Cargando…

Duration of Systole and Diastole for Hydrodynamic Testing of Prosthetic Heart Valves: Comparison Between ISO 5840 Standards and in vivo Studies

OBJECTIVE: To complement the ISO 5840 standards concerning the duration of left ventricular systole and diastole as a function of changes in heart rates according to in vivo studies from the physiologic literature review. METHODS: The systolic and diastolic durations from three in vivo studies were...

Descripción completa

Detalles Bibliográficos
Autores principales: Bazan, Ovandir, Ortiz, Jayme Pinto
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Sociedade Brasileira de Cirurgia Cardiovascular 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5062727/
https://www.ncbi.nlm.nih.gov/pubmed/27556318
http://dx.doi.org/10.5935/1678-9741.20160036
_version_ 1782459840852918272
author Bazan, Ovandir
Ortiz, Jayme Pinto
author_facet Bazan, Ovandir
Ortiz, Jayme Pinto
author_sort Bazan, Ovandir
collection PubMed
description OBJECTIVE: To complement the ISO 5840 standards concerning the duration of left ventricular systole and diastole as a function of changes in heart rates according to in vivo studies from the physiologic literature review. METHODS: The systolic and diastolic durations from three in vivo studies were compared with the durations of systole proposed by the ISO 5840:2010 and ISO 5840-2:2015 for hydrodynamic performance assessment of prosthetic heart valves. RESULTS: Based on the in vivo studies analyzed, the systolic durations proposed by the ISO 5840 standard seemed consistent for 45 and 120 beats per minute (bpm), and showed diverse results for the 70 bpm condition. CONCLUSION: Information on the realistic validation of the operation of left ventricular models for different heart rates were obtained.
format Online
Article
Text
id pubmed-5062727
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Sociedade Brasileira de Cirurgia Cardiovascular
record_format MEDLINE/PubMed
spelling pubmed-50627272016-10-19 Duration of Systole and Diastole for Hydrodynamic Testing of Prosthetic Heart Valves: Comparison Between ISO 5840 Standards and in vivo Studies Bazan, Ovandir Ortiz, Jayme Pinto Braz J Cardiovasc Surg Special Article OBJECTIVE: To complement the ISO 5840 standards concerning the duration of left ventricular systole and diastole as a function of changes in heart rates according to in vivo studies from the physiologic literature review. METHODS: The systolic and diastolic durations from three in vivo studies were compared with the durations of systole proposed by the ISO 5840:2010 and ISO 5840-2:2015 for hydrodynamic performance assessment of prosthetic heart valves. RESULTS: Based on the in vivo studies analyzed, the systolic durations proposed by the ISO 5840 standard seemed consistent for 45 and 120 beats per minute (bpm), and showed diverse results for the 70 bpm condition. CONCLUSION: Information on the realistic validation of the operation of left ventricular models for different heart rates were obtained. Sociedade Brasileira de Cirurgia Cardiovascular 2016 /pmc/articles/PMC5062727/ /pubmed/27556318 http://dx.doi.org/10.5935/1678-9741.20160036 Text en http://creativecommons.org/licenses/by/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Special Article
Bazan, Ovandir
Ortiz, Jayme Pinto
Duration of Systole and Diastole for Hydrodynamic Testing of Prosthetic Heart Valves: Comparison Between ISO 5840 Standards and in vivo Studies
title Duration of Systole and Diastole for Hydrodynamic Testing of Prosthetic Heart Valves: Comparison Between ISO 5840 Standards and in vivo Studies
title_full Duration of Systole and Diastole for Hydrodynamic Testing of Prosthetic Heart Valves: Comparison Between ISO 5840 Standards and in vivo Studies
title_fullStr Duration of Systole and Diastole for Hydrodynamic Testing of Prosthetic Heart Valves: Comparison Between ISO 5840 Standards and in vivo Studies
title_full_unstemmed Duration of Systole and Diastole for Hydrodynamic Testing of Prosthetic Heart Valves: Comparison Between ISO 5840 Standards and in vivo Studies
title_short Duration of Systole and Diastole for Hydrodynamic Testing of Prosthetic Heart Valves: Comparison Between ISO 5840 Standards and in vivo Studies
title_sort duration of systole and diastole for hydrodynamic testing of prosthetic heart valves: comparison between iso 5840 standards and in vivo studies
topic Special Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5062727/
https://www.ncbi.nlm.nih.gov/pubmed/27556318
http://dx.doi.org/10.5935/1678-9741.20160036
work_keys_str_mv AT bazanovandir durationofsystoleanddiastoleforhydrodynamictestingofprostheticheartvalvescomparisonbetweeniso5840standardsandinvivostudies
AT ortizjaymepinto durationofsystoleanddiastoleforhydrodynamictestingofprostheticheartvalvescomparisonbetweeniso5840standardsandinvivostudies