Ex vivo testing of intact eye globes under inflation conditions to determine regional variation of mechanical stiffness
BACKGROUND: The eye globe exhibits significant regional variation of mechanical behaviour. The aim of this present study is to develop a new experimental technique for testing intact eye globes in a form that is representative of in vivo conditions, and therefore suitable for determining the materia...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4979203/ https://www.ncbi.nlm.nih.gov/pubmed/27512719 http://dx.doi.org/10.1186/s40662-016-0052-8 |
_version_ | 1782447285501689856 |
---|---|
author | Whitford, Charles Joda, Akram Jones, Steve Bao, Fangjun Rama, Paolo Elsheikh, Ahmed |
author_facet | Whitford, Charles Joda, Akram Jones, Steve Bao, Fangjun Rama, Paolo Elsheikh, Ahmed |
author_sort | Whitford, Charles |
collection | PubMed |
description | BACKGROUND: The eye globe exhibits significant regional variation of mechanical behaviour. The aim of this present study is to develop a new experimental technique for testing intact eye globes in a form that is representative of in vivo conditions, and therefore suitable for determining the material properties of the complete outer ocular tunic. METHODS: A test rig has been developed to provide closed-loop control of either applied intra-ocular pressure or resulting apical displacement; measurement of displacements across the external surface of the eye globe using high-resolution digital cameras and digital image correlation software; prevention of rigid-body motion and protection of the ocular surface from environmental drying. The method has been demonstrated on one human and one porcine eye globe, which were cyclically loaded. Finite element models based on specimen specific tomography, free from rotational symmetry, were used along with experimental pressure-displacement data in an inverse analysis process to derive the mechanical properties of tissue in different regions of the eye’s outer tunic. RESULTS: The test method enabled monitoring of mechanical response to intraocular pressure variation across the surface of the eye globe. For the two eyes tested, the method showed a gradual change in the sclera’s stiffness from a maximum at the limbus to a minimum at the posterior pole, while in the cornea the stiffness was highest at the centre and lowest in the peripheral zone. Further, for both the sclera and cornea, the load–displacement behaviour did not vary significantly between loading cycles. CONCLUSIONS: The first methodology capable of mechanically testing intact eye globes, with applied loads and boundary conditions that closely represent in vivo conditions is introduced. The method enables determination of the regional variation in mechanical behaviour across the ocular surface. |
format | Online Article Text |
id | pubmed-4979203 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-49792032016-08-11 Ex vivo testing of intact eye globes under inflation conditions to determine regional variation of mechanical stiffness Whitford, Charles Joda, Akram Jones, Steve Bao, Fangjun Rama, Paolo Elsheikh, Ahmed Eye Vis (Lond) Research BACKGROUND: The eye globe exhibits significant regional variation of mechanical behaviour. The aim of this present study is to develop a new experimental technique for testing intact eye globes in a form that is representative of in vivo conditions, and therefore suitable for determining the material properties of the complete outer ocular tunic. METHODS: A test rig has been developed to provide closed-loop control of either applied intra-ocular pressure or resulting apical displacement; measurement of displacements across the external surface of the eye globe using high-resolution digital cameras and digital image correlation software; prevention of rigid-body motion and protection of the ocular surface from environmental drying. The method has been demonstrated on one human and one porcine eye globe, which were cyclically loaded. Finite element models based on specimen specific tomography, free from rotational symmetry, were used along with experimental pressure-displacement data in an inverse analysis process to derive the mechanical properties of tissue in different regions of the eye’s outer tunic. RESULTS: The test method enabled monitoring of mechanical response to intraocular pressure variation across the surface of the eye globe. For the two eyes tested, the method showed a gradual change in the sclera’s stiffness from a maximum at the limbus to a minimum at the posterior pole, while in the cornea the stiffness was highest at the centre and lowest in the peripheral zone. Further, for both the sclera and cornea, the load–displacement behaviour did not vary significantly between loading cycles. CONCLUSIONS: The first methodology capable of mechanically testing intact eye globes, with applied loads and boundary conditions that closely represent in vivo conditions is introduced. The method enables determination of the regional variation in mechanical behaviour across the ocular surface. BioMed Central 2016-08-10 /pmc/articles/PMC4979203/ /pubmed/27512719 http://dx.doi.org/10.1186/s40662-016-0052-8 Text en © The Author(s). 2016 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
spellingShingle | Research Whitford, Charles Joda, Akram Jones, Steve Bao, Fangjun Rama, Paolo Elsheikh, Ahmed Ex vivo testing of intact eye globes under inflation conditions to determine regional variation of mechanical stiffness |
title | Ex vivo testing of intact eye globes under inflation conditions to determine regional variation of mechanical stiffness |
title_full | Ex vivo testing of intact eye globes under inflation conditions to determine regional variation of mechanical stiffness |
title_fullStr | Ex vivo testing of intact eye globes under inflation conditions to determine regional variation of mechanical stiffness |
title_full_unstemmed | Ex vivo testing of intact eye globes under inflation conditions to determine regional variation of mechanical stiffness |
title_short | Ex vivo testing of intact eye globes under inflation conditions to determine regional variation of mechanical stiffness |
title_sort | ex vivo testing of intact eye globes under inflation conditions to determine regional variation of mechanical stiffness |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4979203/ https://www.ncbi.nlm.nih.gov/pubmed/27512719 http://dx.doi.org/10.1186/s40662-016-0052-8 |
work_keys_str_mv | AT whitfordcharles exvivotestingofintacteyeglobesunderinflationconditionstodetermineregionalvariationofmechanicalstiffness AT jodaakram exvivotestingofintacteyeglobesunderinflationconditionstodetermineregionalvariationofmechanicalstiffness AT jonessteve exvivotestingofintacteyeglobesunderinflationconditionstodetermineregionalvariationofmechanicalstiffness AT baofangjun exvivotestingofintacteyeglobesunderinflationconditionstodetermineregionalvariationofmechanicalstiffness AT ramapaolo exvivotestingofintacteyeglobesunderinflationconditionstodetermineregionalvariationofmechanicalstiffness AT elsheikhahmed exvivotestingofintacteyeglobesunderinflationconditionstodetermineregionalvariationofmechanicalstiffness |