Effect of Bayesian-penalized likelihood reconstruction on [13N]-NH3 rest perfusion quantification
OBJECTIVES: Myocardial blood flow (MBF) imaging is used in patients with suspected cardiac sarcoidosis, and also in stress/rest studies. The accuracy of MBF is dependent on imaging parameters such as new reconstruction methodologies. In this work, we aim to assess the impact of a novel PET reconstru...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer US
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5084874/ https://www.ncbi.nlm.nih.gov/pubmed/27435278 http://dx.doi.org/10.1007/s12350-016-0554-8 |
Sumario: | OBJECTIVES: Myocardial blood flow (MBF) imaging is used in patients with suspected cardiac sarcoidosis, and also in stress/rest studies. The accuracy of MBF is dependent on imaging parameters such as new reconstruction methodologies. In this work, we aim to assess the impact of a novel PET reconstruction algorithm (Bayesian-penalized likelihood—BPL) on the values determined from the calculation of [13N]-NH3 MBF values. METHODS: Data from 21 patients undergoing rest MBF evaluation [13N]-NH3 as part of sarcoidosis imaging were retrospectively analyzed. Each scan was reconstructed with a range of BPL coefficients (1-500), and standard clinical FBP and OSEM reconstructions. MBF values were calculated via an automated software routine for all datasets. RESULTS: Reconstruction of [13N]-NH3 dynamic data using the BPL, OSEM, or FBP reconstruction showed no quantitative differences for the calculation of territorial or global MBF (P = .97). Image noise was lower using OSEM or BPL reconstructions than FBP and noise from BPL reached levels seen in OSEM images between B = 300 and B = 400. Intrasubject differences between all reconstructions over all patients in respect of all cardiac territories showed a maximum coefficient of variation of 9.74%. CONCLUSION: Quantitation of MBF via kinetic modeling of cardiac rest MBF by [13N]-NH3 is minimally affected by the use of a BPL reconstruction technique, with BPL images presenting with less noise. |
---|