Wear Scar Similarities between Retrieved and Simulator-Tested Polyethylene TKR Components: An Artificial Neural Network Approach

The aim of this study was to determine how representative wear scars of simulator-tested polyethylene (PE) inserts compare with retrieved PE inserts from total knee replacement (TKR). By means of a nonparametric self-organizing feature map (SOFM), wear scar images of 21 postmortem- and 54 revision-r...

Descripción completa

Detalles Bibliográficos
Autores principales: Orozco Villaseñor, Diego A., Wimmer, Markus A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5002291/
https://www.ncbi.nlm.nih.gov/pubmed/27597955
http://dx.doi.org/10.1155/2016/2071945
_version_ 1782450547020791808
author Orozco Villaseñor, Diego A.
Wimmer, Markus A.
author_facet Orozco Villaseñor, Diego A.
Wimmer, Markus A.
author_sort Orozco Villaseñor, Diego A.
collection PubMed
description The aim of this study was to determine how representative wear scars of simulator-tested polyethylene (PE) inserts compare with retrieved PE inserts from total knee replacement (TKR). By means of a nonparametric self-organizing feature map (SOFM), wear scar images of 21 postmortem- and 54 revision-retrieved components were compared with six simulator-tested components that were tested either in displacement or in load control according to ISO protocols. The SOFM network was then trained with the wear scar images of postmortem-retrieved components since those are considered well-functioning at the time of retrieval. Based on this training process, eleven clusters were established, suggesting considerable variability among wear scars despite an uncomplicated loading history inside their hosts. The remaining components (revision-retrieved and simulator-tested) were then assigned to these established clusters. Six out of five simulator components were clustered together, suggesting that the network was able to identify similarities in loading history. However, the simulator-tested components ended up in a cluster at the fringe of the map containing only 10.8% of retrieved components. This may suggest that current ISO testing protocols were not fully representative of this TKR population, and protocols that better resemble patients' gait after TKR containing activities other than walking may be warranted.
format Online
Article
Text
id pubmed-5002291
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Hindawi Publishing Corporation
record_format MEDLINE/PubMed
spelling pubmed-50022912016-09-05 Wear Scar Similarities between Retrieved and Simulator-Tested Polyethylene TKR Components: An Artificial Neural Network Approach Orozco Villaseñor, Diego A. Wimmer, Markus A. Biomed Res Int Research Article The aim of this study was to determine how representative wear scars of simulator-tested polyethylene (PE) inserts compare with retrieved PE inserts from total knee replacement (TKR). By means of a nonparametric self-organizing feature map (SOFM), wear scar images of 21 postmortem- and 54 revision-retrieved components were compared with six simulator-tested components that were tested either in displacement or in load control according to ISO protocols. The SOFM network was then trained with the wear scar images of postmortem-retrieved components since those are considered well-functioning at the time of retrieval. Based on this training process, eleven clusters were established, suggesting considerable variability among wear scars despite an uncomplicated loading history inside their hosts. The remaining components (revision-retrieved and simulator-tested) were then assigned to these established clusters. Six out of five simulator components were clustered together, suggesting that the network was able to identify similarities in loading history. However, the simulator-tested components ended up in a cluster at the fringe of the map containing only 10.8% of retrieved components. This may suggest that current ISO testing protocols were not fully representative of this TKR population, and protocols that better resemble patients' gait after TKR containing activities other than walking may be warranted. Hindawi Publishing Corporation 2016 2016-08-14 /pmc/articles/PMC5002291/ /pubmed/27597955 http://dx.doi.org/10.1155/2016/2071945 Text en Copyright © 2016 D. A. Orozco Villaseñor and M. A. Wimmer. https://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Orozco Villaseñor, Diego A.
Wimmer, Markus A.
Wear Scar Similarities between Retrieved and Simulator-Tested Polyethylene TKR Components: An Artificial Neural Network Approach
title Wear Scar Similarities between Retrieved and Simulator-Tested Polyethylene TKR Components: An Artificial Neural Network Approach
title_full Wear Scar Similarities between Retrieved and Simulator-Tested Polyethylene TKR Components: An Artificial Neural Network Approach
title_fullStr Wear Scar Similarities between Retrieved and Simulator-Tested Polyethylene TKR Components: An Artificial Neural Network Approach
title_full_unstemmed Wear Scar Similarities between Retrieved and Simulator-Tested Polyethylene TKR Components: An Artificial Neural Network Approach
title_short Wear Scar Similarities between Retrieved and Simulator-Tested Polyethylene TKR Components: An Artificial Neural Network Approach
title_sort wear scar similarities between retrieved and simulator-tested polyethylene tkr components: an artificial neural network approach
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5002291/
https://www.ncbi.nlm.nih.gov/pubmed/27597955
http://dx.doi.org/10.1155/2016/2071945
work_keys_str_mv AT orozcovillasenordiegoa wearscarsimilaritiesbetweenretrievedandsimulatortestedpolyethylenetkrcomponentsanartificialneuralnetworkapproach
AT wimmermarkusa wearscarsimilaritiesbetweenretrievedandsimulatortestedpolyethylenetkrcomponentsanartificialneuralnetworkapproach