Cargando…

Effectiveness of the surgical torque limiter: a model comparing drill- and hand-based screw insertion into locking plates

BACKGROUND: The objective of this study is to analyse the effectiveness of the surgical torque limiter during operative use. The study also investigates the potential differences in torque between hand and drill-based screw insertion into locking plates using a standardised torque limiter. METHODS:...

Descripción completa

Detalles Bibliográficos
Autores principales: Ioannou, Christopher, Knight, Matthew, Daniele, Luca, Flueckiger, Lee, Tan, Ezekiel S. L.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5067914/
https://www.ncbi.nlm.nih.gov/pubmed/27751170
http://dx.doi.org/10.1186/s13018-016-0458-y
_version_ 1782460731337211904
author Ioannou, Christopher
Knight, Matthew
Daniele, Luca
Flueckiger, Lee
Tan, Ezekiel S. L.
author_facet Ioannou, Christopher
Knight, Matthew
Daniele, Luca
Flueckiger, Lee
Tan, Ezekiel S. L.
author_sort Ioannou, Christopher
collection PubMed
description BACKGROUND: The objective of this study is to analyse the effectiveness of the surgical torque limiter during operative use. The study also investigates the potential differences in torque between hand and drill-based screw insertion into locking plates using a standardised torque limiter. METHODS: Torque for both hand and power screw insertion was measured through a load cell, registering 6.66 points per second. This was performed in a controlled environment using synthetic bone, a locking plate and locking screws to simulate plate fixation. Screws were inserted by hand and by drill with torque values measured. RESULTS: The surgical torque limiter (1.5 Nm) was effective as the highest recorded reading in the study was 1.409 Nm. Comparatively, there is a statistically significant difference between screw insertion methods. Torque produced for manually driven screw insertion into locking plates was 1.289 Nm (95 % CI 1.269–1.308) with drill-powered screw insertion at 0.740 Nm (95 % CI 0.723–0.757). CONCLUSIONS: The surgical torque limiter proved to be effective as per product specifications. Screws inserted under power produce significantly less torque when compared to manual insertion by hand. This is likely related to the mechanism of the torque limiter when being used at higher speeds for which it was designed. We conclude that screws may be inserted using power to the plate with the addition of a torque limiter. It is recommended that all screws inserted by drill be hand tightened to achieve adequate torque values.
format Online
Article
Text
id pubmed-5067914
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-50679142016-10-24 Effectiveness of the surgical torque limiter: a model comparing drill- and hand-based screw insertion into locking plates Ioannou, Christopher Knight, Matthew Daniele, Luca Flueckiger, Lee Tan, Ezekiel S. L. J Orthop Surg Res Research Article BACKGROUND: The objective of this study is to analyse the effectiveness of the surgical torque limiter during operative use. The study also investigates the potential differences in torque between hand and drill-based screw insertion into locking plates using a standardised torque limiter. METHODS: Torque for both hand and power screw insertion was measured through a load cell, registering 6.66 points per second. This was performed in a controlled environment using synthetic bone, a locking plate and locking screws to simulate plate fixation. Screws were inserted by hand and by drill with torque values measured. RESULTS: The surgical torque limiter (1.5 Nm) was effective as the highest recorded reading in the study was 1.409 Nm. Comparatively, there is a statistically significant difference between screw insertion methods. Torque produced for manually driven screw insertion into locking plates was 1.289 Nm (95 % CI 1.269–1.308) with drill-powered screw insertion at 0.740 Nm (95 % CI 0.723–0.757). CONCLUSIONS: The surgical torque limiter proved to be effective as per product specifications. Screws inserted under power produce significantly less torque when compared to manual insertion by hand. This is likely related to the mechanism of the torque limiter when being used at higher speeds for which it was designed. We conclude that screws may be inserted using power to the plate with the addition of a torque limiter. It is recommended that all screws inserted by drill be hand tightened to achieve adequate torque values. BioMed Central 2016-10-17 /pmc/articles/PMC5067914/ /pubmed/27751170 http://dx.doi.org/10.1186/s13018-016-0458-y Text en © The Author(s). 2016 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research Article
Ioannou, Christopher
Knight, Matthew
Daniele, Luca
Flueckiger, Lee
Tan, Ezekiel S. L.
Effectiveness of the surgical torque limiter: a model comparing drill- and hand-based screw insertion into locking plates
title Effectiveness of the surgical torque limiter: a model comparing drill- and hand-based screw insertion into locking plates
title_full Effectiveness of the surgical torque limiter: a model comparing drill- and hand-based screw insertion into locking plates
title_fullStr Effectiveness of the surgical torque limiter: a model comparing drill- and hand-based screw insertion into locking plates
title_full_unstemmed Effectiveness of the surgical torque limiter: a model comparing drill- and hand-based screw insertion into locking plates
title_short Effectiveness of the surgical torque limiter: a model comparing drill- and hand-based screw insertion into locking plates
title_sort effectiveness of the surgical torque limiter: a model comparing drill- and hand-based screw insertion into locking plates
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5067914/
https://www.ncbi.nlm.nih.gov/pubmed/27751170
http://dx.doi.org/10.1186/s13018-016-0458-y
work_keys_str_mv AT ioannouchristopher effectivenessofthesurgicaltorquelimiteramodelcomparingdrillandhandbasedscrewinsertionintolockingplates
AT knightmatthew effectivenessofthesurgicaltorquelimiteramodelcomparingdrillandhandbasedscrewinsertionintolockingplates
AT danieleluca effectivenessofthesurgicaltorquelimiteramodelcomparingdrillandhandbasedscrewinsertionintolockingplates
AT flueckigerlee effectivenessofthesurgicaltorquelimiteramodelcomparingdrillandhandbasedscrewinsertionintolockingplates
AT tanezekielsl effectivenessofthesurgicaltorquelimiteramodelcomparingdrillandhandbasedscrewinsertionintolockingplates