The Biomechanics and Optimization of the Needle-Syringe System for Injecting Triamcinolone Acetonide into Keloids

Purpose. Injecting triamcinolone acetonide (TA) into a keloid is physically challenging due to the density of keloids. The purpose was to investigate the effects of various syringe and needle combinations on the injection force to determine the most ergonomic combination. Materials and Methods. A lo...

Descripción completa

Detalles Bibliográficos
Autores principales: Vo, Anthony, Doumit, Marc, Rockwell, Gloria
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5098087/
https://www.ncbi.nlm.nih.gov/pubmed/27843936
http://dx.doi.org/10.1155/2016/5162394
_version_ 1782465713452089344
author Vo, Anthony
Doumit, Marc
Rockwell, Gloria
author_facet Vo, Anthony
Doumit, Marc
Rockwell, Gloria
author_sort Vo, Anthony
collection PubMed
description Purpose. Injecting triamcinolone acetonide (TA) into a keloid is physically challenging due to the density of keloids. The purpose was to investigate the effects of various syringe and needle combinations on the injection force to determine the most ergonomic combination. Materials and Methods. A load cell was used to generate and measure the injection force. Phase 1: the injection force of 5 common syringes was measured by injecting water into air. The syringe that required the lowest injection force was evaluated with various needle gauges (25, 27, and 30 G) and lengths (16, 25, and 38 mm) by injecting TA (40 mg/mL) into air. The needle-syringe combination with the lowest injection force (CLIF) was deemed the most ergonomic combination. Phase 2: comparisons between the CLIF and a standard combination (SC) were performed by injecting TA into air and tap water into a keloid specimen. Intraclass Correlation Coefficient (ICC) and independent t-test were used. Results. Increasing the syringe caliber, injection speed, and needle gauge and length significantly increased the injection force (p value < 0.001). The SC required a maximum force of 40.0 N to inject water into keloid, compared to 25.0 N for the CLIF. Injecting TA into keloid using the SC would require an injection force that was 103.5% of the maximum force female thumbs could exert compared to 64.8% for the CLIF. ICC values were greater than 0.4. Conclusions. The 1 mL polycarbonate syringe with a 25 G, 16 mm needle (CLIF) was the most ergonomic combination. The SC required a substantial injection force, which may represent a physical challenge for female thumbs.
format Online
Article
Text
id pubmed-5098087
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Hindawi Publishing Corporation
record_format MEDLINE/PubMed
spelling pubmed-50980872016-11-14 The Biomechanics and Optimization of the Needle-Syringe System for Injecting Triamcinolone Acetonide into Keloids Vo, Anthony Doumit, Marc Rockwell, Gloria J Med Eng Research Article Purpose. Injecting triamcinolone acetonide (TA) into a keloid is physically challenging due to the density of keloids. The purpose was to investigate the effects of various syringe and needle combinations on the injection force to determine the most ergonomic combination. Materials and Methods. A load cell was used to generate and measure the injection force. Phase 1: the injection force of 5 common syringes was measured by injecting water into air. The syringe that required the lowest injection force was evaluated with various needle gauges (25, 27, and 30 G) and lengths (16, 25, and 38 mm) by injecting TA (40 mg/mL) into air. The needle-syringe combination with the lowest injection force (CLIF) was deemed the most ergonomic combination. Phase 2: comparisons between the CLIF and a standard combination (SC) were performed by injecting TA into air and tap water into a keloid specimen. Intraclass Correlation Coefficient (ICC) and independent t-test were used. Results. Increasing the syringe caliber, injection speed, and needle gauge and length significantly increased the injection force (p value < 0.001). The SC required a maximum force of 40.0 N to inject water into keloid, compared to 25.0 N for the CLIF. Injecting TA into keloid using the SC would require an injection force that was 103.5% of the maximum force female thumbs could exert compared to 64.8% for the CLIF. ICC values were greater than 0.4. Conclusions. The 1 mL polycarbonate syringe with a 25 G, 16 mm needle (CLIF) was the most ergonomic combination. The SC required a substantial injection force, which may represent a physical challenge for female thumbs. Hindawi Publishing Corporation 2016 2016-10-24 /pmc/articles/PMC5098087/ /pubmed/27843936 http://dx.doi.org/10.1155/2016/5162394 Text en Copyright © 2016 Anthony Vo et al. https://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Vo, Anthony
Doumit, Marc
Rockwell, Gloria
The Biomechanics and Optimization of the Needle-Syringe System for Injecting Triamcinolone Acetonide into Keloids
title The Biomechanics and Optimization of the Needle-Syringe System for Injecting Triamcinolone Acetonide into Keloids
title_full The Biomechanics and Optimization of the Needle-Syringe System for Injecting Triamcinolone Acetonide into Keloids
title_fullStr The Biomechanics and Optimization of the Needle-Syringe System for Injecting Triamcinolone Acetonide into Keloids
title_full_unstemmed The Biomechanics and Optimization of the Needle-Syringe System for Injecting Triamcinolone Acetonide into Keloids
title_short The Biomechanics and Optimization of the Needle-Syringe System for Injecting Triamcinolone Acetonide into Keloids
title_sort biomechanics and optimization of the needle-syringe system for injecting triamcinolone acetonide into keloids
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5098087/
https://www.ncbi.nlm.nih.gov/pubmed/27843936
http://dx.doi.org/10.1155/2016/5162394
work_keys_str_mv AT voanthony thebiomechanicsandoptimizationoftheneedlesyringesystemforinjectingtriamcinoloneacetonideintokeloids
AT doumitmarc thebiomechanicsandoptimizationoftheneedlesyringesystemforinjectingtriamcinoloneacetonideintokeloids
AT rockwellgloria thebiomechanicsandoptimizationoftheneedlesyringesystemforinjectingtriamcinoloneacetonideintokeloids
AT voanthony biomechanicsandoptimizationoftheneedlesyringesystemforinjectingtriamcinoloneacetonideintokeloids
AT doumitmarc biomechanicsandoptimizationoftheneedlesyringesystemforinjectingtriamcinoloneacetonideintokeloids
AT rockwellgloria biomechanicsandoptimizationoftheneedlesyringesystemforinjectingtriamcinoloneacetonideintokeloids