Blood Collection Tubes and Storage Temperature Should Be Evaluated when Using the Siemens ADVIA Centaur XP for Measuring 25-Hydroxyvitamin D

A significant bias was found when using the Siemens ADVIA Centaur XP system for measurement of 25-hydroxyvitamin D (25OHD) with VACUETTE(®) tubes with Serum Clot Activator and Gel. Here, we examined whether other commonly used tubes or temperatures affected 25OHD results obtained with the Siemens AD...

Descripción completa

Detalles Bibliográficos
Autores principales: Yu, Songlin, Zhou, Weiyan, Cheng, Xinqi, Fang, Huiling, Zhang, Ruiping, Cheng, Qian, Han, Jianhua, Su, Wei, Xia, Liangyu, Qiu, Ling
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5104342/
https://www.ncbi.nlm.nih.gov/pubmed/27832181
http://dx.doi.org/10.1371/journal.pone.0166327
_version_ 1782466727395721216
author Yu, Songlin
Zhou, Weiyan
Cheng, Xinqi
Fang, Huiling
Zhang, Ruiping
Cheng, Qian
Han, Jianhua
Su, Wei
Xia, Liangyu
Qiu, Ling
author_facet Yu, Songlin
Zhou, Weiyan
Cheng, Xinqi
Fang, Huiling
Zhang, Ruiping
Cheng, Qian
Han, Jianhua
Su, Wei
Xia, Liangyu
Qiu, Ling
author_sort Yu, Songlin
collection PubMed
description A significant bias was found when using the Siemens ADVIA Centaur XP system for measurement of 25-hydroxyvitamin D (25OHD) with VACUETTE(®) tubes with Serum Clot Activator and Gel. Here, we examined whether other commonly used tubes or temperatures affected 25OHD results obtained with the Siemens ADVIA Centaur XP system. Serum was collected into five types of vacuum blood collection tubes from three manufacturers, and 25OHD was analyzed using the Siemens ADVIA Centaur XP system and liquid chromatography tandem mass spectrometry (LC-MS/MS) immediately or after storage at 4°C or -80°C for 48 h. Significantly higher 25OHD values were found when using the Siemens ADVIA Centaur XP system with VACUETTE(®) tubes with serum clot activator and gel and VACUETTE(®) tubes with clot activator but no gel compared with VACUETTE(®) tubes with no additives. The 25OHD values in all of these tubes were not significantly different from those obtained by LC-MS/MS. Moreover, after storage at -80°C for 48 h, the values obtained in IMPROVEVACUTER(®) tubes with serum clot activator and gel significantly increased, with a mean bias of 74.6% compared with the values before storage, on analysis with the Siemens ADVIA Centaur XP system. VACUETTE(®) tubes containing additives significantly affect the accuracy of 25OHD results obtained using the Siemens ADVIA Centaur XP system. Additionally, the composition of serum collected in IMPROVEVACUTER(®) tubes was affected by freezing, resulting in different measurements when using the Siemens 25OHD assay platform.
format Online
Article
Text
id pubmed-5104342
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-51043422016-12-08 Blood Collection Tubes and Storage Temperature Should Be Evaluated when Using the Siemens ADVIA Centaur XP for Measuring 25-Hydroxyvitamin D Yu, Songlin Zhou, Weiyan Cheng, Xinqi Fang, Huiling Zhang, Ruiping Cheng, Qian Han, Jianhua Su, Wei Xia, Liangyu Qiu, Ling PLoS One Research Article A significant bias was found when using the Siemens ADVIA Centaur XP system for measurement of 25-hydroxyvitamin D (25OHD) with VACUETTE(®) tubes with Serum Clot Activator and Gel. Here, we examined whether other commonly used tubes or temperatures affected 25OHD results obtained with the Siemens ADVIA Centaur XP system. Serum was collected into five types of vacuum blood collection tubes from three manufacturers, and 25OHD was analyzed using the Siemens ADVIA Centaur XP system and liquid chromatography tandem mass spectrometry (LC-MS/MS) immediately or after storage at 4°C or -80°C for 48 h. Significantly higher 25OHD values were found when using the Siemens ADVIA Centaur XP system with VACUETTE(®) tubes with serum clot activator and gel and VACUETTE(®) tubes with clot activator but no gel compared with VACUETTE(®) tubes with no additives. The 25OHD values in all of these tubes were not significantly different from those obtained by LC-MS/MS. Moreover, after storage at -80°C for 48 h, the values obtained in IMPROVEVACUTER(®) tubes with serum clot activator and gel significantly increased, with a mean bias of 74.6% compared with the values before storage, on analysis with the Siemens ADVIA Centaur XP system. VACUETTE(®) tubes containing additives significantly affect the accuracy of 25OHD results obtained using the Siemens ADVIA Centaur XP system. Additionally, the composition of serum collected in IMPROVEVACUTER(®) tubes was affected by freezing, resulting in different measurements when using the Siemens 25OHD assay platform. Public Library of Science 2016-11-10 /pmc/articles/PMC5104342/ /pubmed/27832181 http://dx.doi.org/10.1371/journal.pone.0166327 Text en © 2016 Yu et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Yu, Songlin
Zhou, Weiyan
Cheng, Xinqi
Fang, Huiling
Zhang, Ruiping
Cheng, Qian
Han, Jianhua
Su, Wei
Xia, Liangyu
Qiu, Ling
Blood Collection Tubes and Storage Temperature Should Be Evaluated when Using the Siemens ADVIA Centaur XP for Measuring 25-Hydroxyvitamin D
title Blood Collection Tubes and Storage Temperature Should Be Evaluated when Using the Siemens ADVIA Centaur XP for Measuring 25-Hydroxyvitamin D
title_full Blood Collection Tubes and Storage Temperature Should Be Evaluated when Using the Siemens ADVIA Centaur XP for Measuring 25-Hydroxyvitamin D
title_fullStr Blood Collection Tubes and Storage Temperature Should Be Evaluated when Using the Siemens ADVIA Centaur XP for Measuring 25-Hydroxyvitamin D
title_full_unstemmed Blood Collection Tubes and Storage Temperature Should Be Evaluated when Using the Siemens ADVIA Centaur XP for Measuring 25-Hydroxyvitamin D
title_short Blood Collection Tubes and Storage Temperature Should Be Evaluated when Using the Siemens ADVIA Centaur XP for Measuring 25-Hydroxyvitamin D
title_sort blood collection tubes and storage temperature should be evaluated when using the siemens advia centaur xp for measuring 25-hydroxyvitamin d
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5104342/
https://www.ncbi.nlm.nih.gov/pubmed/27832181
http://dx.doi.org/10.1371/journal.pone.0166327
work_keys_str_mv AT yusonglin bloodcollectiontubesandstoragetemperatureshouldbeevaluatedwhenusingthesiemensadviacentaurxpformeasuring25hydroxyvitamind
AT zhouweiyan bloodcollectiontubesandstoragetemperatureshouldbeevaluatedwhenusingthesiemensadviacentaurxpformeasuring25hydroxyvitamind
AT chengxinqi bloodcollectiontubesandstoragetemperatureshouldbeevaluatedwhenusingthesiemensadviacentaurxpformeasuring25hydroxyvitamind
AT fanghuiling bloodcollectiontubesandstoragetemperatureshouldbeevaluatedwhenusingthesiemensadviacentaurxpformeasuring25hydroxyvitamind
AT zhangruiping bloodcollectiontubesandstoragetemperatureshouldbeevaluatedwhenusingthesiemensadviacentaurxpformeasuring25hydroxyvitamind
AT chengqian bloodcollectiontubesandstoragetemperatureshouldbeevaluatedwhenusingthesiemensadviacentaurxpformeasuring25hydroxyvitamind
AT hanjianhua bloodcollectiontubesandstoragetemperatureshouldbeevaluatedwhenusingthesiemensadviacentaurxpformeasuring25hydroxyvitamind
AT suwei bloodcollectiontubesandstoragetemperatureshouldbeevaluatedwhenusingthesiemensadviacentaurxpformeasuring25hydroxyvitamind
AT xialiangyu bloodcollectiontubesandstoragetemperatureshouldbeevaluatedwhenusingthesiemensadviacentaurxpformeasuring25hydroxyvitamind
AT qiuling bloodcollectiontubesandstoragetemperatureshouldbeevaluatedwhenusingthesiemensadviacentaurxpformeasuring25hydroxyvitamind