Lipid suppression via double inversion recovery with symmetric frequency sweep for robust 2D‐GRAPPA‐accelerated MRSI of the brain at 7 T

This work presents a new approach for high‐resolution MRSI of the brain at 7 T in clinically feasible measurement times. Two major problems of MRSI are the long scan times for large matrix sizes and the possible spectral contamination by the transcranial lipid signal. We propose a combination of fre...

Descripción completa

Detalles Bibliográficos
Autores principales: Hangel, Gilbert, Strasser, Bernhard, Považan, Michal, Gruber, Stephan, Chmelík, Marek, Gajdošík, Martin, Trattnig, Siegfried, Bogner, Wolfgang
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4973691/
https://www.ncbi.nlm.nih.gov/pubmed/26370781
http://dx.doi.org/10.1002/nbm.3386
_version_ 1782446436354359296
author Hangel, Gilbert
Strasser, Bernhard
Považan, Michal
Gruber, Stephan
Chmelík, Marek
Gajdošík, Martin
Trattnig, Siegfried
Bogner, Wolfgang
author_facet Hangel, Gilbert
Strasser, Bernhard
Považan, Michal
Gruber, Stephan
Chmelík, Marek
Gajdošík, Martin
Trattnig, Siegfried
Bogner, Wolfgang
author_sort Hangel, Gilbert
collection PubMed
description This work presents a new approach for high‐resolution MRSI of the brain at 7 T in clinically feasible measurement times. Two major problems of MRSI are the long scan times for large matrix sizes and the possible spectral contamination by the transcranial lipid signal. We propose a combination of free induction decay (FID)‐MRSI with a short acquisition delay and acceleration via in‐plane two‐dimensional generalised autocalibrating partially parallel acquisition (2D‐GRAPPA) with adiabatic double inversion recovery (IR)‐based lipid suppression to allow robust high‐resolution MRSI. We performed Bloch simulations to evaluate the magnetisation pathways of lipids and metabolites, and compared the results with phantom measurements. Acceleration factors in the range 2–25 were tested in a phantom. Five volunteers were scanned to verify the value of our MRSI method in vivo. GRAPPA artefacts that cause fold‐in of transcranial lipids were suppressed via double IR, with a non‐selective symmetric frequency sweep. The use of long, low‐power inversion pulses (100 ms) reduced specific absorption rate requirements. The symmetric frequency sweep over both pulses provided good lipid suppression (>90%), in addition to a reduced loss in metabolite signal‐to‐noise ratio (SNR), compared with conventional IR suppression (52–70%). The metabolic mapping over the whole brain slice was not limited to a rectangular region of interest. 2D‐GRAPPA provided acceleration up to a factor of nine for in vivo FID‐MRSI without a substantial increase in g‐factors (<1.1). A 64 × 64 matrix can be acquired with a common repetition time of ~1.3 s in only 8 min without lipid artefacts caused by acceleration. Overall, we present a fast and robust MRSI method, using combined double IR fat suppression and 2D‐GRAPPA acceleration, which may be used in (pre)clinical studies of the brain at 7 T. © 2015 The Authors. NMR in Biomedicine published by John Wiley & Sons Ltd.
format Online
Article
Text
id pubmed-4973691
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-49736912016-08-17 Lipid suppression via double inversion recovery with symmetric frequency sweep for robust 2D‐GRAPPA‐accelerated MRSI of the brain at 7 T Hangel, Gilbert Strasser, Bernhard Považan, Michal Gruber, Stephan Chmelík, Marek Gajdošík, Martin Trattnig, Siegfried Bogner, Wolfgang NMR Biomed Research Articles This work presents a new approach for high‐resolution MRSI of the brain at 7 T in clinically feasible measurement times. Two major problems of MRSI are the long scan times for large matrix sizes and the possible spectral contamination by the transcranial lipid signal. We propose a combination of free induction decay (FID)‐MRSI with a short acquisition delay and acceleration via in‐plane two‐dimensional generalised autocalibrating partially parallel acquisition (2D‐GRAPPA) with adiabatic double inversion recovery (IR)‐based lipid suppression to allow robust high‐resolution MRSI. We performed Bloch simulations to evaluate the magnetisation pathways of lipids and metabolites, and compared the results with phantom measurements. Acceleration factors in the range 2–25 were tested in a phantom. Five volunteers were scanned to verify the value of our MRSI method in vivo. GRAPPA artefacts that cause fold‐in of transcranial lipids were suppressed via double IR, with a non‐selective symmetric frequency sweep. The use of long, low‐power inversion pulses (100 ms) reduced specific absorption rate requirements. The symmetric frequency sweep over both pulses provided good lipid suppression (>90%), in addition to a reduced loss in metabolite signal‐to‐noise ratio (SNR), compared with conventional IR suppression (52–70%). The metabolic mapping over the whole brain slice was not limited to a rectangular region of interest. 2D‐GRAPPA provided acceleration up to a factor of nine for in vivo FID‐MRSI without a substantial increase in g‐factors (<1.1). A 64 × 64 matrix can be acquired with a common repetition time of ~1.3 s in only 8 min without lipid artefacts caused by acceleration. Overall, we present a fast and robust MRSI method, using combined double IR fat suppression and 2D‐GRAPPA acceleration, which may be used in (pre)clinical studies of the brain at 7 T. © 2015 The Authors. NMR in Biomedicine published by John Wiley & Sons Ltd. John Wiley and Sons Inc. 2015-09-15 2015-11 /pmc/articles/PMC4973691/ /pubmed/26370781 http://dx.doi.org/10.1002/nbm.3386 Text en © 2015 The Authors. NMR in Biomedicine published by John Wiley & Sons Ltd. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Articles
Hangel, Gilbert
Strasser, Bernhard
Považan, Michal
Gruber, Stephan
Chmelík, Marek
Gajdošík, Martin
Trattnig, Siegfried
Bogner, Wolfgang
Lipid suppression via double inversion recovery with symmetric frequency sweep for robust 2D‐GRAPPA‐accelerated MRSI of the brain at 7 T
title Lipid suppression via double inversion recovery with symmetric frequency sweep for robust 2D‐GRAPPA‐accelerated MRSI of the brain at 7 T
title_full Lipid suppression via double inversion recovery with symmetric frequency sweep for robust 2D‐GRAPPA‐accelerated MRSI of the brain at 7 T
title_fullStr Lipid suppression via double inversion recovery with symmetric frequency sweep for robust 2D‐GRAPPA‐accelerated MRSI of the brain at 7 T
title_full_unstemmed Lipid suppression via double inversion recovery with symmetric frequency sweep for robust 2D‐GRAPPA‐accelerated MRSI of the brain at 7 T
title_short Lipid suppression via double inversion recovery with symmetric frequency sweep for robust 2D‐GRAPPA‐accelerated MRSI of the brain at 7 T
title_sort lipid suppression via double inversion recovery with symmetric frequency sweep for robust 2d‐grappa‐accelerated mrsi of the brain at 7 t
topic Research Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4973691/
https://www.ncbi.nlm.nih.gov/pubmed/26370781
http://dx.doi.org/10.1002/nbm.3386
work_keys_str_mv AT hangelgilbert lipidsuppressionviadoubleinversionrecoverywithsymmetricfrequencysweepforrobust2dgrappaacceleratedmrsiofthebrainat7t
AT strasserbernhard lipidsuppressionviadoubleinversionrecoverywithsymmetricfrequencysweepforrobust2dgrappaacceleratedmrsiofthebrainat7t
AT povazanmichal lipidsuppressionviadoubleinversionrecoverywithsymmetricfrequencysweepforrobust2dgrappaacceleratedmrsiofthebrainat7t
AT gruberstephan lipidsuppressionviadoubleinversionrecoverywithsymmetricfrequencysweepforrobust2dgrappaacceleratedmrsiofthebrainat7t
AT chmelikmarek lipidsuppressionviadoubleinversionrecoverywithsymmetricfrequencysweepforrobust2dgrappaacceleratedmrsiofthebrainat7t
AT gajdosikmartin lipidsuppressionviadoubleinversionrecoverywithsymmetricfrequencysweepforrobust2dgrappaacceleratedmrsiofthebrainat7t
AT trattnigsiegfried lipidsuppressionviadoubleinversionrecoverywithsymmetricfrequencysweepforrobust2dgrappaacceleratedmrsiofthebrainat7t
AT bognerwolfgang lipidsuppressionviadoubleinversionrecoverywithsymmetricfrequencysweepforrobust2dgrappaacceleratedmrsiofthebrainat7t