Experimental Validation of a Cardiac Simulator for in vitro Evaluation of Prosthetic Heart Valves

OBJECTIVE: This work describes the experimental validation of a cardiac simulator for three heart rates (60, 80 and 100 beats per minute), under physiological conditions, as a suitable environment for prosthetic heart valves testing in the mitral or aortic position. METHODS: In the experiment, an ao...

Descripción completa

Detalles Bibliográficos
Autores principales: Bazan, Ovandir, Ortiz, Jayme Pinto
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Sociedade Brasileira de Cirurgia Cardiovascular 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5062730/
https://www.ncbi.nlm.nih.gov/pubmed/27556315
http://dx.doi.org/10.5935/1678-9741.20160041
_version_ 1782459841603698688
author Bazan, Ovandir
Ortiz, Jayme Pinto
author_facet Bazan, Ovandir
Ortiz, Jayme Pinto
author_sort Bazan, Ovandir
collection PubMed
description OBJECTIVE: This work describes the experimental validation of a cardiac simulator for three heart rates (60, 80 and 100 beats per minute), under physiological conditions, as a suitable environment for prosthetic heart valves testing in the mitral or aortic position. METHODS: In the experiment, an aortic bileaflet mechanical valve and a mitral bioprosthesis were employed in the left ventricular model. A test fluid of 47.6% by volume of glycerin solution in water at 36.5ºC was used as blood analogue fluid. A supervisory control and data acquisition system implemented previously in LabVIEW was applied to induce the ventricular operation and to acquire the ventricular signals. The parameters of the left ventricular model operation were based on in vivo and in vitro data. The waves of ventricular and systemic pressures, aortic flow, stroke volume, among others, were acquired while manual adjustments in the arterial impedance model were also established. RESULTS: The acquired waves showed good results concerning some in vivo data and requirements from the ISO 5840 standard. CONCLUSION: The experimental validation was performed, allowing, in future studies, characterizing the hydrodynamic performance of prosthetic heart valves.
format Online
Article
Text
id pubmed-5062730
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Sociedade Brasileira de Cirurgia Cardiovascular
record_format MEDLINE/PubMed
spelling pubmed-50627302016-10-19 Experimental Validation of a Cardiac Simulator for in vitro Evaluation of Prosthetic Heart Valves Bazan, Ovandir Ortiz, Jayme Pinto Braz J Cardiovasc Surg Original Articles OBJECTIVE: This work describes the experimental validation of a cardiac simulator for three heart rates (60, 80 and 100 beats per minute), under physiological conditions, as a suitable environment for prosthetic heart valves testing in the mitral or aortic position. METHODS: In the experiment, an aortic bileaflet mechanical valve and a mitral bioprosthesis were employed in the left ventricular model. A test fluid of 47.6% by volume of glycerin solution in water at 36.5ºC was used as blood analogue fluid. A supervisory control and data acquisition system implemented previously in LabVIEW was applied to induce the ventricular operation and to acquire the ventricular signals. The parameters of the left ventricular model operation were based on in vivo and in vitro data. The waves of ventricular and systemic pressures, aortic flow, stroke volume, among others, were acquired while manual adjustments in the arterial impedance model were also established. RESULTS: The acquired waves showed good results concerning some in vivo data and requirements from the ISO 5840 standard. CONCLUSION: The experimental validation was performed, allowing, in future studies, characterizing the hydrodynamic performance of prosthetic heart valves. Sociedade Brasileira de Cirurgia Cardiovascular 2016 /pmc/articles/PMC5062730/ /pubmed/27556315 http://dx.doi.org/10.5935/1678-9741.20160041 Text en http://creativecommons.org/licenses/by/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Articles
Bazan, Ovandir
Ortiz, Jayme Pinto
Experimental Validation of a Cardiac Simulator for in vitro Evaluation of Prosthetic Heart Valves
title Experimental Validation of a Cardiac Simulator for in vitro Evaluation of Prosthetic Heart Valves
title_full Experimental Validation of a Cardiac Simulator for in vitro Evaluation of Prosthetic Heart Valves
title_fullStr Experimental Validation of a Cardiac Simulator for in vitro Evaluation of Prosthetic Heart Valves
title_full_unstemmed Experimental Validation of a Cardiac Simulator for in vitro Evaluation of Prosthetic Heart Valves
title_short Experimental Validation of a Cardiac Simulator for in vitro Evaluation of Prosthetic Heart Valves
title_sort experimental validation of a cardiac simulator for in vitro evaluation of prosthetic heart valves
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5062730/
https://www.ncbi.nlm.nih.gov/pubmed/27556315
http://dx.doi.org/10.5935/1678-9741.20160041
work_keys_str_mv AT bazanovandir experimentalvalidationofacardiacsimulatorforinvitroevaluationofprostheticheartvalves
AT ortizjaymepinto experimentalvalidationofacardiacsimulatorforinvitroevaluationofprostheticheartvalves