Blood Collection Tubes and Storage Temperature Should Be Evaluated when Using the Siemens ADVIA Centaur XP for Measuring 25-Hydroxyvitamin D
A significant bias was found when using the Siemens ADVIA Centaur XP system for measurement of 25-hydroxyvitamin D (25OHD) with VACUETTE(®) tubes with Serum Clot Activator and Gel. Here, we examined whether other commonly used tubes or temperatures affected 25OHD results obtained with the Siemens AD...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5104342/ https://www.ncbi.nlm.nih.gov/pubmed/27832181 http://dx.doi.org/10.1371/journal.pone.0166327 |
_version_ | 1782466727395721216 |
---|---|
author | Yu, Songlin Zhou, Weiyan Cheng, Xinqi Fang, Huiling Zhang, Ruiping Cheng, Qian Han, Jianhua Su, Wei Xia, Liangyu Qiu, Ling |
author_facet | Yu, Songlin Zhou, Weiyan Cheng, Xinqi Fang, Huiling Zhang, Ruiping Cheng, Qian Han, Jianhua Su, Wei Xia, Liangyu Qiu, Ling |
author_sort | Yu, Songlin |
collection | PubMed |
description | A significant bias was found when using the Siemens ADVIA Centaur XP system for measurement of 25-hydroxyvitamin D (25OHD) with VACUETTE(®) tubes with Serum Clot Activator and Gel. Here, we examined whether other commonly used tubes or temperatures affected 25OHD results obtained with the Siemens ADVIA Centaur XP system. Serum was collected into five types of vacuum blood collection tubes from three manufacturers, and 25OHD was analyzed using the Siemens ADVIA Centaur XP system and liquid chromatography tandem mass spectrometry (LC-MS/MS) immediately or after storage at 4°C or -80°C for 48 h. Significantly higher 25OHD values were found when using the Siemens ADVIA Centaur XP system with VACUETTE(®) tubes with serum clot activator and gel and VACUETTE(®) tubes with clot activator but no gel compared with VACUETTE(®) tubes with no additives. The 25OHD values in all of these tubes were not significantly different from those obtained by LC-MS/MS. Moreover, after storage at -80°C for 48 h, the values obtained in IMPROVEVACUTER(®) tubes with serum clot activator and gel significantly increased, with a mean bias of 74.6% compared with the values before storage, on analysis with the Siemens ADVIA Centaur XP system. VACUETTE(®) tubes containing additives significantly affect the accuracy of 25OHD results obtained using the Siemens ADVIA Centaur XP system. Additionally, the composition of serum collected in IMPROVEVACUTER(®) tubes was affected by freezing, resulting in different measurements when using the Siemens 25OHD assay platform. |
format | Online Article Text |
id | pubmed-5104342 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-51043422016-12-08 Blood Collection Tubes and Storage Temperature Should Be Evaluated when Using the Siemens ADVIA Centaur XP for Measuring 25-Hydroxyvitamin D Yu, Songlin Zhou, Weiyan Cheng, Xinqi Fang, Huiling Zhang, Ruiping Cheng, Qian Han, Jianhua Su, Wei Xia, Liangyu Qiu, Ling PLoS One Research Article A significant bias was found when using the Siemens ADVIA Centaur XP system for measurement of 25-hydroxyvitamin D (25OHD) with VACUETTE(®) tubes with Serum Clot Activator and Gel. Here, we examined whether other commonly used tubes or temperatures affected 25OHD results obtained with the Siemens ADVIA Centaur XP system. Serum was collected into five types of vacuum blood collection tubes from three manufacturers, and 25OHD was analyzed using the Siemens ADVIA Centaur XP system and liquid chromatography tandem mass spectrometry (LC-MS/MS) immediately or after storage at 4°C or -80°C for 48 h. Significantly higher 25OHD values were found when using the Siemens ADVIA Centaur XP system with VACUETTE(®) tubes with serum clot activator and gel and VACUETTE(®) tubes with clot activator but no gel compared with VACUETTE(®) tubes with no additives. The 25OHD values in all of these tubes were not significantly different from those obtained by LC-MS/MS. Moreover, after storage at -80°C for 48 h, the values obtained in IMPROVEVACUTER(®) tubes with serum clot activator and gel significantly increased, with a mean bias of 74.6% compared with the values before storage, on analysis with the Siemens ADVIA Centaur XP system. VACUETTE(®) tubes containing additives significantly affect the accuracy of 25OHD results obtained using the Siemens ADVIA Centaur XP system. Additionally, the composition of serum collected in IMPROVEVACUTER(®) tubes was affected by freezing, resulting in different measurements when using the Siemens 25OHD assay platform. Public Library of Science 2016-11-10 /pmc/articles/PMC5104342/ /pubmed/27832181 http://dx.doi.org/10.1371/journal.pone.0166327 Text en © 2016 Yu et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. |
spellingShingle | Research Article Yu, Songlin Zhou, Weiyan Cheng, Xinqi Fang, Huiling Zhang, Ruiping Cheng, Qian Han, Jianhua Su, Wei Xia, Liangyu Qiu, Ling Blood Collection Tubes and Storage Temperature Should Be Evaluated when Using the Siemens ADVIA Centaur XP for Measuring 25-Hydroxyvitamin D |
title | Blood Collection Tubes and Storage Temperature Should Be Evaluated when Using the Siemens ADVIA Centaur XP for Measuring 25-Hydroxyvitamin D |
title_full | Blood Collection Tubes and Storage Temperature Should Be Evaluated when Using the Siemens ADVIA Centaur XP for Measuring 25-Hydroxyvitamin D |
title_fullStr | Blood Collection Tubes and Storage Temperature Should Be Evaluated when Using the Siemens ADVIA Centaur XP for Measuring 25-Hydroxyvitamin D |
title_full_unstemmed | Blood Collection Tubes and Storage Temperature Should Be Evaluated when Using the Siemens ADVIA Centaur XP for Measuring 25-Hydroxyvitamin D |
title_short | Blood Collection Tubes and Storage Temperature Should Be Evaluated when Using the Siemens ADVIA Centaur XP for Measuring 25-Hydroxyvitamin D |
title_sort | blood collection tubes and storage temperature should be evaluated when using the siemens advia centaur xp for measuring 25-hydroxyvitamin d |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5104342/ https://www.ncbi.nlm.nih.gov/pubmed/27832181 http://dx.doi.org/10.1371/journal.pone.0166327 |
work_keys_str_mv | AT yusonglin bloodcollectiontubesandstoragetemperatureshouldbeevaluatedwhenusingthesiemensadviacentaurxpformeasuring25hydroxyvitamind AT zhouweiyan bloodcollectiontubesandstoragetemperatureshouldbeevaluatedwhenusingthesiemensadviacentaurxpformeasuring25hydroxyvitamind AT chengxinqi bloodcollectiontubesandstoragetemperatureshouldbeevaluatedwhenusingthesiemensadviacentaurxpformeasuring25hydroxyvitamind AT fanghuiling bloodcollectiontubesandstoragetemperatureshouldbeevaluatedwhenusingthesiemensadviacentaurxpformeasuring25hydroxyvitamind AT zhangruiping bloodcollectiontubesandstoragetemperatureshouldbeevaluatedwhenusingthesiemensadviacentaurxpformeasuring25hydroxyvitamind AT chengqian bloodcollectiontubesandstoragetemperatureshouldbeevaluatedwhenusingthesiemensadviacentaurxpformeasuring25hydroxyvitamind AT hanjianhua bloodcollectiontubesandstoragetemperatureshouldbeevaluatedwhenusingthesiemensadviacentaurxpformeasuring25hydroxyvitamind AT suwei bloodcollectiontubesandstoragetemperatureshouldbeevaluatedwhenusingthesiemensadviacentaurxpformeasuring25hydroxyvitamind AT xialiangyu bloodcollectiontubesandstoragetemperatureshouldbeevaluatedwhenusingthesiemensadviacentaurxpformeasuring25hydroxyvitamind AT qiuling bloodcollectiontubesandstoragetemperatureshouldbeevaluatedwhenusingthesiemensadviacentaurxpformeasuring25hydroxyvitamind |