Registration uncertainties between 3D cone beam computed tomography and different reference CT datasets in lung stereotactic body radiation therapy
BACKGROUND: The aim of this study was to analyze differences in couch shifts (setup errors) resulting from image registration of different CT datasets with free breathing cone beam CTs (FB-CBCT). As well automatic as manual image registrations were performed and registration results were correlated...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5080749/ https://www.ncbi.nlm.nih.gov/pubmed/27782858 http://dx.doi.org/10.1186/s13014-016-0720-9 |
_version_ | 1782462785172537344 |
---|---|
author | Oechsner, Markus Chizzali, Barbara Devecka, Michal Combs, Stephanie Elisabeth Wilkens, Jan Jakob Duma, Marciana Nona |
author_facet | Oechsner, Markus Chizzali, Barbara Devecka, Michal Combs, Stephanie Elisabeth Wilkens, Jan Jakob Duma, Marciana Nona |
author_sort | Oechsner, Markus |
collection | PubMed |
description | BACKGROUND: The aim of this study was to analyze differences in couch shifts (setup errors) resulting from image registration of different CT datasets with free breathing cone beam CTs (FB-CBCT). As well automatic as manual image registrations were performed and registration results were correlated to tumor characteristics. METHODS: FB-CBCT image registration was performed for 49 patients with lung lesions using slow planning CT (PCT), average intensity projection (AIP), maximum intensity projection (MIP) and mid-ventilation CTs (MidV) as reference images. Both, automatic and manual image registrations were applied. Shift differences were evaluated between the registered CT datasets for automatic and manual registration, respectively. Furthermore, differences between automatic and manual registration were analyzed for the same CT datasets. The registration results were statistically analyzed and correlated to tumor characteristics (3D tumor motion, tumor volume, superior-inferior (SI) distance, tumor environment). RESULTS: Median 3D shift differences over all patients were between 0.5 mm (AIPvsMIP) and 1.9 mm (MIPvsPCT and MidVvsPCT) for the automatic registration and between 1.8 mm (AIPvsPCT) and 2.8 mm (MIPvsPCT and MidVvsPCT) for the manual registration. For some patients, large shift differences (>5.0 mm) were found (maximum 10.5 mm, automatic registration). Comparing automatic vs manual registrations for the same reference CTs, ∆AIP achieved the smallest (1.1 mm) and ∆MIP the largest (1.9 mm) median 3D shift differences. The standard deviation (variability) for the 3D shift differences was also the smallest for ∆AIP (1.1 mm). Significant correlations (p < 0.01) between 3D shift difference and 3D tumor motion (AIPvsMIP, MIPvsMidV) and SI distance (AIPvsMIP) (automatic) and also for 3D tumor motion (∆PCT, ∆MidV; automatic vs manual) were found. CONCLUSIONS: Using different CT datasets for image registration with FB-CBCTs can result in different 3D couch shifts. Manual registrations achieved partly different 3D shifts than automatic registrations. AIP CTs yielded the smallest shift differences and might be the most appropriate CT dataset for registration with 3D FB-CBCTs. |
format | Online Article Text |
id | pubmed-5080749 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-50807492016-10-31 Registration uncertainties between 3D cone beam computed tomography and different reference CT datasets in lung stereotactic body radiation therapy Oechsner, Markus Chizzali, Barbara Devecka, Michal Combs, Stephanie Elisabeth Wilkens, Jan Jakob Duma, Marciana Nona Radiat Oncol Research BACKGROUND: The aim of this study was to analyze differences in couch shifts (setup errors) resulting from image registration of different CT datasets with free breathing cone beam CTs (FB-CBCT). As well automatic as manual image registrations were performed and registration results were correlated to tumor characteristics. METHODS: FB-CBCT image registration was performed for 49 patients with lung lesions using slow planning CT (PCT), average intensity projection (AIP), maximum intensity projection (MIP) and mid-ventilation CTs (MidV) as reference images. Both, automatic and manual image registrations were applied. Shift differences were evaluated between the registered CT datasets for automatic and manual registration, respectively. Furthermore, differences between automatic and manual registration were analyzed for the same CT datasets. The registration results were statistically analyzed and correlated to tumor characteristics (3D tumor motion, tumor volume, superior-inferior (SI) distance, tumor environment). RESULTS: Median 3D shift differences over all patients were between 0.5 mm (AIPvsMIP) and 1.9 mm (MIPvsPCT and MidVvsPCT) for the automatic registration and between 1.8 mm (AIPvsPCT) and 2.8 mm (MIPvsPCT and MidVvsPCT) for the manual registration. For some patients, large shift differences (>5.0 mm) were found (maximum 10.5 mm, automatic registration). Comparing automatic vs manual registrations for the same reference CTs, ∆AIP achieved the smallest (1.1 mm) and ∆MIP the largest (1.9 mm) median 3D shift differences. The standard deviation (variability) for the 3D shift differences was also the smallest for ∆AIP (1.1 mm). Significant correlations (p < 0.01) between 3D shift difference and 3D tumor motion (AIPvsMIP, MIPvsMidV) and SI distance (AIPvsMIP) (automatic) and also for 3D tumor motion (∆PCT, ∆MidV; automatic vs manual) were found. CONCLUSIONS: Using different CT datasets for image registration with FB-CBCTs can result in different 3D couch shifts. Manual registrations achieved partly different 3D shifts than automatic registrations. AIP CTs yielded the smallest shift differences and might be the most appropriate CT dataset for registration with 3D FB-CBCTs. BioMed Central 2016-10-26 /pmc/articles/PMC5080749/ /pubmed/27782858 http://dx.doi.org/10.1186/s13014-016-0720-9 Text en © The Author(s). 2016 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
spellingShingle | Research Oechsner, Markus Chizzali, Barbara Devecka, Michal Combs, Stephanie Elisabeth Wilkens, Jan Jakob Duma, Marciana Nona Registration uncertainties between 3D cone beam computed tomography and different reference CT datasets in lung stereotactic body radiation therapy |
title | Registration uncertainties between 3D cone beam computed tomography and different reference CT datasets in lung stereotactic body radiation therapy |
title_full | Registration uncertainties between 3D cone beam computed tomography and different reference CT datasets in lung stereotactic body radiation therapy |
title_fullStr | Registration uncertainties between 3D cone beam computed tomography and different reference CT datasets in lung stereotactic body radiation therapy |
title_full_unstemmed | Registration uncertainties between 3D cone beam computed tomography and different reference CT datasets in lung stereotactic body radiation therapy |
title_short | Registration uncertainties between 3D cone beam computed tomography and different reference CT datasets in lung stereotactic body radiation therapy |
title_sort | registration uncertainties between 3d cone beam computed tomography and different reference ct datasets in lung stereotactic body radiation therapy |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5080749/ https://www.ncbi.nlm.nih.gov/pubmed/27782858 http://dx.doi.org/10.1186/s13014-016-0720-9 |
work_keys_str_mv | AT oechsnermarkus registrationuncertaintiesbetween3dconebeamcomputedtomographyanddifferentreferencectdatasetsinlungstereotacticbodyradiationtherapy AT chizzalibarbara registrationuncertaintiesbetween3dconebeamcomputedtomographyanddifferentreferencectdatasetsinlungstereotacticbodyradiationtherapy AT deveckamichal registrationuncertaintiesbetween3dconebeamcomputedtomographyanddifferentreferencectdatasetsinlungstereotacticbodyradiationtherapy AT combsstephanieelisabeth registrationuncertaintiesbetween3dconebeamcomputedtomographyanddifferentreferencectdatasetsinlungstereotacticbodyradiationtherapy AT wilkensjanjakob registrationuncertaintiesbetween3dconebeamcomputedtomographyanddifferentreferencectdatasetsinlungstereotacticbodyradiationtherapy AT dumamarciananona registrationuncertaintiesbetween3dconebeamcomputedtomographyanddifferentreferencectdatasetsinlungstereotacticbodyradiationtherapy |