Cargando…
Combined uncertainty estimation for the determination of the dissolved iron amount content in seawater using flow injection with chemiluminescence detection
This work assesses the components contributing to the combined uncertainty budget associated with the measurement of the Fe amount content by flow injection chemiluminescence (FI‐CL) in <0.2 μm filtered and acidified seawater samples. Amounts of loaded standard solutions and samples were determin...
Autores principales: | Floor, Geerke H., Clough, Robert, Lohan, Maeve C., Ussher, Simon J., Worsfold, Paul J., Quétel, Christophe R. |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4995610/ https://www.ncbi.nlm.nih.gov/pubmed/27610049 http://dx.doi.org/10.1002/lom3.10057 |
Ejemplares similares
-
Evaluation of marine zooplankton community structure through environmental DNA metabarcoding
por: Djurhuus, Anni, et al.
Publicado: (2018) -
Evaluation of diagnostic pigments to estimate phytoplankton size classes
por: Chase, Alison P., et al.
Publicado: (2020) -
A comparative study of metatranscriptomic assessment methods to characterize
Microcystis
blooms
por: Pound, Helena L., et al.
Publicado: (2021) -
Design of an Automated Flow Injection-Chemiluminescence Instrument Incorporating a Miniature Photomultiplier Tube for Monitoring Picomolar Concentrations of Iron in Seawater
por: Bowie, Andrew R., et al.
Publicado: (2005) -
Integrated luminometer for the determination of trace metals in seawater using fluorescence, phosphorescence and chemiluminescence detection
por: Worsfold, P. J., et al.
Publicado: (2002)