Battery drain in daily practice and medium-term projections on longevity of cardioverter-defibrillators: an analysis from a remote monitoring database
AIMS: The longevity of generators is a crucial determinant of the cost-effectiveness of therapy with implantable cardioverter-defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-D). We evaluated the trend of device-measured residual battery capacity and longevity projecti...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Oxford University Press
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5006961/ https://www.ncbi.nlm.nih.gov/pubmed/26847076 http://dx.doi.org/10.1093/europace/euv436 |
_version_ | 1782451145473523712 |
---|---|
author | Boriani, Giuseppe Ritter, Philippe Biffi, Mauro Ziacchi, Matteo Diemberger, Igor Martignani, Cristian Valzania, Cinzia Valsecchi, Sergio Padeletti, Luigi Gadler, Fredrik |
author_facet | Boriani, Giuseppe Ritter, Philippe Biffi, Mauro Ziacchi, Matteo Diemberger, Igor Martignani, Cristian Valzania, Cinzia Valsecchi, Sergio Padeletti, Luigi Gadler, Fredrik |
author_sort | Boriani, Giuseppe |
collection | PubMed |
description | AIMS: The longevity of generators is a crucial determinant of the cost-effectiveness of therapy with implantable cardioverter-defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-D). We evaluated the trend of device-measured residual battery capacity and longevity projections over 5-year follow-up. We also investigated possible factors associated with battery drain. METHODS AND RESULTS: Data from 4851 patients in the European LATITUDE(®) database who were followed up for a minimum of 3 years were analysed. The factors associated with battery drain (i.e. year-to-year decrease in residual battery capacity), and thus potentially impacting on device longevity, were mainly the pacing parameters in CRT-D devices and the number of shocks delivered and diverted in both ICD and CRT-D (all P < 0.01 on linear regression analysis). Over the first 5 years, the longevity estimates provided by devices showed low intra-patient variability and increased with time. The estimates exceeded 10 years for CRT-D and 13 and 12 years for single- and dual-chamber ICDs, respectively. In CRT-D patients, the expected patient age on replacement was 80 ± 12 years, and the expected probability of undergoing device replacement was 63 ± 13% for New York Heart Association (NYHA) II patients and 37 ± 16% for NYHA III patients. For comparison, the probabilities of replacing a CRT-D lasting 5 years were 78 ± 8 and 59 ± 13%, respectively (both P < 0.001). CONCLUSION: Battery drain was mainly associated with pacing output in CRT-D devices and with the number of capacitor charges in both ICD and CRT-D devices. The longevity estimates provided by the devices were consistent and conservative. According to these estimates, among CRT-D recipients a low proportion of patients should require device replacement. |
format | Online Article Text |
id | pubmed-5006961 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | Oxford University Press |
record_format | MEDLINE/PubMed |
spelling | pubmed-50069612016-09-01 Battery drain in daily practice and medium-term projections on longevity of cardioverter-defibrillators: an analysis from a remote monitoring database Boriani, Giuseppe Ritter, Philippe Biffi, Mauro Ziacchi, Matteo Diemberger, Igor Martignani, Cristian Valzania, Cinzia Valsecchi, Sergio Padeletti, Luigi Gadler, Fredrik Europace CLINICAL RESEARCH AIMS: The longevity of generators is a crucial determinant of the cost-effectiveness of therapy with implantable cardioverter-defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-D). We evaluated the trend of device-measured residual battery capacity and longevity projections over 5-year follow-up. We also investigated possible factors associated with battery drain. METHODS AND RESULTS: Data from 4851 patients in the European LATITUDE(®) database who were followed up for a minimum of 3 years were analysed. The factors associated with battery drain (i.e. year-to-year decrease in residual battery capacity), and thus potentially impacting on device longevity, were mainly the pacing parameters in CRT-D devices and the number of shocks delivered and diverted in both ICD and CRT-D (all P < 0.01 on linear regression analysis). Over the first 5 years, the longevity estimates provided by devices showed low intra-patient variability and increased with time. The estimates exceeded 10 years for CRT-D and 13 and 12 years for single- and dual-chamber ICDs, respectively. In CRT-D patients, the expected patient age on replacement was 80 ± 12 years, and the expected probability of undergoing device replacement was 63 ± 13% for New York Heart Association (NYHA) II patients and 37 ± 16% for NYHA III patients. For comparison, the probabilities of replacing a CRT-D lasting 5 years were 78 ± 8 and 59 ± 13%, respectively (both P < 0.001). CONCLUSION: Battery drain was mainly associated with pacing output in CRT-D devices and with the number of capacitor charges in both ICD and CRT-D devices. The longevity estimates provided by the devices were consistent and conservative. According to these estimates, among CRT-D recipients a low proportion of patients should require device replacement. Oxford University Press 2016-09 2016-02-03 /pmc/articles/PMC5006961/ /pubmed/26847076 http://dx.doi.org/10.1093/europace/euv436 Text en © The Author 2016. Published by Oxford University Press on behalf of the European Society of Cardiology. http://creativecommons.org/licenses/by-nc/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is properly cited. For commercial re-use, please contact journals.permissions@oup.com |
spellingShingle | CLINICAL RESEARCH Boriani, Giuseppe Ritter, Philippe Biffi, Mauro Ziacchi, Matteo Diemberger, Igor Martignani, Cristian Valzania, Cinzia Valsecchi, Sergio Padeletti, Luigi Gadler, Fredrik Battery drain in daily practice and medium-term projections on longevity of cardioverter-defibrillators: an analysis from a remote monitoring database |
title | Battery drain in daily practice and medium-term projections on longevity of cardioverter-defibrillators: an analysis from a remote monitoring database |
title_full | Battery drain in daily practice and medium-term projections on longevity of cardioverter-defibrillators: an analysis from a remote monitoring database |
title_fullStr | Battery drain in daily practice and medium-term projections on longevity of cardioverter-defibrillators: an analysis from a remote monitoring database |
title_full_unstemmed | Battery drain in daily practice and medium-term projections on longevity of cardioverter-defibrillators: an analysis from a remote monitoring database |
title_short | Battery drain in daily practice and medium-term projections on longevity of cardioverter-defibrillators: an analysis from a remote monitoring database |
title_sort | battery drain in daily practice and medium-term projections on longevity of cardioverter-defibrillators: an analysis from a remote monitoring database |
topic | CLINICAL RESEARCH |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5006961/ https://www.ncbi.nlm.nih.gov/pubmed/26847076 http://dx.doi.org/10.1093/europace/euv436 |
work_keys_str_mv | AT borianigiuseppe batterydrainindailypracticeandmediumtermprojectionsonlongevityofcardioverterdefibrillatorsananalysisfromaremotemonitoringdatabase AT ritterphilippe batterydrainindailypracticeandmediumtermprojectionsonlongevityofcardioverterdefibrillatorsananalysisfromaremotemonitoringdatabase AT biffimauro batterydrainindailypracticeandmediumtermprojectionsonlongevityofcardioverterdefibrillatorsananalysisfromaremotemonitoringdatabase AT ziacchimatteo batterydrainindailypracticeandmediumtermprojectionsonlongevityofcardioverterdefibrillatorsananalysisfromaremotemonitoringdatabase AT diembergerigor batterydrainindailypracticeandmediumtermprojectionsonlongevityofcardioverterdefibrillatorsananalysisfromaremotemonitoringdatabase AT martignanicristian batterydrainindailypracticeandmediumtermprojectionsonlongevityofcardioverterdefibrillatorsananalysisfromaremotemonitoringdatabase AT valzaniacinzia batterydrainindailypracticeandmediumtermprojectionsonlongevityofcardioverterdefibrillatorsananalysisfromaremotemonitoringdatabase AT valsecchisergio batterydrainindailypracticeandmediumtermprojectionsonlongevityofcardioverterdefibrillatorsananalysisfromaremotemonitoringdatabase AT padelettiluigi batterydrainindailypracticeandmediumtermprojectionsonlongevityofcardioverterdefibrillatorsananalysisfromaremotemonitoringdatabase AT gadlerfredrik batterydrainindailypracticeandmediumtermprojectionsonlongevityofcardioverterdefibrillatorsananalysisfromaremotemonitoringdatabase |