A Case of Implant Failure in Partial Wrist Fusion Applying Magnesium-Based Headless Bone Screws

This article presents a case of implant failure resulting in mechanical instability of a scaphotrapezotrapezoideal arthrodesis using magnesium-based headless bone screws. During revision surgery osteolysis surrounding the screws was observed as well as degraded screw threads already in existence at...

Descripción completa

Detalles Bibliográficos
Autores principales: Wichelhaus, Alice, Emmerich, Judith, Mittlmeier, Thomas
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5069367/
https://www.ncbi.nlm.nih.gov/pubmed/27800201
http://dx.doi.org/10.1155/2016/7049130
Descripción
Sumario:This article presents a case of implant failure resulting in mechanical instability of a scaphotrapezotrapezoideal arthrodesis using magnesium-based headless bone screws. During revision surgery osteolysis surrounding the screws was observed as well as degraded screw threads already in existence at 6 weeks after implantation. The supposed osseous integration attributed to magnesium-based screws could not be reproduced in this particular case. Thus, it can be reasoned that the use of magnesium-based screws for partial wrist arthrodesis cannot be encouraged, at least not in dual use.