Monitoring Radiofrequency Ablation Using Ultrasound Envelope Statistics and Shear Wave Elastography in the Periablation Period: An In Vitro Feasibility Study

Radiofrequency ablation (RFA) is a minimally invasive method for treating tumors. Shear wave elastography (SWE) has been widely applied in evaluating tissue stiffness and final ablation size after RFA. However, the usefulness of periablation SWE imaging in assessing RFA remains unclear. Therefore, t...

Descripción completa

Detalles Bibliográficos
Autores principales: Tsui, Po-Hsiang, Wang, Chiao-Yin, Zhou, Zhuhuang, Wan, Yung-Liang
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5014384/
https://www.ncbi.nlm.nih.gov/pubmed/27603012
http://dx.doi.org/10.1371/journal.pone.0162488
_version_ 1782452281897123840
author Tsui, Po-Hsiang
Wang, Chiao-Yin
Zhou, Zhuhuang
Wan, Yung-Liang
author_facet Tsui, Po-Hsiang
Wang, Chiao-Yin
Zhou, Zhuhuang
Wan, Yung-Liang
author_sort Tsui, Po-Hsiang
collection PubMed
description Radiofrequency ablation (RFA) is a minimally invasive method for treating tumors. Shear wave elastography (SWE) has been widely applied in evaluating tissue stiffness and final ablation size after RFA. However, the usefulness of periablation SWE imaging in assessing RFA remains unclear. Therefore, this study investigated the correlation between periablation SWE imaging and final ablation size. An in vitro porcine liver model was used for experimental validation (n = 36). During RFA with a power of 50 W, SWE images were collected using a clinical ultrasound system. To evaluate the effects of tissue temperature and gas bubbles during RFA, changes in the ablation temperature were recorded, and image echo patterns were measured using B-mode and ultrasound statistical parametric images. After RFA, the gross pathology of each tissue sample was compared with the region of change in the corresponding periablation SWE image. The experimental results showed that the tissue temperature at the ablation site varied between 70°C and 100°C. Hyperechoic regions and changes were observed in the echo amplitude distribution induced by gas bubbles. Under this condition, the confounding effects (including the temperature increase, tissue stiffness increase, and presence of gas bubbles) resulted in artifacts in the periablation SWE images, and the corresponding region correlated with the estimated final ablation size obtained from the gross pathology (r = 0.8). The findings confirm the feasibility of using periablation SWE imaging in assessing RFA.
format Online
Article
Text
id pubmed-5014384
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-50143842016-09-27 Monitoring Radiofrequency Ablation Using Ultrasound Envelope Statistics and Shear Wave Elastography in the Periablation Period: An In Vitro Feasibility Study Tsui, Po-Hsiang Wang, Chiao-Yin Zhou, Zhuhuang Wan, Yung-Liang PLoS One Research Article Radiofrequency ablation (RFA) is a minimally invasive method for treating tumors. Shear wave elastography (SWE) has been widely applied in evaluating tissue stiffness and final ablation size after RFA. However, the usefulness of periablation SWE imaging in assessing RFA remains unclear. Therefore, this study investigated the correlation between periablation SWE imaging and final ablation size. An in vitro porcine liver model was used for experimental validation (n = 36). During RFA with a power of 50 W, SWE images were collected using a clinical ultrasound system. To evaluate the effects of tissue temperature and gas bubbles during RFA, changes in the ablation temperature were recorded, and image echo patterns were measured using B-mode and ultrasound statistical parametric images. After RFA, the gross pathology of each tissue sample was compared with the region of change in the corresponding periablation SWE image. The experimental results showed that the tissue temperature at the ablation site varied between 70°C and 100°C. Hyperechoic regions and changes were observed in the echo amplitude distribution induced by gas bubbles. Under this condition, the confounding effects (including the temperature increase, tissue stiffness increase, and presence of gas bubbles) resulted in artifacts in the periablation SWE images, and the corresponding region correlated with the estimated final ablation size obtained from the gross pathology (r = 0.8). The findings confirm the feasibility of using periablation SWE imaging in assessing RFA. Public Library of Science 2016-09-07 /pmc/articles/PMC5014384/ /pubmed/27603012 http://dx.doi.org/10.1371/journal.pone.0162488 Text en © 2016 Tsui et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Tsui, Po-Hsiang
Wang, Chiao-Yin
Zhou, Zhuhuang
Wan, Yung-Liang
Monitoring Radiofrequency Ablation Using Ultrasound Envelope Statistics and Shear Wave Elastography in the Periablation Period: An In Vitro Feasibility Study
title Monitoring Radiofrequency Ablation Using Ultrasound Envelope Statistics and Shear Wave Elastography in the Periablation Period: An In Vitro Feasibility Study
title_full Monitoring Radiofrequency Ablation Using Ultrasound Envelope Statistics and Shear Wave Elastography in the Periablation Period: An In Vitro Feasibility Study
title_fullStr Monitoring Radiofrequency Ablation Using Ultrasound Envelope Statistics and Shear Wave Elastography in the Periablation Period: An In Vitro Feasibility Study
title_full_unstemmed Monitoring Radiofrequency Ablation Using Ultrasound Envelope Statistics and Shear Wave Elastography in the Periablation Period: An In Vitro Feasibility Study
title_short Monitoring Radiofrequency Ablation Using Ultrasound Envelope Statistics and Shear Wave Elastography in the Periablation Period: An In Vitro Feasibility Study
title_sort monitoring radiofrequency ablation using ultrasound envelope statistics and shear wave elastography in the periablation period: an in vitro feasibility study
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5014384/
https://www.ncbi.nlm.nih.gov/pubmed/27603012
http://dx.doi.org/10.1371/journal.pone.0162488
work_keys_str_mv AT tsuipohsiang monitoringradiofrequencyablationusingultrasoundenvelopestatisticsandshearwaveelastographyintheperiablationperiodaninvitrofeasibilitystudy
AT wangchiaoyin monitoringradiofrequencyablationusingultrasoundenvelopestatisticsandshearwaveelastographyintheperiablationperiodaninvitrofeasibilitystudy
AT zhouzhuhuang monitoringradiofrequencyablationusingultrasoundenvelopestatisticsandshearwaveelastographyintheperiablationperiodaninvitrofeasibilitystudy
AT wanyungliang monitoringradiofrequencyablationusingultrasoundenvelopestatisticsandshearwaveelastographyintheperiablationperiodaninvitrofeasibilitystudy