Spectral Gini Index for Quantifying the Depth of Consciousness

We propose indices that describe the depth of consciousness (DOC) based on electroencephalograms (EEGs) acquired during anesthesia. The spectral Gini index (SpG) is a novel index utilizing the inequality in the powers of the EEG spectral components; a similar index is the binarized spectral Gini ind...

Descripción completa

Detalles Bibliográficos
Autores principales: You, Kyung-Jin, Noh, Gyu-Jeong, Shin, Hyun-Chool
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5093302/
https://www.ncbi.nlm.nih.gov/pubmed/27840633
http://dx.doi.org/10.1155/2016/2304356
Descripción
Sumario:We propose indices that describe the depth of consciousness (DOC) based on electroencephalograms (EEGs) acquired during anesthesia. The spectral Gini index (SpG) is a novel index utilizing the inequality in the powers of the EEG spectral components; a similar index is the binarized spectral Gini index (BSpG), which has low computational complexity. A set of EEG data from 15 subjects was obtained during the induction and recovery periods of general anesthesia with propofol. The efficacy of the indices as indicators of the DOC was demonstrated by examining Spearman's correlation coefficients between the indices and the effect-site concentration of propofol. A higher correlation was observed for SpG and BSpG (0.633 and 0.770, resp., p < 0.001) compared to the conventional indices. These results show that the proposed indices can achieve a reliable quantification of the DOC with simplified calculations.