Fluid responsiveness prediction using Vigileo FloTrac measured cardiac output changes during passive leg raise test
BACKGROUND: Passive leg raising (PLR) is a so called self-volume challenge used to test for fluid responsiveness. Changes in cardiac output (CO) or stroke volume (SV) measured during PLR are used to predict the need for subsequent fluid loading. This requires a device that can measure CO changes rap...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5052799/ https://www.ncbi.nlm.nih.gov/pubmed/27721980 http://dx.doi.org/10.1186/s40560-016-0188-6 |
_version_ | 1782458289552883712 |
---|---|
author | Krige, Anton Bland, Martin Fanshawe, Thomas |
author_facet | Krige, Anton Bland, Martin Fanshawe, Thomas |
author_sort | Krige, Anton |
collection | PubMed |
description | BACKGROUND: Passive leg raising (PLR) is a so called self-volume challenge used to test for fluid responsiveness. Changes in cardiac output (CO) or stroke volume (SV) measured during PLR are used to predict the need for subsequent fluid loading. This requires a device that can measure CO changes rapidly. The Vigileo™ monitor, using third-generation software, allows continuous CO monitoring. The aim of this study was to compare changes in CO (measured with the Vigileo device) during a PLR manoeuvre to calculate the accuracy for predicting fluid responsiveness. METHODS: This is a prospective study in a 20-bedded mixed general critical care unit in a large non-university regional referral hospital. Fluid responders were defined as having an increase in CO of greater than 15 % following a fluid challenge. Patients meeting the criteria for circulatory shock with a Vigileo™ monitor (Vigileo™; FloTrac; Edwards™; Lifesciences, Irvine, CA, USA) already in situ, and assessed as requiring volume expansion by the clinical team based on clinical criteria, were included. All patients underwent a PLR manoeuvre followed by a fluid challenge. RESULTS: Data was collected and analysed on stroke volume variation (SVV) at baseline and CO and SVV changes during the PLR manoeuvre and following a subsequent fluid challenge in 33 patients. The majority had septic shock. Patient characteristics, baseline haemodynamic variables and baseline vasoactive infusion requirements were similar between fluid responders (10 patients) and non-responders (23 patients). Peak increase in CO occurred within 120 s during the PLR in all cases. Using an optimal cut point of 9 % increase in CO during the PLR produced an area under the receiver operating characteristic curve of 0.85 (95 % CI 0.63 to 1.00) with a sensitivity of 80 % (95 % CI 44 to 96 %) and a specificity of 91 % (95 % CI 70 to 98 %). CONCLUSIONS: CO changes measured by the Vigileo™ monitor using third-generation software during a PLR test predict fluid responsiveness in mixed medical and surgical patients with vasopressor-dependent circulatory shock. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1186/s40560-016-0188-6) contains supplementary material, which is available to authorized users. |
format | Online Article Text |
id | pubmed-5052799 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-50527992016-10-07 Fluid responsiveness prediction using Vigileo FloTrac measured cardiac output changes during passive leg raise test Krige, Anton Bland, Martin Fanshawe, Thomas J Intensive Care Research BACKGROUND: Passive leg raising (PLR) is a so called self-volume challenge used to test for fluid responsiveness. Changes in cardiac output (CO) or stroke volume (SV) measured during PLR are used to predict the need for subsequent fluid loading. This requires a device that can measure CO changes rapidly. The Vigileo™ monitor, using third-generation software, allows continuous CO monitoring. The aim of this study was to compare changes in CO (measured with the Vigileo device) during a PLR manoeuvre to calculate the accuracy for predicting fluid responsiveness. METHODS: This is a prospective study in a 20-bedded mixed general critical care unit in a large non-university regional referral hospital. Fluid responders were defined as having an increase in CO of greater than 15 % following a fluid challenge. Patients meeting the criteria for circulatory shock with a Vigileo™ monitor (Vigileo™; FloTrac; Edwards™; Lifesciences, Irvine, CA, USA) already in situ, and assessed as requiring volume expansion by the clinical team based on clinical criteria, were included. All patients underwent a PLR manoeuvre followed by a fluid challenge. RESULTS: Data was collected and analysed on stroke volume variation (SVV) at baseline and CO and SVV changes during the PLR manoeuvre and following a subsequent fluid challenge in 33 patients. The majority had septic shock. Patient characteristics, baseline haemodynamic variables and baseline vasoactive infusion requirements were similar between fluid responders (10 patients) and non-responders (23 patients). Peak increase in CO occurred within 120 s during the PLR in all cases. Using an optimal cut point of 9 % increase in CO during the PLR produced an area under the receiver operating characteristic curve of 0.85 (95 % CI 0.63 to 1.00) with a sensitivity of 80 % (95 % CI 44 to 96 %) and a specificity of 91 % (95 % CI 70 to 98 %). CONCLUSIONS: CO changes measured by the Vigileo™ monitor using third-generation software during a PLR test predict fluid responsiveness in mixed medical and surgical patients with vasopressor-dependent circulatory shock. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1186/s40560-016-0188-6) contains supplementary material, which is available to authorized users. BioMed Central 2016-10-06 /pmc/articles/PMC5052799/ /pubmed/27721980 http://dx.doi.org/10.1186/s40560-016-0188-6 Text en © The Author(s). 2016 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
spellingShingle | Research Krige, Anton Bland, Martin Fanshawe, Thomas Fluid responsiveness prediction using Vigileo FloTrac measured cardiac output changes during passive leg raise test |
title | Fluid responsiveness prediction using Vigileo FloTrac measured cardiac output changes during passive leg raise test |
title_full | Fluid responsiveness prediction using Vigileo FloTrac measured cardiac output changes during passive leg raise test |
title_fullStr | Fluid responsiveness prediction using Vigileo FloTrac measured cardiac output changes during passive leg raise test |
title_full_unstemmed | Fluid responsiveness prediction using Vigileo FloTrac measured cardiac output changes during passive leg raise test |
title_short | Fluid responsiveness prediction using Vigileo FloTrac measured cardiac output changes during passive leg raise test |
title_sort | fluid responsiveness prediction using vigileo flotrac measured cardiac output changes during passive leg raise test |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5052799/ https://www.ncbi.nlm.nih.gov/pubmed/27721980 http://dx.doi.org/10.1186/s40560-016-0188-6 |
work_keys_str_mv | AT krigeanton fluidresponsivenesspredictionusingvigileoflotracmeasuredcardiacoutputchangesduringpassivelegraisetest AT blandmartin fluidresponsivenesspredictionusingvigileoflotracmeasuredcardiacoutputchangesduringpassivelegraisetest AT fanshawethomas fluidresponsivenesspredictionusingvigileoflotracmeasuredcardiacoutputchangesduringpassivelegraisetest |