Estimation of extravascular lung water using the transpulmonary ultrasound dilution (TPUD) method: a validation study in neonatal lambs

Increased extravascular lung water (EVLW) may contribute to respiratory failure in neonates. Accurate measurement of EVLW in these patients is limited due to the lack of bedside methods. The aim of this pilot study was to investigate the reliability of the transpulmonary ultrasound dilution (TPUD) t...

Descripción completa

Detalles Bibliográficos
Autores principales: Vrancken, S. L., Nusmeier, A., Hopman, J. C., Liem, K. D., van der Hoeven, J. G., Lemson, J., van Heijst, A. F., de Boode, W. P.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Netherlands 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5081382/
https://www.ncbi.nlm.nih.gov/pubmed/26563187
http://dx.doi.org/10.1007/s10877-015-9803-7
Descripción
Sumario:Increased extravascular lung water (EVLW) may contribute to respiratory failure in neonates. Accurate measurement of EVLW in these patients is limited due to the lack of bedside methods. The aim of this pilot study was to investigate the reliability of the transpulmonary ultrasound dilution (TPUD) technique as a possible method for estimating EVLW in a neonatal animal model. Pulmonary edema was induced in 11 lambs by repeated surfactant lavages. In between the lavages, EVLW indexed by bodyweight was estimated by TPUD (EVLWI(tpud)) and transpulmonary dye dilution (EVLWI(tpdd)) (n = 22). Final EVLWI(tpud) measurements were also compared with EVLWI estimations by gold standard post mortem gravimetry (EVLWI(grav)) (n = 6). EVLWI was also measured in two additional lambs without pulmonary edema. Bland–Altman plots showed a mean bias between EVLWI(tpud) and EVLWI(tpdd) of −3.4 mL/kg (LOA ± 25.8 mL/kg) and between EVLWI(tpud) and EVLWI(grav) of 1.7 mL/kg (LOA ± 8.3 mL/kg). The percentage errors were 109 and 43 % respectively. The correlation between changes in EVLW measured by TPUD and TPDD was r(2) = 0.22. Agreement between EVLWI measurements by TPUD and TPDD was low. Trending ability to detect changes between these two methods in EVLWI was questionable. The accuracy of EVLWI(tpud) was good compared to the gold standard gravimetric method but the TPUD lacked precision in its current prototype. Based on these limited data, we believe that TPUD has potential for future use to estimate EVLW after adaptation of the algorithm. Larger studies are needed to support our findings.