Evaluation of gantry speed on image quality and imaging dose for 4D cone-beam CT acquisition

BACKGROUND: This study investigates the effect of gantry speed on 4DCBCT image quality and dose for the Varian On-Board Imager®. METHODS: A thoracic 4DCBCT protocol was designed using a 125 kVp spectrum. Image quality parameters were evaluated for 4DCBCT acquisition using Catphan® phantom with real-...

Descripción completa

Detalles Bibliográficos
Autores principales: Santoso, Andrew P., Song, Kwang H., Qin, Yujiao, Gardner, Stephen J., Liu, Chang, Chetty, Indrin J., Movsas, Benjamin, Ajlouni, Munther, Wen, Ning
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4966562/
https://www.ncbi.nlm.nih.gov/pubmed/27473367
http://dx.doi.org/10.1186/s13014-016-0677-8
_version_ 1782445395852394496
author Santoso, Andrew P.
Song, Kwang H.
Qin, Yujiao
Gardner, Stephen J.
Liu, Chang
Chetty, Indrin J.
Movsas, Benjamin
Ajlouni, Munther
Wen, Ning
author_facet Santoso, Andrew P.
Song, Kwang H.
Qin, Yujiao
Gardner, Stephen J.
Liu, Chang
Chetty, Indrin J.
Movsas, Benjamin
Ajlouni, Munther
Wen, Ning
author_sort Santoso, Andrew P.
collection PubMed
description BACKGROUND: This study investigates the effect of gantry speed on 4DCBCT image quality and dose for the Varian On-Board Imager®. METHODS: A thoracic 4DCBCT protocol was designed using a 125 kVp spectrum. Image quality parameters were evaluated for 4DCBCT acquisition using Catphan® phantom with real-time position management™ system for gantry speeds varying between 1.0 to 6.0°/s. Superior-inferior motion of the phantom was executed using a sinusoidal waveform with five second period. Scans were retrospectively sorted into 4 phases (CBCT-4 ph) and 10 phases (CBCT-10 ph); average 4DCBCT (CBCT-ave), using all image data from the 4DCBCT acquisitions was also evaluated. The 4DCBCT images were evaluated using the following image quality metrics: spatial resolution, contrast-to-noise ratio (CNR), and uniformity index (UI). Additionally, Hounsfield unit (HU) sensitivity compared to a baseline CBCT and percent differences and RMS errors (RMSE) of excursion were also determined. Imaging dose was evaluated using an IBA CC13 ion chamber placed within CIRS Thorax phantom using the same sinusoidal motion and image acquisition settings as mentioned above. RESULTS: Spatial resolution decreased linearly from 5.93 to 3.82 lp/cm as gantry speed increased from 1.0 to 6.0°/s. CNR decreased linearly from 4.80 to 1.82 with gantry speed increasing from 1.0 to 6.0°/s, respectively. No noteworthy variations in UI, HU sensitivity, or excursion metrics were observed with changes in gantry speed. Ion chamber dose rates measured ranged from 2.30 (lung) to 5.18 (bone) E-3 cGy/mAs. CONCLUSIONS: A quantitative analysis of the Varian OBI’s 4DCBCT capabilities was explored. Changing gantry speed changes the number of projections used for reconstruction, affecting both image quality and imaging dose if x-ray tube current is held constant. From the results of this study, a gantry speed between 2 and 3°/s was optimal when considering image quality, dose, and reconstruction time. The future of 4DCBCT clinical utility relies on further investigation of image acquisition and reconstruction optimization.
format Online
Article
Text
id pubmed-4966562
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-49665622016-07-30 Evaluation of gantry speed on image quality and imaging dose for 4D cone-beam CT acquisition Santoso, Andrew P. Song, Kwang H. Qin, Yujiao Gardner, Stephen J. Liu, Chang Chetty, Indrin J. Movsas, Benjamin Ajlouni, Munther Wen, Ning Radiat Oncol Research BACKGROUND: This study investigates the effect of gantry speed on 4DCBCT image quality and dose for the Varian On-Board Imager®. METHODS: A thoracic 4DCBCT protocol was designed using a 125 kVp spectrum. Image quality parameters were evaluated for 4DCBCT acquisition using Catphan® phantom with real-time position management™ system for gantry speeds varying between 1.0 to 6.0°/s. Superior-inferior motion of the phantom was executed using a sinusoidal waveform with five second period. Scans were retrospectively sorted into 4 phases (CBCT-4 ph) and 10 phases (CBCT-10 ph); average 4DCBCT (CBCT-ave), using all image data from the 4DCBCT acquisitions was also evaluated. The 4DCBCT images were evaluated using the following image quality metrics: spatial resolution, contrast-to-noise ratio (CNR), and uniformity index (UI). Additionally, Hounsfield unit (HU) sensitivity compared to a baseline CBCT and percent differences and RMS errors (RMSE) of excursion were also determined. Imaging dose was evaluated using an IBA CC13 ion chamber placed within CIRS Thorax phantom using the same sinusoidal motion and image acquisition settings as mentioned above. RESULTS: Spatial resolution decreased linearly from 5.93 to 3.82 lp/cm as gantry speed increased from 1.0 to 6.0°/s. CNR decreased linearly from 4.80 to 1.82 with gantry speed increasing from 1.0 to 6.0°/s, respectively. No noteworthy variations in UI, HU sensitivity, or excursion metrics were observed with changes in gantry speed. Ion chamber dose rates measured ranged from 2.30 (lung) to 5.18 (bone) E-3 cGy/mAs. CONCLUSIONS: A quantitative analysis of the Varian OBI’s 4DCBCT capabilities was explored. Changing gantry speed changes the number of projections used for reconstruction, affecting both image quality and imaging dose if x-ray tube current is held constant. From the results of this study, a gantry speed between 2 and 3°/s was optimal when considering image quality, dose, and reconstruction time. The future of 4DCBCT clinical utility relies on further investigation of image acquisition and reconstruction optimization. BioMed Central 2016-07-29 /pmc/articles/PMC4966562/ /pubmed/27473367 http://dx.doi.org/10.1186/s13014-016-0677-8 Text en © The Author(s). 2016 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research
Santoso, Andrew P.
Song, Kwang H.
Qin, Yujiao
Gardner, Stephen J.
Liu, Chang
Chetty, Indrin J.
Movsas, Benjamin
Ajlouni, Munther
Wen, Ning
Evaluation of gantry speed on image quality and imaging dose for 4D cone-beam CT acquisition
title Evaluation of gantry speed on image quality and imaging dose for 4D cone-beam CT acquisition
title_full Evaluation of gantry speed on image quality and imaging dose for 4D cone-beam CT acquisition
title_fullStr Evaluation of gantry speed on image quality and imaging dose for 4D cone-beam CT acquisition
title_full_unstemmed Evaluation of gantry speed on image quality and imaging dose for 4D cone-beam CT acquisition
title_short Evaluation of gantry speed on image quality and imaging dose for 4D cone-beam CT acquisition
title_sort evaluation of gantry speed on image quality and imaging dose for 4d cone-beam ct acquisition
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4966562/
https://www.ncbi.nlm.nih.gov/pubmed/27473367
http://dx.doi.org/10.1186/s13014-016-0677-8
work_keys_str_mv AT santosoandrewp evaluationofgantryspeedonimagequalityandimagingdosefor4dconebeamctacquisition
AT songkwangh evaluationofgantryspeedonimagequalityandimagingdosefor4dconebeamctacquisition
AT qinyujiao evaluationofgantryspeedonimagequalityandimagingdosefor4dconebeamctacquisition
AT gardnerstephenj evaluationofgantryspeedonimagequalityandimagingdosefor4dconebeamctacquisition
AT liuchang evaluationofgantryspeedonimagequalityandimagingdosefor4dconebeamctacquisition
AT chettyindrinj evaluationofgantryspeedonimagequalityandimagingdosefor4dconebeamctacquisition
AT movsasbenjamin evaluationofgantryspeedonimagequalityandimagingdosefor4dconebeamctacquisition
AT ajlounimunther evaluationofgantryspeedonimagequalityandimagingdosefor4dconebeamctacquisition
AT wenning evaluationofgantryspeedonimagequalityandimagingdosefor4dconebeamctacquisition