Cargando…
Novel device, a temporary guidewire fixator
A novel device for distal fixation of a guidewire was tested in regards to deployment and retrieval, deposition in the blood stream and force of fixation in a pig model. Eleven pigs were subjected to full anaesthesia and heparinized to active clotting time 250–350 s. Uninterrupted blood flow during...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5098692/ https://www.ncbi.nlm.nih.gov/pubmed/26864777 http://dx.doi.org/10.1177/1708538116629265 |
_version_ | 1782465811080806400 |
---|---|
author | Bosaeus, Linus Mani, Kevin Wanhainen, Anders Liungman, Krister |
author_facet | Bosaeus, Linus Mani, Kevin Wanhainen, Anders Liungman, Krister |
author_sort | Bosaeus, Linus |
collection | PubMed |
description | A novel device for distal fixation of a guidewire was tested in regards to deployment and retrieval, deposition in the blood stream and force of fixation in a pig model. Eleven pigs were subjected to full anaesthesia and heparinized to active clotting time 250–350 s. Uninterrupted blood flow during 4 h deposition was assessed by angiography and inspected for thrombus deposition upon retrieval. The force of fixation was investigated up to the level of loss of fixation (displacement force). The device was successfully deployed and retrieved in over 40 cases. In one case, an alternative method for bailout retrieval was used. Deposition for 4 h was performed, and uninterrupted blood flow was verified by angiography. No instances of arterial occlusion or thrombosis were detected. The median dislocation force was 7.6 N. No arterial rupture or dissection was detected following the loss of fixation. As a conclusion, the device was considered safe and functional in this animal test model. |
format | Online Article Text |
id | pubmed-5098692 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | SAGE Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-50986922016-11-14 Novel device, a temporary guidewire fixator Bosaeus, Linus Mani, Kevin Wanhainen, Anders Liungman, Krister Vascular Original Articles A novel device for distal fixation of a guidewire was tested in regards to deployment and retrieval, deposition in the blood stream and force of fixation in a pig model. Eleven pigs were subjected to full anaesthesia and heparinized to active clotting time 250–350 s. Uninterrupted blood flow during 4 h deposition was assessed by angiography and inspected for thrombus deposition upon retrieval. The force of fixation was investigated up to the level of loss of fixation (displacement force). The device was successfully deployed and retrieved in over 40 cases. In one case, an alternative method for bailout retrieval was used. Deposition for 4 h was performed, and uninterrupted blood flow was verified by angiography. No instances of arterial occlusion or thrombosis were detected. The median dislocation force was 7.6 N. No arterial rupture or dissection was detected following the loss of fixation. As a conclusion, the device was considered safe and functional in this animal test model. SAGE Publications 2016-02-09 2016-12 /pmc/articles/PMC5098692/ /pubmed/26864777 http://dx.doi.org/10.1177/1708538116629265 Text en © The Author(s) 2016 http://creativecommons.org/licenses/by-nc/3.0/ This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 3.0 License (http://www.creativecommons.org/licenses/by-nc/3.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access page(https://us.sagepub.com/en-us/nam/open-access-at-sage). |
spellingShingle | Original Articles Bosaeus, Linus Mani, Kevin Wanhainen, Anders Liungman, Krister Novel device, a temporary guidewire fixator |
title | Novel device, a temporary guidewire fixator |
title_full | Novel device, a temporary guidewire fixator |
title_fullStr | Novel device, a temporary guidewire fixator |
title_full_unstemmed | Novel device, a temporary guidewire fixator |
title_short | Novel device, a temporary guidewire fixator |
title_sort | novel device, a temporary guidewire fixator |
topic | Original Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5098692/ https://www.ncbi.nlm.nih.gov/pubmed/26864777 http://dx.doi.org/10.1177/1708538116629265 |
work_keys_str_mv | AT bosaeuslinus noveldeviceatemporaryguidewirefixator AT manikevin noveldeviceatemporaryguidewirefixator AT wanhainenanders noveldeviceatemporaryguidewirefixator AT liungmankrister noveldeviceatemporaryguidewirefixator |