Precise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using Gafchromic™ EBT3 films

PURPOSE: The purpose of this study is to evaluate the dosimetric uncertainty associated with Gafchromic™ (EBT3) films and establish a practical and efficient film dosimetry protocol for Stereotactic Radiosurgery (SRS) and Stereotactic Body Radiotherapy (SBRT). METHOD AND MATERIALS: EBT3 films were i...

Descripción completa

Detalles Bibliográficos
Autores principales: Wen, Ning, Lu, Siming, Kim, Jinkoo, Qin, Yujiao, Huang, Yimei, Zhao, Bo, Liu, Chang, Chetty, Indrin J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5050597/
https://www.ncbi.nlm.nih.gov/pubmed/27716323
http://dx.doi.org/10.1186/s13014-016-0709-4
_version_ 1782457912462934016
author Wen, Ning
Lu, Siming
Kim, Jinkoo
Qin, Yujiao
Huang, Yimei
Zhao, Bo
Liu, Chang
Chetty, Indrin J.
author_facet Wen, Ning
Lu, Siming
Kim, Jinkoo
Qin, Yujiao
Huang, Yimei
Zhao, Bo
Liu, Chang
Chetty, Indrin J.
author_sort Wen, Ning
collection PubMed
description PURPOSE: The purpose of this study is to evaluate the dosimetric uncertainty associated with Gafchromic™ (EBT3) films and establish a practical and efficient film dosimetry protocol for Stereotactic Radiosurgery (SRS) and Stereotactic Body Radiotherapy (SBRT). METHOD AND MATERIALS: EBT3 films were irradiated at each of seven different dose levels between 1 and 15 Gy with open fields and standard deviations of dose maps were calculated at each color channel for evaluation. A scanner non-uniform response correction map was built by registering and comparing film doses to the reference ion chamber array-based dose map delivered with the same doses. To determine the temporal dependence of EBT3 films, the average correction factors of different dose levels as a function of time were evaluated up to 4 days after irradiation. An integrated film dosimetry protocol was developed for dose calibration, calibration curve fitting, dose mapping, and profile/gamma analysis. Patient specific quality assurance (PSQA) was performed for 83 SRS/SBRT treatment plans, and analysis of the measurements and calculations are presented here. RESULTS: The scanner response varied within 1 % for the field sizes less than 5 × 5 cm(2), and up to 5 % for the field sizes of 10 × 10 cm(2) for all color channels. The scanner correction method was able to remove visually evident, irregular detector responses for larger field sizes. The dose response of the film changed rapidly (~10 %) in the first two hours and became smooth plateaued afterwards, ~3 % change between 2 and 24 h. The uncertainties were approximately 1.5, 1.7 and 4.8 % over the dose range of 3~15 Gy for the red, green and blue channels. The green channel showed very high sensitivity and low uncertainty in the dose range between 10 and 15 Gy, which is suitable for SRS/SBRT commissioning and PSQA. The difference between the calculated dose and measured dose of ion chamber measurement at isocenter was −0.64 ± 2.02 for all plans, corresponding to a 95 % confidence interval of (−1.09, −0.26). The percentage of points passing the 3 %/1 mm gamma criteria in absolute dose, averaged over all tests was 95.0 ± 4.2. CONCLUSION: We have developed the EBT3 films based dosimetry protocol to obtain absolute dose values. The overall uncertainty has been established to be 1.5 % for SRS and SBRT PSQA.
format Online
Article
Text
id pubmed-5050597
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-50505972016-10-05 Precise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using Gafchromic™ EBT3 films Wen, Ning Lu, Siming Kim, Jinkoo Qin, Yujiao Huang, Yimei Zhao, Bo Liu, Chang Chetty, Indrin J. Radiat Oncol Research PURPOSE: The purpose of this study is to evaluate the dosimetric uncertainty associated with Gafchromic™ (EBT3) films and establish a practical and efficient film dosimetry protocol for Stereotactic Radiosurgery (SRS) and Stereotactic Body Radiotherapy (SBRT). METHOD AND MATERIALS: EBT3 films were irradiated at each of seven different dose levels between 1 and 15 Gy with open fields and standard deviations of dose maps were calculated at each color channel for evaluation. A scanner non-uniform response correction map was built by registering and comparing film doses to the reference ion chamber array-based dose map delivered with the same doses. To determine the temporal dependence of EBT3 films, the average correction factors of different dose levels as a function of time were evaluated up to 4 days after irradiation. An integrated film dosimetry protocol was developed for dose calibration, calibration curve fitting, dose mapping, and profile/gamma analysis. Patient specific quality assurance (PSQA) was performed for 83 SRS/SBRT treatment plans, and analysis of the measurements and calculations are presented here. RESULTS: The scanner response varied within 1 % for the field sizes less than 5 × 5 cm(2), and up to 5 % for the field sizes of 10 × 10 cm(2) for all color channels. The scanner correction method was able to remove visually evident, irregular detector responses for larger field sizes. The dose response of the film changed rapidly (~10 %) in the first two hours and became smooth plateaued afterwards, ~3 % change between 2 and 24 h. The uncertainties were approximately 1.5, 1.7 and 4.8 % over the dose range of 3~15 Gy for the red, green and blue channels. The green channel showed very high sensitivity and low uncertainty in the dose range between 10 and 15 Gy, which is suitable for SRS/SBRT commissioning and PSQA. The difference between the calculated dose and measured dose of ion chamber measurement at isocenter was −0.64 ± 2.02 for all plans, corresponding to a 95 % confidence interval of (−1.09, −0.26). The percentage of points passing the 3 %/1 mm gamma criteria in absolute dose, averaged over all tests was 95.0 ± 4.2. CONCLUSION: We have developed the EBT3 films based dosimetry protocol to obtain absolute dose values. The overall uncertainty has been established to be 1.5 % for SRS and SBRT PSQA. BioMed Central 2016-10-04 /pmc/articles/PMC5050597/ /pubmed/27716323 http://dx.doi.org/10.1186/s13014-016-0709-4 Text en © The Author(s). 2016 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research
Wen, Ning
Lu, Siming
Kim, Jinkoo
Qin, Yujiao
Huang, Yimei
Zhao, Bo
Liu, Chang
Chetty, Indrin J.
Precise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using Gafchromic™ EBT3 films
title Precise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using Gafchromic™ EBT3 films
title_full Precise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using Gafchromic™ EBT3 films
title_fullStr Precise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using Gafchromic™ EBT3 films
title_full_unstemmed Precise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using Gafchromic™ EBT3 films
title_short Precise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using Gafchromic™ EBT3 films
title_sort precise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using gafchromic™ ebt3 films
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5050597/
https://www.ncbi.nlm.nih.gov/pubmed/27716323
http://dx.doi.org/10.1186/s13014-016-0709-4
work_keys_str_mv AT wenning precisefilmdosimetryforstereotacticradiosurgeryandstereotacticbodyradiotherapyqualityassuranceusinggafchromicebt3films
AT lusiming precisefilmdosimetryforstereotacticradiosurgeryandstereotacticbodyradiotherapyqualityassuranceusinggafchromicebt3films
AT kimjinkoo precisefilmdosimetryforstereotacticradiosurgeryandstereotacticbodyradiotherapyqualityassuranceusinggafchromicebt3films
AT qinyujiao precisefilmdosimetryforstereotacticradiosurgeryandstereotacticbodyradiotherapyqualityassuranceusinggafchromicebt3films
AT huangyimei precisefilmdosimetryforstereotacticradiosurgeryandstereotacticbodyradiotherapyqualityassuranceusinggafchromicebt3films
AT zhaobo precisefilmdosimetryforstereotacticradiosurgeryandstereotacticbodyradiotherapyqualityassuranceusinggafchromicebt3films
AT liuchang precisefilmdosimetryforstereotacticradiosurgeryandstereotacticbodyradiotherapyqualityassuranceusinggafchromicebt3films
AT chettyindrinj precisefilmdosimetryforstereotacticradiosurgeryandstereotacticbodyradiotherapyqualityassuranceusinggafchromicebt3films