Cargando…

Implantation of single lead cardioverter defibrillator with floating atrial sensing dipole in a pregnant patient without using fluoroscopy

In this case report, we look into the implant procedure of a single-lead ICD with floating atrial sensing dipole in a pregnant woman, without using fluoroscopy. This system benefits the proper positioning of the lead. This is possible thanks to the simultaneous display of both the atrial and ventric...

Descripción completa

Detalles Bibliográficos
Autores principales: Quartieri, Fabio, Giacopelli, Daniele, Iori, Matteo, Bottoni, Nicola
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5031860/
https://www.ncbi.nlm.nih.gov/pubmed/27676164
http://dx.doi.org/10.1016/j.ipej.2016.08.004