Cargando…

Storing of Extracts in Polypropylene Microcentrifuge Tubes Yields Contaminant Peak During Ultra-flow Liquid Chromatographic Analysis

BACKGROUND AND AIM: This study was designed to understand the effect of storage in polypropylene microcentrifuge tubes and glass vials during ultra-flow liquid chromatographic (UFLC) analysis. MATERIALS AND METHODS: One ml of methanol was placed in polypropylene microcentrifuge tubes (PP material, A...

Descripción completa

Detalles Bibliográficos
Autores principales: Kshirsagar, Parthraj R., Hegde, Harsha, Pai, Sandeep R.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4971948/
https://www.ncbi.nlm.nih.gov/pubmed/27563216
http://dx.doi.org/10.4103/0973-1296.185719
Descripción
Sumario:BACKGROUND AND AIM: This study was designed to understand the effect of storage in polypropylene microcentrifuge tubes and glass vials during ultra-flow liquid chromatographic (UFLC) analysis. MATERIALS AND METHODS: One ml of methanol was placed in polypropylene microcentrifuge tubes (PP material, Autoclavable) and glass vials (Borosilicate) separately for 1, 2, 4, 8, 10, 20, 40, and 80 days intervals stored at −4°C. RESULTS: Contaminant peak was detected in methanol stored in polypropylene microcentrifuge tubes using UFLC analysis. The contaminant peak detected was prominent, sharp detectable at 9.176 ± 0.138 min on a Waters 250–4.6 mm, 4 μ, Nova-Pak C18 column with mobile phase consisting of methanol:water (70:30). CONCLUSION: It was evident from the study that long-term storage of biological samples prepared using methanol in polypropylene microcentrifuge tubes produce contaminant peak. Further, this may mislead in future reporting an unnatural compound by researchers. SUMMARY: Long-term storage of biological samples prepared using methanol in polypropylene microcentrifuge tubes produce contaminant peak. Contamination peak with higher area under the curve (609993) was obtained in ultra-flow liquid chromatographic run for methanol stored in PP microcentrifuge tubes. Contamination peak was detected at retention time 9.113 min with a lambda max of 220.38 nm and 300 mAU intensity on the given chromatographic conditions. Glass vials serve better option over PP microcentrifuge tubes for storing biological samples. Abbreviations used: UFLC: Ultra Flow Liquid Chromatography; LC: Liquid Chromatography; MS: Mass spectrometry; AUC: Area Under Curve.