Comparison and verification of the reconstruction method of the catheters for interstitial brachytherapy

PURPOSE: 1. Comparison and verification of accuracy of the implant reconstruction method based on images from IBU and CT. 2. Estimation of influence of the implant reconstruction method on dose disposition in selected reference points. MATERIAL AND METHODS: Paraffin-wax phantom with three catheters,...

Descripción completa

Detalles Bibliográficos
Autores principales: Blasiak, Barbara, Zwierzchowski, Grzegorz, Skowronek, Janusz, Stefaniak, Patrycja, Dymnicka, Magdalena
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Termedia Publishing House 2009
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5080396/
https://www.ncbi.nlm.nih.gov/pubmed/27799955
_version_ 1782462703918383104
author Blasiak, Barbara
Zwierzchowski, Grzegorz
Skowronek, Janusz
Stefaniak, Patrycja
Dymnicka, Magdalena
author_facet Blasiak, Barbara
Zwierzchowski, Grzegorz
Skowronek, Janusz
Stefaniak, Patrycja
Dymnicka, Magdalena
author_sort Blasiak, Barbara
collection PubMed
description PURPOSE: 1. Comparison and verification of accuracy of the implant reconstruction method based on images from IBU and CT. 2. Estimation of influence of the implant reconstruction method on dose disposition in selected reference points. MATERIAL AND METHODS: Paraffin-wax phantom with three catheters, central marker and control point were prepared. IBU unit were used for obtaining two series of images for reconstruction. The Earth magnetic field correction algorithm was used to correct S-shape distortions of the images. CT images (1 mm slice) were prepared. In the treatment planning system positions of 15 catheter points (MP, measure points), control point (CP) and central marker (CM) were reconstructed for each series of images. Distances between 15 catheter points and control point, and between catheter points and central marker were calculated. RESULTS: There were no statistically significant differences observed for IBU and CT based reconstructions for all orientations of the phantom (p > 0.05, U-Mann Whitney Test). There were no statistically significant differences observed between reconstruction based on IBU images with and without Earth magnetic field correction algorithm for phantom located perpendicular to the IBU table (p > 0.05, Wilcoxon Test). Statistically significant differences were observed only for images set with long axis of the phantom located parallel to the table (p < 0.05, Wilcoxon Test). There were no statistically significant differences observed for values doses in reference points for reconstruction based on IBU images and CT for all orientations of the phantom. CONCLUSIONS: 1. Obtained results showed that IBU (radiographs based) reconstruction of the catheter placement is the reliable and accurate method for interstitial implants when reconstruction based on CT “catheter tracking” is not possible or not necessary. 2. The Earth magnetic field correction algorithm should be always use to correct S-shape distortions; reconstruction will be more accurate in particular orientations of image intensifier of the imaging unit.
format Online
Article
Text
id pubmed-5080396
institution National Center for Biotechnology Information
language English
publishDate 2009
publisher Termedia Publishing House
record_format MEDLINE/PubMed
spelling pubmed-50803962016-10-31 Comparison and verification of the reconstruction method of the catheters for interstitial brachytherapy Blasiak, Barbara Zwierzchowski, Grzegorz Skowronek, Janusz Stefaniak, Patrycja Dymnicka, Magdalena J Contemp Brachytherapy Original Article PURPOSE: 1. Comparison and verification of accuracy of the implant reconstruction method based on images from IBU and CT. 2. Estimation of influence of the implant reconstruction method on dose disposition in selected reference points. MATERIAL AND METHODS: Paraffin-wax phantom with three catheters, central marker and control point were prepared. IBU unit were used for obtaining two series of images for reconstruction. The Earth magnetic field correction algorithm was used to correct S-shape distortions of the images. CT images (1 mm slice) were prepared. In the treatment planning system positions of 15 catheter points (MP, measure points), control point (CP) and central marker (CM) were reconstructed for each series of images. Distances between 15 catheter points and control point, and between catheter points and central marker were calculated. RESULTS: There were no statistically significant differences observed for IBU and CT based reconstructions for all orientations of the phantom (p > 0.05, U-Mann Whitney Test). There were no statistically significant differences observed between reconstruction based on IBU images with and without Earth magnetic field correction algorithm for phantom located perpendicular to the IBU table (p > 0.05, Wilcoxon Test). Statistically significant differences were observed only for images set with long axis of the phantom located parallel to the table (p < 0.05, Wilcoxon Test). There were no statistically significant differences observed for values doses in reference points for reconstruction based on IBU images and CT for all orientations of the phantom. CONCLUSIONS: 1. Obtained results showed that IBU (radiographs based) reconstruction of the catheter placement is the reliable and accurate method for interstitial implants when reconstruction based on CT “catheter tracking” is not possible or not necessary. 2. The Earth magnetic field correction algorithm should be always use to correct S-shape distortions; reconstruction will be more accurate in particular orientations of image intensifier of the imaging unit. Termedia Publishing House 2009-03-23 2009-03 /pmc/articles/PMC5080396/ /pubmed/27799955 Text en Copyright: © 2009 Termedia Sp. z o. o. http://creativecommons.org/licenses/by-nc-sa/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License, allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.
spellingShingle Original Article
Blasiak, Barbara
Zwierzchowski, Grzegorz
Skowronek, Janusz
Stefaniak, Patrycja
Dymnicka, Magdalena
Comparison and verification of the reconstruction method of the catheters for interstitial brachytherapy
title Comparison and verification of the reconstruction method of the catheters for interstitial brachytherapy
title_full Comparison and verification of the reconstruction method of the catheters for interstitial brachytherapy
title_fullStr Comparison and verification of the reconstruction method of the catheters for interstitial brachytherapy
title_full_unstemmed Comparison and verification of the reconstruction method of the catheters for interstitial brachytherapy
title_short Comparison and verification of the reconstruction method of the catheters for interstitial brachytherapy
title_sort comparison and verification of the reconstruction method of the catheters for interstitial brachytherapy
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5080396/
https://www.ncbi.nlm.nih.gov/pubmed/27799955
work_keys_str_mv AT blasiakbarbara comparisonandverificationofthereconstructionmethodofthecathetersforinterstitialbrachytherapy
AT zwierzchowskigrzegorz comparisonandverificationofthereconstructionmethodofthecathetersforinterstitialbrachytherapy
AT skowronekjanusz comparisonandverificationofthereconstructionmethodofthecathetersforinterstitialbrachytherapy
AT stefaniakpatrycja comparisonandverificationofthereconstructionmethodofthecathetersforinterstitialbrachytherapy
AT dymnickamagdalena comparisonandverificationofthereconstructionmethodofthecathetersforinterstitialbrachytherapy