Combined uncertainty estimation for the determination of the dissolved iron amount content in seawater using flow injection with chemiluminescence detection

This work assesses the components contributing to the combined uncertainty budget associated with the measurement of the Fe amount content by flow injection chemiluminescence (FI‐CL) in <0.2 μm filtered and acidified seawater samples. Amounts of loaded standard solutions and samples were determin...

Descripción completa

Detalles Bibliográficos
Autores principales: Floor, Geerke H., Clough, Robert, Lohan, Maeve C., Ussher, Simon J., Worsfold, Paul J., Quétel, Christophe R.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4995610/
https://www.ncbi.nlm.nih.gov/pubmed/27610049
http://dx.doi.org/10.1002/lom3.10057
_version_ 1782449504194134016
author Floor, Geerke H.
Clough, Robert
Lohan, Maeve C.
Ussher, Simon J.
Worsfold, Paul J.
Quétel, Christophe R.
author_facet Floor, Geerke H.
Clough, Robert
Lohan, Maeve C.
Ussher, Simon J.
Worsfold, Paul J.
Quétel, Christophe R.
author_sort Floor, Geerke H.
collection PubMed
description This work assesses the components contributing to the combined uncertainty budget associated with the measurement of the Fe amount content by flow injection chemiluminescence (FI‐CL) in <0.2 μm filtered and acidified seawater samples. Amounts of loaded standard solutions and samples were determined gravimetrically by differential weighing. Up to 5% variations in the loaded masses were observed during measurements, in contradiction to the usual assumptions made when operating under constant loading time conditions. Hence signal intensities (V) were normalised to the loaded mass and plots of average normalised intensities (in V kg(−1)) vs. values of the Fe amount content (in nmol kg(−1)) added to a “low level” iron seawater matrix were used to produce the calibration graphs. The measurement procedure implemented and the uncertainty estimation process developed were validated from the agreement obtained with consensus values for three SAFe and GEOTRACES reference materials (D2, GS, and GD). Relative expanded uncertainties for peak height and peak area based results were estimated to be around 12% and 10% (coverage factor k = 2), respectively. The most important contributory factors were the uncertainty on the sensitivity coefficient (i.e., calibration slope) and the within‐sequence‐stability (i.e., the signal stability over several hours of operation; here 32 h). For GD, using peak height measurements, these factors contributed respectively 69.7% and 21.6% while the short‐term repeatability accounted for only 7.9%. Therefore, an uncertainty estimation based on the intensity repeatability alone, as is often done in FI‐CL studies, is not a realistic estimation of the overall uncertainty of the procedure.
format Online
Article
Text
id pubmed-4995610
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-49956102016-09-06 Combined uncertainty estimation for the determination of the dissolved iron amount content in seawater using flow injection with chemiluminescence detection Floor, Geerke H. Clough, Robert Lohan, Maeve C. Ussher, Simon J. Worsfold, Paul J. Quétel, Christophe R. Limnol Oceanogr Methods Evaluations of Existing Methods This work assesses the components contributing to the combined uncertainty budget associated with the measurement of the Fe amount content by flow injection chemiluminescence (FI‐CL) in <0.2 μm filtered and acidified seawater samples. Amounts of loaded standard solutions and samples were determined gravimetrically by differential weighing. Up to 5% variations in the loaded masses were observed during measurements, in contradiction to the usual assumptions made when operating under constant loading time conditions. Hence signal intensities (V) were normalised to the loaded mass and plots of average normalised intensities (in V kg(−1)) vs. values of the Fe amount content (in nmol kg(−1)) added to a “low level” iron seawater matrix were used to produce the calibration graphs. The measurement procedure implemented and the uncertainty estimation process developed were validated from the agreement obtained with consensus values for three SAFe and GEOTRACES reference materials (D2, GS, and GD). Relative expanded uncertainties for peak height and peak area based results were estimated to be around 12% and 10% (coverage factor k = 2), respectively. The most important contributory factors were the uncertainty on the sensitivity coefficient (i.e., calibration slope) and the within‐sequence‐stability (i.e., the signal stability over several hours of operation; here 32 h). For GD, using peak height measurements, these factors contributed respectively 69.7% and 21.6% while the short‐term repeatability accounted for only 7.9%. Therefore, an uncertainty estimation based on the intensity repeatability alone, as is often done in FI‐CL studies, is not a realistic estimation of the overall uncertainty of the procedure. John Wiley and Sons Inc. 2015-09-18 2015-12 /pmc/articles/PMC4995610/ /pubmed/27610049 http://dx.doi.org/10.1002/lom3.10057 Text en © 2015 The Authors Limnology and Oceanography: Methods published by Wiley Periodicals, Inc. on behalf of Association for the Sciences of Limnology and Oceanography This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Evaluations of Existing Methods
Floor, Geerke H.
Clough, Robert
Lohan, Maeve C.
Ussher, Simon J.
Worsfold, Paul J.
Quétel, Christophe R.
Combined uncertainty estimation for the determination of the dissolved iron amount content in seawater using flow injection with chemiluminescence detection
title Combined uncertainty estimation for the determination of the dissolved iron amount content in seawater using flow injection with chemiluminescence detection
title_full Combined uncertainty estimation for the determination of the dissolved iron amount content in seawater using flow injection with chemiluminescence detection
title_fullStr Combined uncertainty estimation for the determination of the dissolved iron amount content in seawater using flow injection with chemiluminescence detection
title_full_unstemmed Combined uncertainty estimation for the determination of the dissolved iron amount content in seawater using flow injection with chemiluminescence detection
title_short Combined uncertainty estimation for the determination of the dissolved iron amount content in seawater using flow injection with chemiluminescence detection
title_sort combined uncertainty estimation for the determination of the dissolved iron amount content in seawater using flow injection with chemiluminescence detection
topic Evaluations of Existing Methods
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4995610/
https://www.ncbi.nlm.nih.gov/pubmed/27610049
http://dx.doi.org/10.1002/lom3.10057
work_keys_str_mv AT floorgeerkeh combineduncertaintyestimationforthedeterminationofthedissolvedironamountcontentinseawaterusingflowinjectionwithchemiluminescencedetection
AT cloughrobert combineduncertaintyestimationforthedeterminationofthedissolvedironamountcontentinseawaterusingflowinjectionwithchemiluminescencedetection
AT lohanmaevec combineduncertaintyestimationforthedeterminationofthedissolvedironamountcontentinseawaterusingflowinjectionwithchemiluminescencedetection
AT usshersimonj combineduncertaintyestimationforthedeterminationofthedissolvedironamountcontentinseawaterusingflowinjectionwithchemiluminescencedetection
AT worsfoldpaulj combineduncertaintyestimationforthedeterminationofthedissolvedironamountcontentinseawaterusingflowinjectionwithchemiluminescencedetection
AT quetelchristopher combineduncertaintyestimationforthedeterminationofthedissolvedironamountcontentinseawaterusingflowinjectionwithchemiluminescencedetection