Toric intraocular lens orientation and residual refractive astigmatism: an analysis

PURPOSE: To analyze intraocular lens (IOL) orientation data from an online toric back-calculator (astigmatismfix.com) for determining if differences were apparent by lens type. METHODS: A retrospective review of astigmatismfix.com toric back-calculations that included IOL identification and intended...

Descripción completa

Detalles Bibliográficos
Autores principales: Potvin, Rick, Kramer, Brent A, Hardten, David R, Berdahl, John P
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove Medical Press 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5036610/
https://www.ncbi.nlm.nih.gov/pubmed/27703323
http://dx.doi.org/10.2147/OPTH.S114118
Descripción
Sumario:PURPOSE: To analyze intraocular lens (IOL) orientation data from an online toric back-calculator (astigmatismfix.com) for determining if differences were apparent by lens type. METHODS: A retrospective review of astigmatismfix.com toric back-calculations that included IOL identification and intended orientation axis. RESULTS: Of 12,812 total validated calculation records, 8,229 included intended orientation and lens identification data. Of the latter, 5,674 calculations (69%) involved lenses oriented 5° or more from their intended position. Using estimated toric lens usage data, the percentage of lenses with orientation ≥5° from intended was 0.89% overall, but the percentage varied significantly between specific toric lens brands (P<0.05). The percentage of back-calculations related to lenses that were not oriented as intended was also statistically significantly different by lens brand (P<0.05). When IOLs were misoriented, they were significantly more likely to be misoriented in a counterclockwise direction (P<0.05). This was found to be due to a bias toward counterclockwise orientation observed with one specific brand, a bias that was not observed with the other three brands analyzed here. CONCLUSION: The percentage of eyes with lens orientation ≥5° from intended in the Toric Results Analyzer data set was <1% of toric IOLs in general, with the relative percentage of Tecnis(®) Toric IOLs significantly higher than AcrySof(®) Toric IOLs. Both of these had higher rates than the Staar(®) Toric and Trulign(®) Toric lenses, with the availability of higher Tecnis and AcrySof cylinder powers a likely contributing factor. The AcrySof Toric IOL appears to be less likely than the Tecnis Toric IOL to cause residual astigmatism as a result of misorientation. The Tecnis Toric IOL appears more likely to be misoriented in a counterclockwise direction; no such bias was observed with the AcrySof Toric, the Trulign(®) Toric, or the Staar Toric IOLs.