Cargando…
The Use of a Pressure-Indicating Sensor Film to Provide Feedback upon Hydrogel-Forming Microneedle Array Self-Application In Vivo
PURPOSE: To evaluate the combination of a pressure-indicating sensor film with hydrogel-forming microneedle arrays, as a method of feedback to confirm MN insertion in vivo. METHODS: Pilot in vitro insertion studies were conducted using a Texture Analyser to insert MN arrays, coupled with a pressure-...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer US
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5093199/ https://www.ncbi.nlm.nih.gov/pubmed/27633885 http://dx.doi.org/10.1007/s11095-016-2032-z |
_version_ | 1782464868610211840 |
---|---|
author | Vicente-Pérez, Eva M. Quinn, Helen L. McAlister, Emma O’Neill, Shannon Hanna, Lezley-Anne Barry, Johanne G. Donnelly, Ryan F. |
author_facet | Vicente-Pérez, Eva M. Quinn, Helen L. McAlister, Emma O’Neill, Shannon Hanna, Lezley-Anne Barry, Johanne G. Donnelly, Ryan F. |
author_sort | Vicente-Pérez, Eva M. |
collection | PubMed |
description | PURPOSE: To evaluate the combination of a pressure-indicating sensor film with hydrogel-forming microneedle arrays, as a method of feedback to confirm MN insertion in vivo. METHODS: Pilot in vitro insertion studies were conducted using a Texture Analyser to insert MN arrays, coupled with a pressure-indicating sensor film, at varying forces into excised neonatal porcine skin. In vivo studies involved twenty human volunteers, who self-applied two hydrogel-forming MN arrays, one with a pressure-indicating sensor film incorporated and one without. Optical coherence tomography was employed to measure the resulting penetration depth and colorimetric analysis to investigate the associated colour change of the pressure-indicating sensor film. RESULTS: Microneedle insertion was achieved in vitro at three different forces, demonstrating the colour change of the pressure-indicating sensor film upon application of increasing pressure. When self-applied in vivo, there was no significant difference in the microneedle penetration depth resulting from each type of array, with a mean depth of 237 μm recorded. When the pressure-indicating sensor film was present, a colour change occurred upon each application, providing evidence of insertion. CONCLUSIONS: For the first time, this study shows how the incorporation of a simple, low-cost pressure-indicating sensor film can indicate microneedle insertion in vitro and in vivo, providing visual feedback to assure the user of correct application. Such a strategy may enhance usability of a microneedle device and, hence, assist in the future translation of the technology to widespread clinical use. |
format | Online Article Text |
id | pubmed-5093199 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | Springer US |
record_format | MEDLINE/PubMed |
spelling | pubmed-50931992016-11-17 The Use of a Pressure-Indicating Sensor Film to Provide Feedback upon Hydrogel-Forming Microneedle Array Self-Application In Vivo Vicente-Pérez, Eva M. Quinn, Helen L. McAlister, Emma O’Neill, Shannon Hanna, Lezley-Anne Barry, Johanne G. Donnelly, Ryan F. Pharm Res Research Paper PURPOSE: To evaluate the combination of a pressure-indicating sensor film with hydrogel-forming microneedle arrays, as a method of feedback to confirm MN insertion in vivo. METHODS: Pilot in vitro insertion studies were conducted using a Texture Analyser to insert MN arrays, coupled with a pressure-indicating sensor film, at varying forces into excised neonatal porcine skin. In vivo studies involved twenty human volunteers, who self-applied two hydrogel-forming MN arrays, one with a pressure-indicating sensor film incorporated and one without. Optical coherence tomography was employed to measure the resulting penetration depth and colorimetric analysis to investigate the associated colour change of the pressure-indicating sensor film. RESULTS: Microneedle insertion was achieved in vitro at three different forces, demonstrating the colour change of the pressure-indicating sensor film upon application of increasing pressure. When self-applied in vivo, there was no significant difference in the microneedle penetration depth resulting from each type of array, with a mean depth of 237 μm recorded. When the pressure-indicating sensor film was present, a colour change occurred upon each application, providing evidence of insertion. CONCLUSIONS: For the first time, this study shows how the incorporation of a simple, low-cost pressure-indicating sensor film can indicate microneedle insertion in vitro and in vivo, providing visual feedback to assure the user of correct application. Such a strategy may enhance usability of a microneedle device and, hence, assist in the future translation of the technology to widespread clinical use. Springer US 2016-09-15 2016 /pmc/articles/PMC5093199/ /pubmed/27633885 http://dx.doi.org/10.1007/s11095-016-2032-z Text en © The Author(s) 2016 Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. |
spellingShingle | Research Paper Vicente-Pérez, Eva M. Quinn, Helen L. McAlister, Emma O’Neill, Shannon Hanna, Lezley-Anne Barry, Johanne G. Donnelly, Ryan F. The Use of a Pressure-Indicating Sensor Film to Provide Feedback upon Hydrogel-Forming Microneedle Array Self-Application In Vivo |
title | The Use of a Pressure-Indicating Sensor Film to Provide Feedback upon Hydrogel-Forming Microneedle Array Self-Application In Vivo |
title_full | The Use of a Pressure-Indicating Sensor Film to Provide Feedback upon Hydrogel-Forming Microneedle Array Self-Application In Vivo |
title_fullStr | The Use of a Pressure-Indicating Sensor Film to Provide Feedback upon Hydrogel-Forming Microneedle Array Self-Application In Vivo |
title_full_unstemmed | The Use of a Pressure-Indicating Sensor Film to Provide Feedback upon Hydrogel-Forming Microneedle Array Self-Application In Vivo |
title_short | The Use of a Pressure-Indicating Sensor Film to Provide Feedback upon Hydrogel-Forming Microneedle Array Self-Application In Vivo |
title_sort | use of a pressure-indicating sensor film to provide feedback upon hydrogel-forming microneedle array self-application in vivo |
topic | Research Paper |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5093199/ https://www.ncbi.nlm.nih.gov/pubmed/27633885 http://dx.doi.org/10.1007/s11095-016-2032-z |
work_keys_str_mv | AT vicenteperezevam theuseofapressureindicatingsensorfilmtoprovidefeedbackuponhydrogelformingmicroneedlearrayselfapplicationinvivo AT quinnhelenl theuseofapressureindicatingsensorfilmtoprovidefeedbackuponhydrogelformingmicroneedlearrayselfapplicationinvivo AT mcalisteremma theuseofapressureindicatingsensorfilmtoprovidefeedbackuponhydrogelformingmicroneedlearrayselfapplicationinvivo AT oneillshannon theuseofapressureindicatingsensorfilmtoprovidefeedbackuponhydrogelformingmicroneedlearrayselfapplicationinvivo AT hannalezleyanne theuseofapressureindicatingsensorfilmtoprovidefeedbackuponhydrogelformingmicroneedlearrayselfapplicationinvivo AT barryjohanneg theuseofapressureindicatingsensorfilmtoprovidefeedbackuponhydrogelformingmicroneedlearrayselfapplicationinvivo AT donnellyryanf theuseofapressureindicatingsensorfilmtoprovidefeedbackuponhydrogelformingmicroneedlearrayselfapplicationinvivo AT vicenteperezevam useofapressureindicatingsensorfilmtoprovidefeedbackuponhydrogelformingmicroneedlearrayselfapplicationinvivo AT quinnhelenl useofapressureindicatingsensorfilmtoprovidefeedbackuponhydrogelformingmicroneedlearrayselfapplicationinvivo AT mcalisteremma useofapressureindicatingsensorfilmtoprovidefeedbackuponhydrogelformingmicroneedlearrayselfapplicationinvivo AT oneillshannon useofapressureindicatingsensorfilmtoprovidefeedbackuponhydrogelformingmicroneedlearrayselfapplicationinvivo AT hannalezleyanne useofapressureindicatingsensorfilmtoprovidefeedbackuponhydrogelformingmicroneedlearrayselfapplicationinvivo AT barryjohanneg useofapressureindicatingsensorfilmtoprovidefeedbackuponhydrogelformingmicroneedlearrayselfapplicationinvivo AT donnellyryanf useofapressureindicatingsensorfilmtoprovidefeedbackuponhydrogelformingmicroneedlearrayselfapplicationinvivo |