http://dx.doi.org/10.1016/j.rmcr.2016.07.007 Text en © 2016 The Author(s) http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Case Report
So-ngern, Apichart
Boonsarngsuk, Viboon
Fractured metallic tracheostomy tube: A rare complication of tracheostomy
title Fractured metallic tracheostomy tube: A rare complication of tracheostomy
title_full Fractured metallic tracheostomy tube: A rare complication of tracheostomy
title_fullStr Fractured metallic tracheostomy tube: A rare complication of tracheostomy
title_full_unstemmed Fractured metallic tracheostomy tube: A rare complication of tracheostomy
title_short Fractured metallic tracheostomy tube: A rare complication of tracheostomy
title_sort fractured metallic tracheostomy tube: a rare complication of tracheostomy
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4961798/
https://www.ncbi.nlm.nih.gov/pubmed/27489759
http://dx.doi.org/10.1016/j.rmcr.2016.07.007
work_keys_str_mv AT songernapichart fracturedmetallictracheostomytubeararecomplicationoftracheostomy
AT boonsarngsukviboon fracturedmetallictracheostomytubeararecomplicationoftracheostomy