Cargando…

Investigation into Deep Brain Stimulation Lead Designs: A Patient-Specific Simulation Study

New deep brain stimulation (DBS) electrode designs offer operation in voltage and current mode and capability to steer the electric field (EF). The aim of the study was to compare the EF distributions of four DBS leads at equivalent amplitudes (3 V and 3.4 mA). Finite element method (FEM) simulation...

Descripción completa

Detalles Bibliográficos
Autores principales: Alonso, Fabiola, Latorre, Malcolm A., Göransson, Nathanael, Zsigmond, Peter, Wårdell, Karin
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5039468/
https://www.ncbi.nlm.nih.gov/pubmed/27618109
http://dx.doi.org/10.3390/brainsci6030039