Cargando…
Modeling Canadian Quality Control Test Program for Steroid Hormone Receptors in Breast Cancer: Diagnostic Accuracy Study
The Canadian Immunohistochemistry Quality Control program monitors clinical laboratory performance for estrogen receptor and progesterone receptor tests used in breast cancer treatment management in Canada. Current methods assess sensitivity and specificity at each time point, compared with a refere...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Lippincott Williams & Wilkins
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5106078/ https://www.ncbi.nlm.nih.gov/pubmed/26200834 http://dx.doi.org/10.1097/PAI.0000000000000249 |
_version_ | 1782466991216394240 |
---|---|
author | Pérez, Teresa Makrestsov, Nikita Garatt, John Torlakovic, Emina Gilks, C. Blake Mallett, Susan |
author_facet | Pérez, Teresa Makrestsov, Nikita Garatt, John Torlakovic, Emina Gilks, C. Blake Mallett, Susan |
author_sort | Pérez, Teresa |
collection | PubMed |
description | The Canadian Immunohistochemistry Quality Control program monitors clinical laboratory performance for estrogen receptor and progesterone receptor tests used in breast cancer treatment management in Canada. Current methods assess sensitivity and specificity at each time point, compared with a reference standard. We investigate alternative performance analysis methods to enhance the quality assessment. We used 3 methods of analysis: meta-analysis of sensitivity and specificity of each laboratory across all time points; sensitivity and specificity at each time point for each laboratory; and fitting models for repeated measurements to examine differences between laboratories adjusted by test and time point. Results show 88 laboratories participated in quality control at up to 13 time points using typically 37 to 54 histology samples. In meta-analysis across all time points no laboratories have sensitivity or specificity below 80%. Current methods, presenting sensitivity and specificity separately for each run, result in wide 95% confidence intervals, typically spanning 15% to 30%. Models of a single diagnostic outcome demonstrated that 82% to 100% of laboratories had no difference to reference standard for estrogen receptor and 75% to 100% for progesterone receptor, with the exception of 1 progesterone receptor run. Laboratories with significant differences to reference standard identified with Generalized Estimating Equation modeling also have reduced performance by meta-analysis across all time points. The Canadian Immunohistochemistry Quality Control program has a good design, and with this modeling approach has sufficient precision to measure performance at each time point and allow laboratories with a significantly lower performance to be targeted for advice. |
format | Online Article Text |
id | pubmed-5106078 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | Lippincott Williams & Wilkins |
record_format | MEDLINE/PubMed |
spelling | pubmed-51060782016-11-22 Modeling Canadian Quality Control Test Program for Steroid Hormone Receptors in Breast Cancer: Diagnostic Accuracy Study Pérez, Teresa Makrestsov, Nikita Garatt, John Torlakovic, Emina Gilks, C. Blake Mallett, Susan Appl Immunohistochem Mol Morphol Research Articles The Canadian Immunohistochemistry Quality Control program monitors clinical laboratory performance for estrogen receptor and progesterone receptor tests used in breast cancer treatment management in Canada. Current methods assess sensitivity and specificity at each time point, compared with a reference standard. We investigate alternative performance analysis methods to enhance the quality assessment. We used 3 methods of analysis: meta-analysis of sensitivity and specificity of each laboratory across all time points; sensitivity and specificity at each time point for each laboratory; and fitting models for repeated measurements to examine differences between laboratories adjusted by test and time point. Results show 88 laboratories participated in quality control at up to 13 time points using typically 37 to 54 histology samples. In meta-analysis across all time points no laboratories have sensitivity or specificity below 80%. Current methods, presenting sensitivity and specificity separately for each run, result in wide 95% confidence intervals, typically spanning 15% to 30%. Models of a single diagnostic outcome demonstrated that 82% to 100% of laboratories had no difference to reference standard for estrogen receptor and 75% to 100% for progesterone receptor, with the exception of 1 progesterone receptor run. Laboratories with significant differences to reference standard identified with Generalized Estimating Equation modeling also have reduced performance by meta-analysis across all time points. The Canadian Immunohistochemistry Quality Control program has a good design, and with this modeling approach has sufficient precision to measure performance at each time point and allow laboratories with a significantly lower performance to be targeted for advice. Lippincott Williams & Wilkins 2016-11 2016-11-08 /pmc/articles/PMC5106078/ /pubmed/26200834 http://dx.doi.org/10.1097/PAI.0000000000000249 Text en Copyright © 2015 Wolters Kluwer Health, Inc. All rights reserved. This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial-No Derivatives License 4.0 (CCBY-NC-ND), where it is permissible to download and share the work provided it is properly cited. The work cannot be changed in any way or used commercially. http://creativecommons.org/licenses/by-nc-nd/4.0/. |
spellingShingle | Research Articles Pérez, Teresa Makrestsov, Nikita Garatt, John Torlakovic, Emina Gilks, C. Blake Mallett, Susan Modeling Canadian Quality Control Test Program for Steroid Hormone Receptors in Breast Cancer: Diagnostic Accuracy Study |
title | Modeling Canadian Quality Control Test Program for Steroid Hormone Receptors in Breast Cancer: Diagnostic Accuracy Study |
title_full | Modeling Canadian Quality Control Test Program for Steroid Hormone Receptors in Breast Cancer: Diagnostic Accuracy Study |
title_fullStr | Modeling Canadian Quality Control Test Program for Steroid Hormone Receptors in Breast Cancer: Diagnostic Accuracy Study |
title_full_unstemmed | Modeling Canadian Quality Control Test Program for Steroid Hormone Receptors in Breast Cancer: Diagnostic Accuracy Study |
title_short | Modeling Canadian Quality Control Test Program for Steroid Hormone Receptors in Breast Cancer: Diagnostic Accuracy Study |
title_sort | modeling canadian quality control test program for steroid hormone receptors in breast cancer: diagnostic accuracy study |
topic | Research Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5106078/ https://www.ncbi.nlm.nih.gov/pubmed/26200834 http://dx.doi.org/10.1097/PAI.0000000000000249 |
work_keys_str_mv | AT perezteresa modelingcanadianqualitycontroltestprogramforsteroidhormonereceptorsinbreastcancerdiagnosticaccuracystudy AT makrestsovnikita modelingcanadianqualitycontroltestprogramforsteroidhormonereceptorsinbreastcancerdiagnosticaccuracystudy AT garattjohn modelingcanadianqualitycontroltestprogramforsteroidhormonereceptorsinbreastcancerdiagnosticaccuracystudy AT torlakovicemina modelingcanadianqualitycontroltestprogramforsteroidhormonereceptorsinbreastcancerdiagnosticaccuracystudy AT gilkscblake modelingcanadianqualitycontroltestprogramforsteroidhormonereceptorsinbreastcancerdiagnosticaccuracystudy AT mallettsusan modelingcanadianqualitycontroltestprogramforsteroidhormonereceptorsinbreastcancerdiagnosticaccuracystudy |