Cargando…

Computed Tomography Angiography with a 192-slice Dual-source Computed Tomography System: Improvements in Image Quality and Radiation Dose

PURPOSE: This study aims to compare image quality, radiation dose, and the influence of the heart rate on image quality of high-pitch spiral coronary computed tomography angiography (CCTA) using 128-slice (second generation) dual-source CT (DSCT) and a 192-slice DSCT (third generation) scanner. MATE...

Descripción completa

Detalles Bibliográficos
Autores principales: Linsen, Philip V M, Coenen, Adriaan, Lubbers, Marisa M, Dijkshoorn, Marcel L, Ouhlous, Mohamed, Nieman, Koen
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5093882/
https://www.ncbi.nlm.nih.gov/pubmed/27833784
http://dx.doi.org/10.4103/2156-7514.192840
_version_ 1782465015412948992
author Linsen, Philip V M
Coenen, Adriaan
Lubbers, Marisa M
Dijkshoorn, Marcel L
Ouhlous, Mohamed
Nieman, Koen
author_facet Linsen, Philip V M
Coenen, Adriaan
Lubbers, Marisa M
Dijkshoorn, Marcel L
Ouhlous, Mohamed
Nieman, Koen
author_sort Linsen, Philip V M
collection PubMed
description PURPOSE: This study aims to compare image quality, radiation dose, and the influence of the heart rate on image quality of high-pitch spiral coronary computed tomography angiography (CCTA) using 128-slice (second generation) dual-source CT (DSCT) and a 192-slice DSCT (third generation) scanner. MATERIALS AND METHODS: Two consecutive cohorts of fifty patients underwent CCTA by high-pitch spiral scan mode using 128 or 192-slice DSCT. The 192-slice DSCT system has a more powerful roentgen tube (2 × 120 kW) that allows CCTA acquisition at lower tube voltages, wider longitudinal coverage for faster table speed (732 m/s), and the use of iterative reconstruction. Objective image quality was measured as the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). Subjective image quality was evaluated using a Likert scale. RESULTS: While the effective dose was lower with 192-slice DSCT (1.2 ± 0.5 vs. 0.6 ± 0.3 mSv; P < 0.001), the SNR (18.9 ± 4.3 vs. 11.0 ± 2.9; P < 0.001) and CNR (23.5 ± 4.8 vs. 14.3 ± 4.1; P < 0.001) were superior to 128-slice DSCT. Although patients scanned with 192-slice DSCT had a faster heart rate (59 ± 7 vs. 56 ± 6; P = 0.045), subjective image quality was scored higher (4.2 ± 0.8 vs. 3.0 ± 0.7; P < 0.001) compared to 128-slice DSCT. CONCLUSIONS: High-pitch spiral CCTA by 192-slice DSCT provides better image quality, despite a higher average heart rate, at lower radiation doses compared to 128-slice DSCT.
format Online
Article
Text
id pubmed-5093882
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-50938822016-11-10 Computed Tomography Angiography with a 192-slice Dual-source Computed Tomography System: Improvements in Image Quality and Radiation Dose Linsen, Philip V M Coenen, Adriaan Lubbers, Marisa M Dijkshoorn, Marcel L Ouhlous, Mohamed Nieman, Koen J Clin Imaging Sci Original Article PURPOSE: This study aims to compare image quality, radiation dose, and the influence of the heart rate on image quality of high-pitch spiral coronary computed tomography angiography (CCTA) using 128-slice (second generation) dual-source CT (DSCT) and a 192-slice DSCT (third generation) scanner. MATERIALS AND METHODS: Two consecutive cohorts of fifty patients underwent CCTA by high-pitch spiral scan mode using 128 or 192-slice DSCT. The 192-slice DSCT system has a more powerful roentgen tube (2 × 120 kW) that allows CCTA acquisition at lower tube voltages, wider longitudinal coverage for faster table speed (732 m/s), and the use of iterative reconstruction. Objective image quality was measured as the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). Subjective image quality was evaluated using a Likert scale. RESULTS: While the effective dose was lower with 192-slice DSCT (1.2 ± 0.5 vs. 0.6 ± 0.3 mSv; P < 0.001), the SNR (18.9 ± 4.3 vs. 11.0 ± 2.9; P < 0.001) and CNR (23.5 ± 4.8 vs. 14.3 ± 4.1; P < 0.001) were superior to 128-slice DSCT. Although patients scanned with 192-slice DSCT had a faster heart rate (59 ± 7 vs. 56 ± 6; P = 0.045), subjective image quality was scored higher (4.2 ± 0.8 vs. 3.0 ± 0.7; P < 0.001) compared to 128-slice DSCT. CONCLUSIONS: High-pitch spiral CCTA by 192-slice DSCT provides better image quality, despite a higher average heart rate, at lower radiation doses compared to 128-slice DSCT. Medknow Publications & Media Pvt Ltd 2016-10-21 /pmc/articles/PMC5093882/ /pubmed/27833784 http://dx.doi.org/10.4103/2156-7514.192840 Text en Copyright: © 2016 Journal of Clinical Imaging Science http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 3.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms.
spellingShingle Original Article
Linsen, Philip V M
Coenen, Adriaan
Lubbers, Marisa M
Dijkshoorn, Marcel L
Ouhlous, Mohamed
Nieman, Koen
Computed Tomography Angiography with a 192-slice Dual-source Computed Tomography System: Improvements in Image Quality and Radiation Dose
title Computed Tomography Angiography with a 192-slice Dual-source Computed Tomography System: Improvements in Image Quality and Radiation Dose
title_full Computed Tomography Angiography with a 192-slice Dual-source Computed Tomography System: Improvements in Image Quality and Radiation Dose
title_fullStr Computed Tomography Angiography with a 192-slice Dual-source Computed Tomography System: Improvements in Image Quality and Radiation Dose
title_full_unstemmed Computed Tomography Angiography with a 192-slice Dual-source Computed Tomography System: Improvements in Image Quality and Radiation Dose
title_short Computed Tomography Angiography with a 192-slice Dual-source Computed Tomography System: Improvements in Image Quality and Radiation Dose
title_sort computed tomography angiography with a 192-slice dual-source computed tomography system: improvements in image quality and radiation dose
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5093882/
https://www.ncbi.nlm.nih.gov/pubmed/27833784
http://dx.doi.org/10.4103/2156-7514.192840
work_keys_str_mv AT linsenphilipvm computedtomographyangiographywitha192slicedualsourcecomputedtomographysystemimprovementsinimagequalityandradiationdose
AT coenenadriaan computedtomographyangiographywitha192slicedualsourcecomputedtomographysystemimprovementsinimagequalityandradiationdose
AT lubbersmarisam computedtomographyangiographywitha192slicedualsourcecomputedtomographysystemimprovementsinimagequalityandradiationdose
AT dijkshoornmarcell computedtomographyangiographywitha192slicedualsourcecomputedtomographysystemimprovementsinimagequalityandradiationdose
AT ouhlousmohamed computedtomographyangiographywitha192slicedualsourcecomputedtomographysystemimprovementsinimagequalityandradiationdose
AT niemankoen computedtomographyangiographywitha192slicedualsourcecomputedtomographysystemimprovementsinimagequalityandradiationdose